Download presentation
1
Cardiac Implantable Devices
Nursing Care: The Basics and Beyond
2
Welcome! Terri Rhodes, RN, BSN Laura Hess, RN, BSN
Clinical Level III, CEP Lab Nurse Laura Hess, RN, BSN Clinical Level II, CEP Lab Nurse Please feel free to ask questions during the presentation!
3
Objectives: Examine device terminology
Examine the components, functions and indications for a pacemaker Inventory the components, indications and functions of an internal cardiac defibrillator (ICD) Compare the pacing modes using NBG pacing code system Assess patient needs preoperatively Manage patient postoperatively Analyze rhythm strips for appropriate pacemaker and ICD functioning
4
Outline 1. Welcome and general information 2. Pacemakers 3. ICD’s
4. NBG codes 5. Biventricular Pacing 6. Nursing Considerations 7. Pacemaker Practice Strips
5
Normal Conduction System
6
A Brief History of Implantable Devices
First human implant Dr. Senning in Stockholm, only lasted 3 hours 1960- First clinically successful human implant Dr’s Chardack and Gage in the US William Greatbatch, engineer 1965- First VVI implanted 1972- Partially programmable 1977-Multiprogrammable 1981- Dual chamber multi-programmable
7
Along Came ICD’s… 1980 - First human implant Thoracotomy
Epicardial patch & lead Large device placed in abdomen Not programmable; i.e. only one setting Second generation ICD Transvenous electrode Bradycardia & anti-tachycardia pacing Fifth generation Dual-chamber rate responsive pacing Improved recognition of SVT The Next Generation Remote interrogation CHF Management S-ICD- subcutaneous ICD
8
General “Device” Terms to Understand
Sense Fire Capture
9
Sense Sense: the ability of the device to recognize the presence or absence of an innate “p” wave or “qrs” complex
10
Fire Fire: the device has sensed a missed “p” wave or “qrs” complex, and has sent energy down the pacing wire to the tissue
11
Capture Capture: the energy has contracted the myocardial tissue, and resulted in a “p” wave or “qrs” complex on skin leads
12
Device Terms Continued…
Failure to Capture: A spike is noted on strip, but is not followed by appropriate “p” or “qrs” wave form Failure to Sense Spike (energy) is missing during absence of “p” or “qrs” Spike noted at inappropriate times R on T Occurs when device fails to sense, and delivers energy during vulnerable T wave - or – if programmed at VOO/AOO, the pacemaker delivers the energy in spite of intrinsic activity and paces on the t-wave. Failure to Fire Device does not send energy (pacer spike) when indicated ***If you notice any of these, check your patient, check pulse and notify physician***
13
What Do You Need To Have a Paced Beat?
Atrial Paced Beat: “a” pacing spike P wave immediately following pacer spike Ventricular Paced Beat: “v” pacing spike QRS immediately follows pacing spike
15
Examples of Paced “a”, Paced “v”, and Both
16
Pacemakers
17
What is a pacemaker? A internal device that regulates electrical impulses through the heart. Sense Fire Capture Single Chamber, Dual Chamber and Bi-Ventricular
18
Pacemaker Components Pulse generator- battery which provides the energy. Controls the rate, output, and sensitivity. The “Can” Leads- carries the impulse to the heart tissue Atrial Right Ventricle Left Ventricle Coronary Sinus
20
Indications for pacemakers
Symptomatic 2nd degree, Mobitz Type II heart block Complete heart block (3rd degree) Asystole Symptomatic bradycardia Sinus node dysfunction Carotid sinus syndrome and hypersensitivity An exaggerated response to carotid sinus baroreceptor stimulation. Sometimes even mild stimulation in the neck region causes a marked decrease in heart rate, blood pressure, and causes syncope.
21
Other Indications S/P Alcohol Septal Ablation
Hypertrophic Obstructive Cardiomyopathy (HOCM) S/P Alcohol Septal Ablation Congestive heart failure (CHF) Biventricular pacing
22
Magnet Placement for a Pacemaker
Temporarily changes the mode of pacing to asynchronous (VOO, DOO) while magnet is in place. Paces regardless of rhythm This is programmable feature of the device; NOT ONE SIZE FITS ALL
23
Break???
