Download presentation
1
Fluoroscopy Notes Rad Tech 290
2
Fluoroscopy Notes Ch 1 Approximately 5% of the US population has a fluoro procedure each year The average number of fluoro exams per person is 1.3 The average number of spot films is 4.6
3
The MC exam is a GI tract at 53%
A 2 minute UGI exam can produce an exposure ranging from 5-15 rads, comparatively a KUB is between mrads. Fluoro is defined as a rad exam utilizing fluorescence for the observation of the transient image.
4
Fluoro was first used as a dynamic procedure
Fluoro was first used as a dynamic procedure. Second as a means of positioning for spot films. Medical exposure accounts for about 20% of the total radiation people receive. Even though the percentage is small, for medical exposure, it is the only exposure that is controllable. FLUOROSCOPY TO POSITION PATIENTS IS PROHIBITED.
5
Notes Chapter 2 Factors Directly Affect Exposure
mA kVp Collimation Filtration Exposure time Total fluoro time Target to panel distance (TPD) Patient to II distance Sensitivity of the image receptor Essentially speed RSV
6
The following will reduce exposure
Collimating Last frame hold Shortest possible patient to II distance Highest possible kVp Pulsed fluoroscopy Using the largest II mode with collimation
7
Factors indirectly influencing exposure
Room illumination Image receptor quality Absorption of the table top
8
mA 0.5 – 5 mA Usually 1 – 3 mA Spot films 100 mA or higher Output and dose are directly proportional to mA
9
kVp Maximum photon energy Beam quality Penetrability of the beam
Tube potential
10
collimation Required by law
Image quality improves as the beam is collimated
11
Collimation Collimate tightly to the area of interest.
Reduces the patient’s total entrance skin exposure. Improves image contrast. Scatter radiation to the operator will also decrease.
12
Factor affecting staff doses
FIELD SIZE DEPENDENCE Scattered dose rate is higher when field size increases 11x11 cm 17x17 cm 17x17 cm 100 kV 0.8 mGy/h 1.3 mGy/h 1 mA 0.6 mGy/h 1.1 mGy/h 0.3 mGy/h 0.7 mGy/h 1m patient distance Patient thickness 18 cm
13
filtration If the tube is operated above 125 kVp, 3 mm Al eq is required. Filtration reduces patient dose
14
Source to table top target to panel distance
Cannot be less than 12” and should be 18” Mobiles are required to be at least 12” Fixed units, 15”
15
Patient to II distance The closer the II, the lower the dose
This is more pronounced with fixed units. Decreases the SID
16
Tabletop Less than 1 mm Al eq at 100 kVp Exposure switch Dead man type
17
Primary protective barrier
The II is the primary barrier and must have 2 mm Pb eq for systems operating above 125 kVp The II has to be in place for the tube to energize
18
Protective Actions Bucky Slot Cover Protective curtains
Automatically covered, 0.25 mm Pb eq Protective curtains 0.25 mm Pb eq Not required on c-arms Scatter at 1 foot can reach 500 mrad/hr
19
Allowable exposure rates
Cannot exceed 5 rad/minutes Unless, ABC or image recording Cumulative timer Cannot exceed 5 minutes Illumination
20
II Considerations Purpose Brightness Gain
The basic purpose of the II is to make the fluoro image brighter When the image is brighter it is easier to visualize structures Brightness Gain Minification gain multiplied by electronic (flux) gain
21
II Facts Input phosphor, cesium iodide Photcathode, danium antimony
Output phosphor, zinc cadnium sulfide
22
Image Quality Issues Quantum Mottle Contrast Resolution
Caused by too few photons Contrast Subject Detector Image Resolution
23
The first step in obtaining optimum contrast in an x-ray image is to adjust the x-ray beam spectrum for the specific anatomy and clinical purpose. The penetration and the resulting contrast of a specific object or structure in the body generally depends on the photon energy spectrum. Contrast is not the only thing that must be considered in selecting the spectrum for a specific procedure. The spectrum also affects the penetration through the body section being imaged. This has a significant effect on the radiation dose to the patient. Also, as the penetration through a body section is reduced, the amount of radiation required from the x-ray tube is increased with a resulting increase in x-ray tube heating. Effect of X ray Beam Penetration on Contrast, Body Penetration, and Dose t
24
Dose vs. Noise 15 µR per frame 24 µR per frame 2 µR per frame
25
Magnification tubes, Multi-mode
Distortion Size, shape, pincushion Lag Vignetting Less bright at the edges than center of image Magnification tubes, Multi-mode Variable FoV
26
Closed Circuit TV Systems
Camera MC is the vidicon Camera control unit Video amplifier Monitor CRT, etc.
