Staple-Line Reinforcement Dan Geisler, MD, FACS, FASCRS West Penn Allegheny Health System
2
2 Disclosures CovidienConsultant
3
3 Staple Line Reinforcement WHY? Anastomotic Leak Staple Line Bleeding
4
4 Staple Line Reinforcement WHAT? Synthetic, bioabsorbable staple-line reinforcement Collagen matrix Polyglycolic acid
5
5 Staple Line Reinforcement WHAT? Gore Seamguard Synovis Staple Line Reinforcement Cook biodesign staple line reinforcement Covidien Duet TRS
6
6 GORE Seamguard Bioabsorbable Staple Line Reinforcement Synthetic Absorbable Sleeves (“socks”) held by non-absorbable polyester braided suture Increases thickness of tissue by up to 0.5mm
7
7 Staple Line Reinforcement Synovis® Peri-Strips Dry® Biologic tissue prepared from bovine pericardium (U.S. cattle) Strips attached by gel (“glue”) Increases thickness of tissue by 0.5mm to 1.5mm Synovis® Peri-Strips Dry® with Veritas® Collagen Matrix Prepared from dehydrated bovine pericardium (U.S. cattle) Similar assembly to Peri-Strips Dry®
8
8 Staple Line Reinforcement PeriPatch™ Aegis Processed Bovine Pericardial Staple Line Reinforcement Made ONLY for Ethicon staplers (ETS/Flex 45 and Echelon 60) Easier to assemble, but more preparation involved (pre-soak) IFU recommends use of 15mm trocar PeriPatch™ Aegis EQ Identical in design to PeriPatch™ Aegis, but it is made from equine pericardium
9
9 Staple Line Reinforcement DUET Thinnest On the Market will increase the thickness of the tissue to be stapled by approximately 0.12mm – 0.18mm Clear easily view staple formation Synthetic Absorbable Consistent Thickness Optimal Absorption Profile Strength During Critical Wound Healing Time
10
10 Staple Line Reinforcement WHAT?
11
11 Covidien | August 21, 2015 | Confidential11 | Duet Tissue Reinforcement – Absorption Profile Reinforcement material is a synthetic absorbable film prepared from synthetic polyester composed of glycolide, dioxanone and trimethylene carbonate Measurable mechanical strength through 14 days Absorption should be complete between 90-110 days
12
12
13
13
14
14 Staple Line Reinforcement WHEN? High-risk anastomosis Bleeding Anastomotic leaks
15
15
16
16 DATA?
17
17 Staple Line Reinforcement Gastric Bypass N = 1073 (568 bovine collagen matrix) Abdominal septic events 4% versus 1% Staple line leakage Purulent drainage Abscess Callery CD, Filiciotto S, Neil KL
18
18 Staple Line Reinforcement Colorectal Anastomosis N = 117 laparoscopic + open Circular bioabsorbable Seanguard 4 intraoperative leaks Use of Seamguard may result in a lower incidence of anastomotic leakage Portillo & Franklin
19
19 Staple Line Reinforcement Colorectal Anastomosis “most…need to be accepted and optimized” “not much experience…further investigation is needed”
20
20 Staple Line Reinforcement Seems promising Preventing leakage + bleeding More studies are needed