24
Intracardiac Cardioverter Defibrillators or ICD’s
25
What is an ICD? An internal device that can regulate electrical impulses through the heart, but its main function is to detect and terminate tachy arrhythmias. Defibrillation Override pacing Cardioversion Pacemaker Functions (Single/Dual/BiV)
26
Components of an ICD Pulse generator- battery which provides the energy. Detects tachy arrhythmias and delivers defibrillation energy when indicated. Controls the rate, output, and sensitivity of the pacemaker function. The “Can” Leads- carries the impulse to the heart tissue Right Ventricle Endo Coil – High output lead Atrium Pacemaker lead Left Ventricle Placed via the Coronary Sinus when placed in EP lab, and epicardial when placed in OR
27
Unipolar ICD
28
Indications for ICDs Secondary prevention (already had event)
Sudden Cardiac Death; NSVT, Sustained VT, V-fib arrest Inducible VT (EP testing) Primary prevention (trying to treat FIRST event) Cardiomyopathy (SCD-HeFT) At risk for sudden cardiac death Unknown etiology Long QT Brugada Syndrome (Na channel abnormality resulting in RBBB with J point elevation and concave ST elevation) Cardiac Sarcoid
29
And the Latest… S-ICD The S-ICD System is intended to provide defibrillation therapy for the treatment of life-threatening ventricular tachyarrhythmias in patients who do not have: *symptomatic bradycardia *incessant VT *spontaneous, frequently recurring VT that is reliably terminated with anti-tachycardia pacing
30
Which one do you want? Traditional ICD S-ICD
*Provides effective defibrillation *Provides effective defib for for ventricular arrhythmias ventricular arrhythmias *Provides brady pacing *No risk of vascular injury *Provides ATP pacing *Low risk of systemic injury *Provides atrial diagnostics *Preserves venous access *Familiarity of implant technique *Avoids risk of endovascular lead extraction
31
Magnet Placement for an ICD
Suspends tachycardia detection while the magnet is in place Pacing parameters remain unchanged This is a programmable feature of the ICD, and may be different
32
Caution! Place magnet on device ONLY under guidance or supervision from a physician or Electrophysiology Department nurse. Examples of when placing magnet is appropriate: ICD “ shocking” at inappropriate times During OR procedures requiring cautery. Stat pads must be placed on patient. During a code situation when you want to take ‘control of the shocking’
33
Special Considerations for Pt’s with ICD’s
If ICD discharges? 1. Check your pt: Think BLS/ACLS! ABC’s, is pt. responsive, what rhythm are they in? Take appropriate action if pt. is not stable 2. If pt. is stable notify EP department During a CODE? DO NOT place STAT pads directly over device UCH policy: Place external defibrillator pads 4-6 inches away from the device laterally if possible. Pt. is going for another OR procedure Notify Anesthesia that pt. has device, tell them the company and they will notify the EP department
34
Break?
35
NBG Codes Generic code created for NASPE and BPEG. (NASPE is the North American Society of Pacing and Electrophysiology. BPEG is the British Pacing and Electrophysiology Group.) Pacemaker programming codes that identifies how the pacemaker is programmed to function.
36
NBG Codes: Programming the pacemaker
I- What chamber do you want to pace? II- What chamber do you want to sense? III-What do you want to do with the sensed information? Inhibit pacing or trigger pacing? Tracking the Atrial activity IV-Do you want to increase the rate with the patient’s activity? Generic code created for NASPE and BPEG. NASPE is the North American Society of Pacing and Electrophysiology. BPEG is the British Pacing and Electrophysiology Group.