27
Cinefluoroscopy Synchronization Framing frequency F-number
Record with the x-ray pulses Framing frequency Division of 60 The higher the rate the higher the dose F-number Video disk recording (electronic radiography) Exposure ends when image is formed Basically fluoro phototiming 95% dose reduction
28
Video tape Spot films Instant playback and no additional dose
Conventional cassettes Photospot cameras ½ to 1/3 dose of convent. Cass. Lower image quality
29
Accessories Gonadal shields Grids Cassettes Cine film
Required when possible Grids Fluoro uses low ratio grids Cassettes Cine film Per frame basis 10 x the dose than fluoro
30
Factors affecting an Increase in Scatter
High kVp Large field size Thick body part
31
Advantages of 3 phase and medium/high frequency generators
Relatively high mA Higher effective kVp Near constant potential Less ripple
32
Notes for Chapter 3 Fluoro Image Production
Fluoro units have 2 basic components X-ray tube Image intensifier
33
Fluoro X-ray Tube Regular rotating anode x-ray tube Runs at a lower mA
Less than 5 mA Small focal spot Possible because of the low mA
34
II The primary purpose is to increase the brightness of the fluoro image Components Glass envelope that provides a vacuum Input layer Converts x-ray photons to electrons Electronic (electostatic) lens Output layer
35
Input layer Converts x-ray photons to light photons
Light photons then strike the photocathode and convert into electrons Electrons are then accelerated across the II Electrons strike the output phosphor and are converted back into light photons
36
The image intensifier (I.I.)
I.I. Input Screen Electrode E1 Electrode E2 Electrode E3 Electrons Path I.I.Output Screen Photocathode +
37
Image intensifier component
Input screen: conversion of incident X Rays into light photons (CsI) 1 X Ray photon creates 3,000 light photons Photocathode: conversion of light photons into electrons only 10 to 20% of light photons are converted into photoelectrons Electrodes : focalization of electrons onto the output screen electrodes provide the electronic magnification Output screen: conversion of accelerated electrons into light photons
38
Image Intensifier Magnification Modes
Same area Output phosphor Input Phosphor 9 inch field 6.5 inch field
39
RELATIVE PATIENT ENTRANCE DOSE RATE
FOR SOME UNITS IMAGE INTENSIFIER Active Field-of-View (FOV) 12" (32 cm) 100 9" (22 cm) 6" (16 cm) 300 4.5" (11 cm) 400
40
Brightness Gain BG is the product of minification gain and flux(electronic) gain Minification Gain Input phosphor dia.2/output phosphor dia2 Making the image smaller will make it brighter. The same number of photons are contained in a smaller area In most IIs the output phosphor is 1 inch.
41
Flux Gain Caused by the conversion efficiency of the output phosphor and the acceleration of the electrons across the II As the electrons accelerate they gain kinetic energy Flux gain is usually between 50 and 150.
42
Measuring Brightness Gain
The actual measurement is done by calculating the conversion factor Intensity of output phosphor (candelas)/mrads/sec Brightness gain will deteriorate 10% annually. This will ultimately decrease image contrast
43
Beam Splitter Mirror 10% of the output light goes to the vidicon (video camera) the remainder goes to the photospot device. NOTE: not all units have a beam splitter
44
Vignetting and Pincushion Distortion
The loss of shape at the edges of the fluoro image Vignetting Loss of brightness at the edge of the image
45
Veiling Glare Occurs when the light from the output phosphor ‘reflects’ back into the II. Remember, the photocathode is stimulated by light, so light ‘reflecting’ from the output phosphor would also trigger electron production. Decreased contrast results
46
Keeps light output of the II constant.
