Presentation is loading. Please wait.

Presentation is loading. Please wait.

| Mechanical Robustness of Cryogenic Temperature Sensors Packaged in a Flat, Hermetically Sealed Package 2015 CEC-ICMC, Tucson, AZ, C30rG-1, 1 July 2015.

Similar presentations


Presentation on theme: "| Mechanical Robustness of Cryogenic Temperature Sensors Packaged in a Flat, Hermetically Sealed Package 2015 CEC-ICMC, Tucson, AZ, C30rG-1, 1 July 2015."— Presentation transcript:

1 | Mechanical Robustness of Cryogenic Temperature Sensors Packaged in a Flat, Hermetically Sealed Package 2015 CEC-ICMC, Tucson, AZ, C30rG-1, 1 July 2015 Dr. Scott Courts

2 | Background – Strain Free Mounting 2  Historically, cryogenic thermometers required strain free mounting for optimum stability Platinum helix Sensor support Glass-platinum seal Glass sheath Platinum lead wires Fluke 5686 Glass Capsule Standard Platinum Resistance Thermometer Construction

3 | Background 3  The most stable devices were fragile and subject to damage from mechanical shock  True for Platinum, Rhodium-Iron, and Germanium resistance thermometers  With a small sacrifice in stability, newer devices can be packaged much more robustly This work examines the mechanical robustness of cryogenic temperature sensors mounted in a flat, hermetically sealed package

4 | Device types tested in –SD package  All devices manufactured by Lake Shore Cryotronics, Inc.  2 Models packaged in flat, hermetically sealed –SD package: – Cernox  model CX-1050-SD – Diode model DT-670-SD 4 -SD package 2 mm × 3 mm × 1 mm height

5 | Device type tested in copper A-can  For comparison, also tested a Cernox  model CX-1050-AA 5 CX-AA package 3 mm diameter × 8 mm long

6 | Test Protocol – 20 Devices of Each Type 6 Calibration, 1.4-325 K Drop Test Vibration Mechanical Shock Calibration, 1.4-325 K

7 | Test Protocol – 20 Devices of Each Type 7 Calibration, 1.4-325 K Drop Test 20 cm, 50 cm, 1 m, and 4 m, + control samples Vibration Mechanical Shock Calibration, 1.4-325 K

8 | Test Protocol – 20 Devices of Each Type 8 Calibration, 1.4-325 K Drop Test 20 cm, 50 cm, 1 m, and 4 m, + control samples Vibration Mechanical Shock Calibration, 1.4-325 K MIL-STD-202, Method 214, Table 2, Condition H Overall G RMS = 34g’s, 60 sec along each axis

9 | Test Protocol – 20 Devices of Each Type 9 Calibration, 1.4-325 K Drop Test 20 cm, 50 cm, 1 m, and 4 m, + control samples Vibration Mechanical Shock Calibration, 1.4-325 K MIL-STD-883, Method 2002, Condition B G PEAK =1,500g’s, 3 shocks along each axis MIL-STD-202, Method 214, Table 2, Condition H Overall G RMS = 34g’s, 60 sec along each axis

10 | CX-1050-SD Drop Test Results 10 No discernable effects for any drop height

11 | CX-1050-SD Vibration/Mechanical Shock Results No discernable effects from vibration or mechanical shock 11

12 | DT-670-SD Drop Test Results 12 No discernable effects for any drop height

13 | DT-670-SD Vibration/Mechanical Shock Results 13 No discernable effects from vibration or mechanical shock

14 | CX-1050-AA Drop Test Results 14 No discernable effects for any drop height

15 | CX-1050-AA Vibration Test Results 15 Negative trend in  T, but results skewed by two devices. Offset uncorrelated to initial drop height.

16 | CX-1050-AA Mechanical Shock Test Results 16 Negative trend in  T, but results skewed by three devices. Offset uncorrelated to initial drop height.

17 | CX-1050-AA Offset Cause 17 Negative  T  Positive  R; Most likely cause of increased resistance is stretching of the  3 mm long 2-lead portion of the 50 μm diameter gold lead connecting the sensor die to the phosphor bronze post. (For comparison, the connecting gold leads in the –SD package are 25 μm diameter by  0.75 mm long)

18 | Conclusions This work has shown:  For models Cernox  CX-1050-SD and diode DT-670-SD – No discernable effect when dropped from heights up to 4 m – No discernable effect from vibration with G RMS = 34Gs – No discernable effect from mechanical shock of G PEAK = 1,500Gs 18

19 | Conclusions This work has shown:  For models Cernox  CX-1050-SD and diode DT-670-SD – No discernable effect when dropped from heights up to 4 m – No discernable effect from vibration with G RMS = 34Gs – No discernable effect from mechanical shock of G PEAK = 1,500Gs  For model Cernox  CX-1050-AA – No discernable effect when dropped from heights up to 4 m – A negative trend in  T after vibration with average offset approaching -60 mK at 300 K, but uncorrelated to drop height – A negative trend in  T after mechanical shock with average offset approaching -60 mK at 300 K, but uncorrelated to drop height 19

20 | Thank You! 20


Download ppt "| Mechanical Robustness of Cryogenic Temperature Sensors Packaged in a Flat, Hermetically Sealed Package 2015 CEC-ICMC, Tucson, AZ, C30rG-1, 1 July 2015."

Similar presentations


Ads by Google