2 nd ANNUAL ARDENT WORKSHOP Quality assessment of hadrontherapy fields with TEPCs Paolo Colautti, INFN 2 nd Annual ARDENT Workshop, Milano, 14-18 October 2013
2
P.Colautti 2 nd ARDENT workshop. Milano 14-18 October 2013 The radiation quality concept The radiation field quality is meant as a physical measurable quantity, which is significant for primary effects on a biological system We define the measured y-spectrum in 1- 2 µm sites as the radiation-field microdosimetric quality Some detectors measure only, which is used as quality mark of the radiation field. The couple of values and is a better quality mark of the radiation field. RBE µ, the microdosimetric assessment of RBE, can be an accurate quality mark for a given biological end-point.
3
P.Colautti 2 nd ARDENT workshop. Milano 14-18 October 2013 Therapeutic beam constrains Fluence rates Proton continuos beam BNCT facility 1 cm TEPC ought to measure at counting rates bigger than 10 6 s -1 1 mm TEPC will measure at counting rates bigger than 10 4 s -1 CNAO protons per spill CNAO carbon ions per spill
4
P.Colautti 2 nd ARDENT workshop. Milano 14-18 October 2013 Energy calibration technique in ~ 1cm cylindrical TEPCs Vacuum tide Al cylinder Tissue-equivalent conductive plastic
5
P.Colautti 2 nd ARDENT workshop. Milano 14-18 October 2013 Energy calibration technique in ~ 1mm cylindrical TEPCs -source Gas IN Gas OUT Cathode H.V. Anode pulse Guard tubes H.V.
6
P.Colautti 2 nd ARDENT workshop. Milano 14-18 October 2013 Measurements at the Nice therapeutic proton beam
7
P.Colautti 2 nd ARDENT workshop. Milano 14-18 October 2013 Insulator surfaces damage without electric-field tubes 75 µm
8
P.Colautti 2 nd ARDENT workshop. Milano 14-18 October 2013 Solution: a cavity inside the insulator to switch off the the electronic avalanche 0.9 mm 0.150 mm 0.200 mm
9
P.Colautti 2 nd ARDENT workshop. Milano 14-18 October 2013 Mini TEPC of 2.7 mm of external diameter
10
P.Colautti 2 nd ARDENT workshop. Milano 14-18 October 2013 Vacuum and gas flow apparatus Membrane pump Pressure readout Exhaust Gas MFC MFC controller Turbo pump Solenoid valve
11
P.Colautti 2 nd ARDENT workshop. Milano 14-18 October 2013 Electronic chain HV anode HV cathode Pulser +18/-6 volt Low LET amplifier Medium LET amplifier High LET amplifier Low LET MCA Medium LET MCA High LET MCA
12
P.Colautti 2 nd ARDENT workshop. Milano 14-18 October 2013 Electronic chain
13
P.Colautti 2 nd ARDENT workshop. Milano 14-18 October 2013 Before measuring: linearity checks Integral linearity High LET
14
P.Colautti 2 nd ARDENT workshop. Milano 14-18 October 2013 Initial data from the 3 MCA
15
P.Colautti 2 nd ARDENT workshop. Milano 14-18 October 2013 Volt calibration
16
P.Colautti 2 nd ARDENT workshop. Milano 14-18 October 2013 Logaritmic compaction 1 st 2 nd B th log y i log y i+1 1 st ….. Equal log intervals Example: B = 60 First decade y 1 = 1.0391, y 2 = 1.0798, y 3 …. Δ 1 = 0.04065 Second decade y 1 = 10.0391, y 2 = 10.0798, y 3 …. Δ 1 = 0.4065 Original counts Count density in compacted intervals
17
P.Colautti 2 nd ARDENT workshop. Milano 14-18 October 2013 Sub-spectra junction Sub-spectra junction is not feasible if they do not superimpose
18
P.Colautti 2 nd ARDENT workshop. Milano 14-18 October 2013 The pulse-height frequency distribution
19
P.Colautti 2 nd ARDENT workshop. Milano 14-18 October 2013 The pulse-height weighted distribution
20
P.Colautti 2 nd ARDENT workshop. Milano 14-18 October 2013 Lineal energy calibration ICRU 49 data 142 keV/µm Calibration factor: 142/475 = 0.299
21
P.Colautti 2 nd ARDENT workshop. Milano 14-18 October 2013 ∆E calculation in CSDA
22
P.Colautti 2 nd ARDENT workshop. Milano 14-18 October 2013 Maximum ∆E-lost calculation in CSDA
23
P.Colautti 2 nd ARDENT workshop. Milano 14-18 October 2013 Lineal-energy dose distribution Fractional part of the visual area is the fractional contribution to the the absorbed dose by the corresponding y-events Blue points are used to linearly extrapolate the frequency y- values down to 0.01 keV/µm
24
P.Colautti 2 nd ARDENT workshop. Milano 14-18 October 2013 Weighting the dose distribution with a biological-significant function Fractional part of the visual area is the fractional contribution to the effective-dose of corresponding y- events The weight is 1 for y-values < 10 keV/µm The weight is 200 keV/µm
25
P.Colautti 2 nd ARDENT workshop. Milano 14-18 October 2013 Splitting the dose spectrum in its neutron and gamma components
26
P.Colautti 2 nd ARDENT workshop. Milano 14-18 October 2013 Quality of the conjunctive-melanoma proton-therapeutic beam
27
P.Colautti 2 nd ARDENT workshop. Milano 14-18 October 2013 The therapeutic beam quality Iso dose map Iso effective-dose map Iso RBEµ map
28
P.Colautti 2 nd ARDENT workshop. Milano 14-18 October 2013 Two mini TEPCs for quality measurements of BNCT radiation fields
29
P.Colautti 2 nd ARDENT workshop. Milano 14-18 October 2013 The twin TEPC for BNCT
30
30 BNCT dose components twin TEPC measurements
31
P.Colautti 2 nd ARDENT workshop. Milano 14-18 October 2013 Suggested lectures Glenn F.Knoll. Radiation Detection and Measurements. John Wiley & Sons, NY, 1979 ICRU Report 36. Microdosimetry. 1983