Automatic Brightness Stabilization (ABS) Automatic Brightness Control (ABC) Keeps light output of the II constant. Brightness of the image varies with changes to kVp and mA. Increase mA increase brightness; direct relationship Increase kVp 10% double brightness
47
Brightness Sensing II photocathode current
Television camera signal sensing Lens coupled phototube sensing
48
Types of ABS Variable mA, preset kVp Variable mA with kVp following
Set the kVp and the unit adjusts mA Variable mA with kVp following If the mA range is exceeded the unit will automatically adjust the kVp to compensate Variable kVp, preset mA Set the mA and the unit adjusts kVp Variable kVp, variable mA
49
Closed Circuit TV Systems
Camera Camera control unit Power supply and video amplifier Monitor
50
Cameras Vidicon MC, inexpensive, lag, 525 raster lines Plumbicon
Cardiac cath labs Fixed gain (better contrast) and low lag Increased quantum mottle Image orthicon Not widely used CCD Solid state semiconductor Small, low power consumption, low price, long life
51
Photoconductive camera tube
Focussing optical lens Input plate Steering coils Deviation coil Alignement coil Accelarator grids Control grid Electron beam Video Signal Signal electrode Field grid Electrode Electron gun Iris Photoconductive layer
52
Schematic structure of a charged couple device (CCD)
53
Monitor 525 lines 30 times per second
Combined with the video camera improves image contrast
54
TV Image Quality Horizontal resolution Vertical resolution Contrast
Brightness Lag
55
Horizontal resolution
Bandwidth or bandpass Increase frequency bandwidth increase horizontal resolution Vertical resolution Determined by number of scan lines Kell factor Ratio of vertical resolution and scan lines
56
Contrast Brightness Lag Adjust contrast first and brightness after
Occurs when the II is moved rapidly.
57
Dynamic Image Recording
Video Tape 2 advantages Instant replay No additional patient exposure Disadvantages Poor image quality, fixed frame rate,
58
Cinefluoroscopy 16 or 35 mm (MC high patient dose, better image)
Synchronization Camera shutters open at the same rate as x-ray pulses Framing rate F-number The lower the number the more light hitting the camera the lower the patient dose; however, more distortion at the edges
59
Framing Underframing should be avoided
Exact framing, diameter of the II fits in the shortest dimension of the film Overframing, diameter of the II fits the largest dimension of the film. Part of the image is lost Total overframing, diameter of the II is equal to the diagonal of the film
60
Static Image Recording
Video disk Last image freeze (hold) (sticky fluoroscopy) Electronic radiography, similar to AEC x-ray only until image is made. Decrease dose up to 95% 1 to 30 frames per second Spot film Conventional cassettes
61
Spot film Photospot Cassette spots
Image is taken from the II output phosphor Dose 20 – 50 X higher per frame than fluoro because of higher mA Currently, 70 mm roll, 105 roll, 100 mm chip 105 mm roll ½ the dose of cassette spots Cassette spots Slower ‘frame’ rate Higher dose and better spatial resolution
62
Digital Fluoroscopy Digital image is obtained from the output phosphor. A vidicon then a digital image processor Or, digital video camera Digital photospot Instant playback, possible image enlargement
63
Pulsed Fluoroscopy Variable frame rates are possible with a corresponding decrease in patient dose
64
High level (boost) fluoroscopy
Higher tube currents than normal 10-20 mA usual 40 mA potentially Increase patient dose 2-10 times reg. Fluoro 10-50 rads per minute Limited to 20 rad/minute unless recording the image Key points Special activation required, audible signal, dose rate limited to 20 rad.minute
67
Notes from Chapter 4 Conducting the Fluoro Exam
Operator dose is directly proportional to patient dose Image brightness is directly proportional to dose rate at input phosphor
68
Technical factors which directly influence dose rate at the table top
mA kVp Collimation Filtration Exposure time Target panel distance
69
Technical factors which indirectly influence dose by affecting technical factors
Room lighting Image receptor quality tabletop
70