37
NBG Code Review P: Simple programmable V: Ventricle V: Ventricle
II III IV Chamber Chamber Response Programmable Paced Sensed to Sensing Functions/Rate Modulation P: Simple programmable V: Ventricle V: Ventricle T: Triggered M: Multi- programmable A: Atrium A: Atrium I: Inhibited D: Dual (A+V) D: Dual (A+V) D: Dual (T+I) C: Communicating O: None O: None O: None R: Rate modulating S: Single (A or V) S: Single (A or V) O: None
38
The NBG pacing code I II III IV Position Category Letters Used
Manufac- turer’s Designation Only I II III Chamber(s) Paced Sensed Response to Sensing Programmability, rate modulation O-None R-Rate modulation A-Atrium V-Ventricle D-Dual (A+V) S- Single (A or V) T-Triggered I-Inhibited (T+I) IV
39
The NBG pacing code I II III IV Position Category Letters Used
Manufac- turer’s Designation Only I II III Chamber(s) Paced Sensed Response to Sensing Programmability, rate modulation O-None R-Rate modulation A-Atrium V-Ventricle D-Dual (A+V) S- Single (A or V) T-Triggered I-Inhibited (T+I) IV
40
The NBG pacing code I II III IV Position Category Letters Used
Manufac- turer’s Designation Only I II III Chamber(s) Paced Sensed Response to Sensing Programmability, rate modulation O-None R-Rate modulation A-Atrium V-Ventricle D-Dual (A+V) S- Single (A or V) T-Triggered I-Inhibited (T+I) IV
41
The NBG pacing code I II III IV Position Category Letters Used
Manufac- turer’s Designation Only I II III Chamber(s) Paced Sensed Response to Sensing Programmability, rate modulation O-None R-Rate modulation A-Atrium V-Ventricle D-Dual (A+V) S- Single (A or V) T-Triggered I-Inhibited (T+I) IV
42
How Do We Use The NBG Language?
Single Chamber Pacing How Do We Use The NBG Language?
43
* VOO Ventricular pacing No sensing
Ventricular asynchronous pacing at lower programmed pacing rate Used for: surgical procedures with cautery Ventricular lead *
44
* VVI I Ventricular pacing Ventricular sensing
Sensed intrinsic QRS inhibits ventricular pacing Used if patient is in A-fib, do not want to tract the atrial rate I Ventricular lead *
45
* AOO Atrial pacing No sensing
Atrial asynchronous pacing at lower programmed pacing rate Atrial lead *
46
* AAI Atrial pacing Atrial sensing
Intrinsic P wave inhibits atrial pacing Atrial lead * Indications: Sinus Node Dysfunction
47
Tracking Mode: Both triggers and inhibits pacing
Dual Chamber Pacing Tracking Mode: Both triggers and inhibits pacing
48
Benefits of Dual Chamber Pacing
Provides AV synchrony Lower incidence of atrial fibrillation Lower risk of systemic embolism and stroke Lower incidence of new congestive heart failure Lower mortality and higher survival rates Studies have been done that demonstrate the differences in outcome, hemodynamic improvement, and quality of life assessment by using AV synchronous, or "atrial-based," pacing modes instead of VVI/R. Some of the benefits of using an atrial-based pacing mode include: AV synchrony–Clinical benefits such as increased cardiac output, augmentation of ventricular filling (especially important for the majority of the pacing population with LVD and reduced compliance from effects of aging). Providing AV synchrony minimizes valvular regurgitation, and preserves atrial electrical stability. In the Framingham Study, the development of chronic AF was associated with a doubling of overall mortality and of mortality from cardiovascular disease (Kannel, 1982) The following emphasize the importance of preventing atrial fibrillation: Patients with AF unrelated to rheumatic or prosthetic valvular disease have a risk of ischemic stroke about five times higher than those with normal sinus rhythm. AF is associated with over 75,000 cases of stroke per year. See bibliography for listing of studies cited.
49
Atrial undersensing only (capturing ok)
49
50
DDD * * I Pacing in both the atrium and ventricle
Sensing in both the atrium and ventricle Atrial lead * Intrinsic P wave and intrinsic QRS can inhibit pacing Ventricular Lead I * Intrinsic P Wave can “trigger” a paced QRS Maintain AV synchronization
51
DDD pacing Dual-chamber pacing capable of pacing and sensing in both the atrial and ventricular chambers of the heart 4 distinct patterns can be observed with DDD pacing
52
DDD pacing Sensing in both the atrium and the ventricle (inhibiting in both the atrium and the ventricle)
53
DDD pacing Pacing in the atrium with sensing (inhibition of pacing) in the ventricle
54
DDD pacing Sensing in the atrium (inhibition of atrial pacing) and pacing in the ventricle Also known as “P wave tracking”
55
DDD pacing Atrial pacing and ventricular pacing (no inhibition of pacing)
56
DDD mode May resemble other modes of pacing
Will strive to maintain AV synchrony with variable atrial rates and AV conduction
57
Dual Chamber Timing Parameters
Lower rate Upper rate intervals Dual-chamber pacing requires attention to these parameters: Lower rate AV and V-A intervals Upper rates Refractory periods Blanking periods
58
