Download presentation
Presentation is loading. Please wait.
1
Prof. Zoltán Juvancz D.Sc.
Chromatography Prof. Zoltán Juvancz D.Sc.
2
The first chromatography by Tswett
3
Aim of chromatography The compounds of interest has to be separated from matrix components. The signal to noise ratio must be as high as possible. The time consumption of analysis must be short.
4
Gas chromatograms of 104 volatile organic compounds (VOC)
5
If the peaks are narrow, more compounds can be separated
6
Definitions Chromatography is a separation method.
(Chromatography is not an identification method like NMR, IR, MS) Chromatography consist of two phases: mobile and stationary phase. Mobile phase is forced along the column from injection to detector as a flowing media. Stationary phase is anchored to the column wall or to the particles, which are packed into the column.
7
Chromatographic process
The injected sample is dragged by mobile phase along the column. The components of the sample distribute between the stationary phase and mobile phase. If X compound has bigger affinity to stationary phase than affinity of Y compound to stationary phase, the X compound elutes later than Y from column. The sharp injected peaks become broader and broader during their run.
8
Schematic view of aGC
9
Cross section of open tubular (capillary)
General parameters Column length: 5-100 m Column diameters: mm Film thickness: 0.15 – 5 µm Fused silica wall has no metal content. The metals can cause destruction of sample and stationary phase with their catalytic effects.
10
Advantages of chromatography
Exact quantitative analysis is done even from trace compounds. Disturbing effects of matrix compound can be eliminate. Small material consumption The quantization has a broad linearity range. Analyses of several compound can be done during one run. Chromatography is a fast analysis method. On-line coupling are routinely solved to compound identification methods, Well establishes instrumentation with high level automation is commercially available.
11
Trace analysis
12
Single cell analysis
13
Determination of trace compound from cpmplex matrix
14
Analyses of chlorinated pesticides with GC x GC system
15
Determination 106 compounds during one run
16
Determination of trace impurity of a compounds
Determination of compounds having different magnitudes in their volume.
17
Fast analysis
18
On-line GC/MS coupling with selected ion monitoring
19
Chromatography is highly automitized
Auotomatic processes allow whole day work without human inspection. The automated processes have high reproducibility and accuracy.
20
On line HPLC/MS coupling
21
Determination of origin of extasy tablets
22
Recognition of adulteration of bergamot oil
23
Elution profile of a chromatogram
Intensity Time
24
Chromatographic peak An ideal chromatographic has Gaus shape
Qualitative measure: retention time (tR) Quantitative measure: Peak area (A)
25
Chromatographic peak A certain compound shows same retention times under same conditions (column, flow, temperature etc.) independently from its injected amount (in the linear range of Langmuir isotherm). The identity of the compounds is partly based their retention time. (Qualitative parameter) Peak area is linear function of the quantity of certain compound (Quantitative parameter).
26
Chromatographic peak An ideal chromatographic has Gaus shape
Qualitative measure: retention time (tr) Quantitative measure: Peak area (A)
27
Chromatographic peak The given compound shows same retention times under same conditions (column, flow, temperature etc.) independently from its injected amount (in the linear range of Langmuir isotherm). The identity of the compounds is partly based their retention time. Peak area is linear function of the quantity of certain compound.
28
Chromatographic expressions
Retention time: tR (qualitative measure) Hold up time: tm (The time, what the compounds spend in mobile phase) Peak with in half high: wh Peak area (A) (quantitative measure) Theoretical plate high: N (efficiency, sharpness ) N = 5,54 (tR/wh)2 Capacity ratio: k (Strength of stationary phase) k = tR-tm/tm
29
Theoretical plate height
The peaks become broader and broader during the chromatography The later eluting peak are broader than early ones. The theoretical plate height is same for all peaks. The theoretical plate height characterize the chromatographic system. The capacity ratio shows the interaction of compound of interests (distribution between the stationary and mobile phases.
30
Chromatographic expressions
Selectivity: (Measure of interaction ratio between two compound and stationary phase) = tR2’/ tR1’ Resolution: Rs (Measure the separation of two peaks) Rs = 1,177 (tR1-tR2)/(Wh1+Wh2) Rs : 1.5 baseline resolution
31
Increasing of capacity ratio with increasing of volume of stationary phase
32
Changing of capacity ratio with decreasing solvent strength of mobile phase
Decreasing solvent strength results in longer retention times.
33
Distribution of the compounds between two phases
Kd = Cm/Cst = p/q E = KdV/(1 +KdV) Where Kd distribution constant C m concentration of a compound in mobile phase Cst concentration of a compound in stationary phase E extraction ratio V phase ratio Two compounds can be separated if their distribution constants are not equal (Kdx ≠ Kdy).