Collimation A border needs to be visible when the II is 14 inches above the tabletop and the collimator is fully opened With an automatic collimator, a border should always be visible Image is not brighter with a less collimation (bigger field size)
71
Filtration Total filtration
2.5 mm Al eq < 125 kVp 3.0 mm Al eq at 125 kVp and above Total filtration includes inherent and added Exposure rate should be less than 2.2 rads/min at 80 kVp HVL
72
Allowable exposure rates
Limited to 5 rads per minute If the unit has ABC/ABS then 10 rads/minute is allowed However, if the unit has ‘boost’ then the limit is 5rads/minute ABC/ABS units have to be checked by a physicist annually Also have to have weekly fluoro checks of mA and kVp No ABS 3 year check by physicist
73
TPD TPD increases from 12 to 18 inches
Pt. Dose decreases by 30% II as close as possible to the patient
74
Room Lighting Affects visual acuity
Photopic acuity is 10 X better than scotopic acuity Day versus night vision Normal viewing distance is 12 – 15 inches Image recognition in 0.2 seconds
75
Gonadal Shields 0.5 mm Pb eq
97% effective at 100 kVp and 3 mm Al filtration
76
Notes from Chapter 5 Basic Operational Procedures
Minimize exposure time by utilizing short looks Use the cumulative timer Use the highest applicable kVp Collimate Use mag and boost only when necessary Use last image hold
77
Use a photospot instead of cassettes
Use video tape Use II with good contrast Monitor the TV monitor for brightness and contrast Minimize the pt. II distance Position the II prior to exposure
78
Prevent pt. Motion with instructions
Use gonadal shielding when possible Use compression devices
79
Factor affecting staff doses
ANGLE DEPENDENCE Scattered dose rate is higher near the area into which the X-ray beam enters the patient 100 kV 0.9 mGy/h 1 mA 0.6 mGy/h 11x11 cm 0.3 mGy/h 1m patient distance patient thickness 18 cm
80
Factor affecting staff doses
DISTANCE VARIATION mGy/h at 0.5m mGy/h at 1m Scattered dose rate is lower when distance to the patient increases 100 kV 1 mA 11x11 cm
81
Factor affecting staff doses
Tube undercouch position reduces, in general, high dose rates to the specialist’s eye lens X-Ray tube mGy/h 100 kV 2.2 (100%) 1 m 2.0 (91%) 20x20 cm 1.3 (59%) mGy/h 1 Gy/h (17mGy/min) 1.2 (55%) 1.2 (55%) 1.2 (55%) 1 Gy/h 1m patient distance (17 mGy/min) 1.3 (59%) 20x20 cm 2.2 (100%) 100 kV 1 m 1m patient distance X-Ray tube
82
Mobile Concerns Audible indicator
Cumulative timer visible on the monitor Video storage Last frame hold Longest possible TPD
83
Notes for Chapter 6 Pediatric Fluoroscopy
Motion Sedation Mechanical devices Personnel and Parental Protection Everyone in the room needs lead Gonadal shielding Artifacts
84
Photospots instead of cassette spots if possible
ABS/ABC Watch putting the II directly over large concentrations of contrast Distance Collimation Photospots instead of cassette spots if possible
85
Notes for Chapter 7 Mobile Fluoroscopic Equipment
Primary beam is intercepted by the II If the II is used routinely in a single location it needs to have secondary shielding SSD has to be at least 12 inches Must have an II Collimation has to be used or unit will not energize Unit cannot energize unless the II is in the primary beam
86
Maximum dose rate of 5 rads/minute
Personnel monitoring in required for all persons operating fluoro equipment Protective aprons are required if exposed to more than 5mrads/hour Boost mode should only be used after areas of interest have been localized
87
Notes for Chapter 8 Responsibilities of X-ray Supervisor
Chief Radiologist or designees are responsible Licentiates who can use and supervise fluoro Radiology supervisor and Operator Fluoro supervisor and Operator Techs with Fluoro permits can only use fluoro when supervised by above
88
Specific requirements
Establish a fluoro procedures manual Annual review of manual Assure that techs don’t practice medicine Observe tech performance Assure techs are offered training Assure equipment monitoring is adhered to
89
Personnel Protection Operator is adequately protected from scatter
Individuals in the room need to wear aprons and film badges Use protective devices as applicable
90
Protection Devices SCREEN AND GOGGLES CURTAIN THYROID
91