Lower Rate The lowest rate the pacemaker will pace the atrium in the absence of intrinsic atrial events Lower Rate Interval In order to provide optimal hemodynamic benefit to the patient, dual-chamber pacemakers strive to mimic the normal heart rhythm. In dual-chamber pacemakers, the lower rate is the rate at which the pacemaker will pace the atrium in the absence of intrinsic atrial activity. Similar to single-chamber timing, the lower rate can be converted to a lower rate interval (A-A interval), or the longest period of time allowed between atrial events. AP AP VP VP DDD 60 / 120 New Slide
59
DDDR 60 / 100 (upper tracking rate)
The maximum rate the ventricle can be paced in response to sensed atrial events { Lower Rate Interval Upper Tracking Rate Limit SAV VA SAV VA The sequence of an atrial intrinsic event being sensed, starting an SAV interval, timing out the SAV interval, and pacing in the ventricle can be referred to as "tracking." If the atrial rate begins to increase and continues to increase, is it desirable to let the ventricle "track" to extremely high rates? No. It is desirable to limit the rate at which the ventricle can pace in the presence of high atrial rates. This limit is called the upper tracking rate. AS VP DDDR 60 / 100 (upper tracking rate) Sinus rate: 100 bpm New Slide
60
Rate responsiveness/ adaptive-rate pacing
The 4th Letter in the NBG Code
61
Rate responsiveness/adaptive-rate pacing
Rate response attempts to mimic the sinus node by increasing heart rate in response to increasing metabolic demand
62
Rate responsiveness/adaptive-rate pacing
Sensor(s) detect changes in physiologic needs and increase the pacing rate accordingly
63
Rate responsiveness/adaptive-rate pacing
The sensor detects changes by: Sensing motion (crystal or accelerometer) Sensing changes in intrathoracic impedance, e.g., minute ventilation
64
Example of Dual-Chamber Rate-Responsive pacing
DDDR pacing Example of Dual-Chamber Rate-Responsive pacing
65
Biventricular PPM or ICD
A Brief Overview of What It Means To BiV Pace
66
Three lead system: Right atrial Right ventricular Left ventricular
Biventricular pacing Three lead system: Right atrial Right ventricular Left ventricular
67
Biventricular pacing
68
Cardiac Resynchronization Therapy (CRT)
Patient Indications Bi-Ventricular ICD Moderate to severe HF (NYHA Class III/IV) patients Symptomatic despite optimal, medical therapy QRS 130 msec LVEF 35%
69
Biventricular pacing Also known as cardiac resynchronization therapy, keeps the right and left ventricles pumping together by sending small electrical impulses to the heart muscle coordinating their contractions. The heart is able to fill and pump blood more effectively. This along with medical therapy, helps to improve heart failure symptoms. Improves quality of life in many.
70
Biventricular pacing Achieved by:
Inhibiting intrinsic ventricular rhythm Ensure pacing in RV and LV Short A-V delays to promote pacing in the ventricle
71
Break?
72
When Devices Go Bad!!!!
73
Complications of Device Implantation:
Pocket hematoma Pocket infection Pneumothorax Cardiac perforation Cardiac tamponade Vascular damage Lead dislodgement Lead fracture Lead infection Inappropriate shocks
74
Laser Lead Extraction Program
Implemented at UCH in 2008 by Chancey Weaver RN and Dr. Michelle Khoo M.D. First laser lead extraction in January 2009 ~30 leads extracted/year Reasons for a lead extraction: Fractured Leads Infected Lead(s) Non-functional leads/too many leads Regaining venous access
75
Unexplained Dents!
76
Device Erosion
77
Lead Fracture
78
Intraprocedure
79
Extracted Lead
80
Extracted Generator and Lead
81
Nursing Considerations
Preoperative ICD Placement and Postoperative Care
82
Preoperative Left/right arm IV Reinforce patient and family education
EP department performs education prior to and after procedure, any further questions, please call the EP lab NPO Surgical site Pre-op medications Antibiotics Blood work (WBC, Platelets, INR, Basic) Anesthesia in the procedure Restrictions after procedure
83
Postoperative Vital signs
Changes may indicate pericardial effusion or pneumothorax Type of device and settings ECG interpretation and documentation, as per unit guidelines Activity HOB <30 degrees for the first 4 hours Antibiotics Incision site X-ray within 1 hour of arriving back in room and X-ray in AM as well **Pt. placed in sling for 24 hours to allow leads to adhere to tissue**
84
Documentation According to hospital policy:
University of Colorado Hospital Call report to telemetry: Include device manufacturer and model number, mode (VVI, DDD, etc.), and lower and upper programmed rates (should be given in report). Place in computerized documentation: Device manufacturer, mode, rate, rate cut off, therapies, and date of implant. If the device fires, document any therapies of the device including the precipitating dysrhythmia and outcome in your charting. Include ECG strips, if available, documenting the dysrhythmia, the delivery of the therapy via the ICD and the resultant rhythm and the patient response.