34
Distribution of a compound between the mobile and stationary phases
35
Material distribution between the stationary phase and mobile phase
K= p/q
36
Band broadening in mobile phase
37
Stones in river Light stone water flow Heavy stone
base water flow Heavy stone Simulation of separation
38
Material distribution between the stationary phase and mobile phase
39
Separation processes Number of the equilibrium Intensity of signal
Place of materials from injection point
40
If the peaks are narrow, more compounds can be separated
41
The slow mass transfer between the phases causes peak broadening
Resistance of mass transfer processes cause that the material zone are retained in stationary phase from the material zone in mobile phase.
42
Band broadening in open tube
The band of sample is getting broader caused by slow diffusion from the middle of mobile phase to the stationary phase.
43
Band broadening caused by uneven flow
The flow velocity is not uniform across the column.
44
Band broadening caused by badly washed holes
The compounds come in and out from the holes slowly. The diffusion processes are much slower than the flow of the mobile phase.
45
The ideal flow of mobile phase is result of compromises
HETP: high equivalent theoretical plate, column HETP: N/L (L: length of ) HETP =A +B/u + Cu A=Eddy Diffusion, B =Molecular Diffusion, C =Resistance to mass transfer, U =Linear velocity of mobile phase (cm/s)
46
Resolution vs. efficiency, capacity ratio, selectivity
P. Sandra JHRC 12 (1989) 82.
47
Resolution-efficiency- selectivity
48
Ways how to improve the resolution
Increase the efficiency Longer columns Smaller diameter of column or smaller diameter of particles Thinner stationary phase Optimated flow of mobile phase Increase the capacity ratio (3-10) Bigger amount of stationary phase Weaker mobile phase (HPLC) Lower analysis temperature (GC) Increased selectivity More selective phases Lower analysis temperature Derivatization
49
Higher efficiency with smaller diameter column
GC Chirasil-Dex 10 m
50
Higher column length Longer column give better resolution,
but longer analysis times.
51
Increasing of capacity ratio with increasing of volume of stationary phase
52
Weaker mobile phase increases the capacity ratio
53
Selectivity steeply increases with decreasing temperature
54
Selectivity vs. analysis temperature
= tR2’/ tR1’ ln α = Δ(ΔS0)/R- Δ(ΔH0)/RT α: selectivity S: entrophy H: entalphy R: gas constant T: absolute temperature
55
Selectivity increase with derivatisation
GC Oszlop: 10 m x 0.1 mm CSP: Chirasil-Dex Vivő gáz: H2 Hőmérséklet: 180C
56
Selectivity increase using selectivity stationary phase
57
Resolution The small value of resolution can produce false
retention time. Peaks with different magnitudes need more than Rs 1.5 value for their baseline separations.
58
„Minor peak first” is important with peaks in different magnitudes
GC Nem csak a szórás, de az abszolult érték is más. Ok a tailing, ami a csatlakozás és adszorpciós helyek. A: 20 m x 0,2 mm, ChNEB, 160° C. B: 20 m x 0,2 mm, ChDA, 150° C.
59
The peak area is the base of quantitative analysis
The area of a peak is the sum of regularly measured signals Threshold value of Steepness Area Sampling frequency Mode of baseline corrections
60
Distorted peak shapes
61
The determination of area of badly separated peaks is errorneous
V.R. Meyer, Chromatographia 40 (1995) 15.
62
Asymmetric peaks give false area count.
63
Difficulties in area determination
Noisy baseline Drifting baseline
64
Overloding If the concentration of a compound exceed the saturation value overloading effects occures. Errors are caused the overloading: Incorrect retention times, Incorrect peak areas
65
Effect of overloading
66
Overloading destroys the resolution
67
Compensation of overloading
68
Different types of chromatography according to mobile phase
Gas chromatography, (GC) Liquid chromatography (LC), High performance liquid chromatography (HPLC) Supercritical fluid chromatography (SFC) Electro kinetic chromatography (EKC)
69
Chromatogry according to the column
Packed column Open tubular column (capillary) Chip Thin layer (TLC) Chromatography according to interaction types Distribution Adsorption Exclusion Ion exchange
70
The role of different interaction types in various chromatographic modes
GC SFC HPLC EKC Dispersion ++++ +++ ++ + - Dipole-dipole Hydrogen bridge Ionic / Repulsion A királis felismerő kölcsönhatásoknak legjobban megfelelő módot érdemes kiválasztani. GC a funkciós csoport nélküli szénhidrogén enatiomerek elválasztására is alkalmas. The GC is mostly (70-95%) boiling point selective method.
71
Advantages of various chromatographic modes
Tulajdonság GC SFC HPLC EKC Efficiency ++++ +++ ++ Analyses temperature + Variability of mobile phase / Speed of analyses Sensitivity Established instrumentation GC is very efficient (long columns), but less selective method (no mobile phase selectivity).
Similar presentations
© 2024 SlidePlayer.com. Inc.
All rights reserved.