Restrictions Techs can only use fluoro equip. under the supervision of a supervisor operator Techs cannot interpret films Techs cannot use a title implying the right to practice medicine
92
Notification Requirements
Immediate notification (prompt phone call and timely letter) Total dose 25 rems Eye dose 75 rems Skin or extremity dose of 250 rems 24 hour notification ( call within 24 hours and a letter follow-up) Total dose 5 rems Eye dose 15 rems Skin or extremity 50 rems
93
Notes on Chapter 9 Supervision of Techs w/Fluoro permits
Clear the room of unnecessary personnel Collimate Use shields Use correct technical factors Position the patient correctly Avoid patient motion
94
Direct supervision Indirect supervision Use equipment only as trained
Spot filming and video taping
95
Notes on Chapter 10 Health Effects of Low Level Radiation dose
Somatic dose indicators Injuries to superficial tissue Induction of cancer Cataracts, fertility issues, life-span shortening Injuries to developing fetus Based on dose at specific locations or points
96
Genetic dose indicators
High marrow dose exams BE, UGI, abdominal angio Genetic dose indicators 50 rads temp male sterility 30 rads temp female sterility Genetically significiant dose Number of future kids X-ray exam rate Mean gonadal dose/exam
97
Notes on Chapter 11 Biological Effects and Significance of Dose
Effects appear to follow a linear non-threshold dose curve Dose rate to tissue Total dose Type of cell exposed
98
Radiobiological injury
Cellular amplification Gross cellular effects MC effect is the cessation of cell division Latent Period Short term, weeks or less Immediate or early effects Long term, years or longer Delayed or late effects
99
Dose relationship curves
Non-threshold linear Threshold Non-linear Regulations are based on non-threshold linear curves
100
Variations in Cell Sensitivity
Bergonie and Tribondeau Number of undifferentiated cells Degree of mitotic activity Duration of active proliferation Radiation induced mitotic delay is usually reversable
101
Cell Sensitivity Lymphocytes or white blood cells RBCs Epithelial
Endothelial Connective tissue cells Bone Nerve Brain Muscle
102
Short Term Effects 25 rads or less demonstrate no effects
103
Long Term Effects No specific effect associated with radiation exposure Somatic damage Increased incidence of cancer Embryological effects Cataracts Life span shortening Genetic mutations
104
Carcinogenic Effects Human evidence Early radiologists and dentists
Radium dial painters Uranium miners Survivors of Hiroshima and Nagasaki
105
Radiation Induced Cancers
Female breast Thyroid Hemopoitic tissue Lungs GI tract Bones
106
Embryological Effects
As little as 10 rads demonstrates effects in animal models 50 rads can cause spontaneous abortion
107
Notes on Chapter 12 Personnel Radiation Protection
ALARA Basis for radiation protection requirements Stochastic effects Probability of an event occurring, ie cancer Non-stochastic effects (deterministic) Severity of the effects varies with exposure
108
Operator Exposure Distance Apparel Aprons, 0.25 mm Pb 97% effective
Should be placed on hangers when not in use Aprons cover 80% of the bone marrow
109
Notes on Chapter 13 Personnel Monitoring
Record exposure Measure accumulated exposure Indicate type of exposure Provide a record of exposure
110
Types Film badge TLD Others 10 mrad to 700 rads +/- 25% accuracy
Lithium fluoride +/- 9%, cannot be reread Others Pocket dosimeter Audible device
111
Maximum Permissible Dose For adults over 18 y/o
Whole body - head, trunk, arms above the elbow, and legs above the knee 5 Rem Skin and extremities 50 rem Lens 15 Rem Occupational dose for people under 18 y/o 10% of adult dose
112
Occupational Exposure Limit
Whole Body – 5rem/year Extremities – 50rem/year Eye – 15rem/year Pregnant workers – 0.5rem/gestation period General public Limited to 0.1 rem/year (Addition to the background radiation) 0.002 Rem (2 mrem) per hour
113
Who must be monitored? Persons in high radiation area
0.1 Rem per hour at 30 cm Fluoro rooms Persons operating mobile x-ray equipment Radiation Area 0.005 Rem per hour at 30 cm
114
Typical Exposures mrem/yr X-Ray Tech Pain Mgnt. MPD Cardio. 5000 4000
3000 2000 1000 X-Ray Tech Pain Mgnt. MPD Cardio.
Similar presentations
© 2025 SlidePlayer.com. Inc.
All rights reserved.