85
Strip Documentation According to hospital policy, and individual unit guidelines. Minimal information includes “running” a strip every 12 hours or with a change in rhythm Documentation: date, time, patient's name, medical record #, heart rate, PR, QRS, and QT intervals, and rhythm analysis
86
Patient and Family Education
Wound care S/S of infection No submersion under water for 3 weeks No direct water spray (shower spray) for 1 week Coughing and deep breathing Activity All information in Post Op packet NO lifting arm above shoulder for 6 weeks Follow-up appointment Remote interrogations Electromagnetic interference Identification card
87
Patients Admitted With a PPM or an ICD
88
Patients admitted with a PPM/ICD
Ask patient for device information, i.e. registration card EP does not need to be consulted if a patient is admitted for a non-device related problem and the device appears to be working appropriately. MRI not recommended (except new Medtronic PPM) Pre-op/Post-op patients may require device programming changes ICD- tachy therapies off, or may fire during cautery PPM- reprogram to VOO, or may fail to pace appropriately
89
Pacemaker Practice Strips
90
What You Need to Document
Underlying rhythm? Is it “a” paced, “v” paced or both Is the device doing what it is programmed to do?
91
Troubleshooting Failure to: Sense Fire Capture
92
How to interpret a “paced” strip: One method of many…
Is intrinsic activity present? Are pacing spikes present: “A”, “V”, or both? Is 1:1 capture present? Is intrinsic activity sensed appropriately? Over sensing- sensing of an inappropriate single Leads to underpacing Under sensing- failure to sense intrinsic cardiac signal Results in overpacing What is the heart rate? What is the programmed pacing rate? Compliments of Northwestern Memorial Hospital, October 2002
93
Is This Normal Device Operation?
Answer: Yes. Device is programmed to VVI; rate 60.
94
Is This Normal Device Operation?
Answer: Yes. This is DDD Pacing. Programmed parameters are: DDD, Lower Rate 60, Upper Rate 120.
95
What Device Operation is This?
Answer: Ventricular Safety Pacing, coincident with a PVC. Programmed parameters: DDD Lower rate 60 ppm Upper tracking rate 120 ppm PAV 180 ms SAV 150 ms PVARP 310 ms VSP “on”
96
Is This Normal Device Operation?
Answer: Yes. The device is programmed to the DDI mode–hence, no atrial tracking. Programmed parameters: DDI; 60; PAV 150.
97
Is This Normal Device Operation?
Answer: Yes. This ECG is an example of atrial synchronous pacing in the VDD mode in the first five complexes, followed by ventricular pacing as the atrial rate drops below the lower rate in the last two complexes. Programmed parameters: VDD; lower rate 60; upper rate 120.
99
What is missing? Lack of V capture 99
102
Atrial pacemaker
104
DOO pacing mode 104
107
Atrial undersensing only (capturing ok)
107
108
Thank You Cardiac Electrophysiology William H. Sauer, MD
Dena Keilman, RN Kari Jackson, RN Noelle Hernandez, RN Amanda Lange, RN Heidi Huber, RN Terri Rhodes, RN Dan Sullivan, RN Claire Rutherford, RN Matt Upton, RN Laura Hess, RN Diane Ridgway, RN Ann Czyz. RN William H. Sauer, MD Duy Nguyen, MD Paul Varosy, MD Ryan Aleong, MD Joe Schulller, MD Wendy Tzou, MD Christine Tompkins, MD David Katz, MD Cathy Kenny, ANP
109
References Burke M, et al. Safety and Efficacy of a Subcutaneous Implantable-Defibrillator (S-ICD System US IDE Study). Late-Breaking Abstract Session. HRS 2012.
Similar presentations
© 2025 SlidePlayer.com. Inc.
All rights reserved.