Operator Course. Trace Sampling Equipment Overview Analysis Calibration Course Overview.

Similar presentations


Presentation on theme: "Operator Course. Trace Sampling Equipment Overview Analysis Calibration Course Overview."— Presentation transcript:

1 Operator Course

2 Trace Sampling Equipment Overview Analysis Calibration Course Overview

3 Trace Traces are microscopic amounts of particles and vapors. Narcotic substances are organic and constantly emit ‘traces’. The VaporTracer detects traces down to nanogram levels.

4 The VT2 is just one more tool to help you improve your ability to detect narcotic traces. Use it along with your X-ray, metal detectors, canine units, physical inspections, and current security measures. Trace

5 Substances Detected READY 10 100 1k 10k COC l _ _ HER l _ _ THC l _ _ MET l _ _ EDM l _ _ MDM l _ _ P+

6 Traces cling to anything they touch. Traces transfer from person to person, person to object, object to object and airborne. Trace How Do Traces Get Transferred?

7 Cargo, mail, and packages Conveyance access - Security access points, border crossings, checkpoints Baggage Facilities protection - government buildings, correctional institutions Trace Applications

8 Gun Radio lined with narcotics Pepper spray Laptop with concealed narcotics Stun gun Narcotics traces on a person’s hands Trace Which of the following can be detected with the VT2?

9 What are some of the traces the VT2 can detect? What are some things the VT2 cannot detect? Are narcotics traces illegal? How are traces transferred? Trace

10 ·Particle – Most reliable method. Uses a different collection procedure. ·Vapor – Certain substances give off less vapor than others. Temperature is a factor for collecting vapor. Sample vapor release points. Sampling

11 l Sampling Traps l Sampling Wand l Gloves l Sampling Table or Area l Alcohol (isopropyl 70% or greater) Sampling Materials and Equipment for Particle Sampling

12 Vapor Plate White Spacer Locking Ring Check Sample Sampling Materials and Equipment for Vapor Sampling

13 l Person whose item you are sampling should be present. l Sample one item at a time. l Use firm pressure. l Sample in one direction, one time. l Sample most touched areas last. Sampling Particle Sampling Fundamentals

14 People: l Hands l Belts l Pockets Sampling Particle Sampling Fundamentals

15 Vehicles: l Door handle l Gas cap l Steering wheel l Knobs l Seat Belt l Luggage l Wherever there may be a fingerprint. l Avoid sampling engine parts. Sampling Particle Sampling Fundamentals

16 Luggage/Bags/Briefcases: l Handles l Straps l Zipper tabs l Latches l Interior seams and linings l Commonly touched areas Sampling Particle Sampling Fundamentals

17 Electronic Devices: l Vents l Seams l Battery compartments l Disk drives l Edges l Latches AVOID SAMPLING: l Monitors l Keyboards l Camera lenses Sampling Particle Sampling Fundamentals

18 l Sample vapor release points. l Get as close as possible to the vapor release point without touching it. l Temperature affects vapor samples. l Substances emit different levels of vapor. Sampling Vapor Sampling Fundamentals

19 What are some particle sampling fundamentals? Vapor? What areas could we sample on a vehicle? Electronic item? Luggage? What areas do we avoid sampling? Why? When do we discard a sample trap? Sampling

20 Desorber Start button Navigation buttons Display screen Power switch Power receptacle Battery Equipment Overview

21 READY 10 100 1k 10k COC l _ _ HER l _ _ THC l _ _ MET l _ _ EDM l _ _ MDM l _ _ P+ Equipment Overview

22 MENU 1 1: 19 SELECT SUBSTANCE BEEPER OFF LIGHT ON METHOD VAP CAL EXT INT ON OFF PART Equipment Overview

23 STATUS BAT DETACHED HTR l 200 FLO OK PMP OK DATE 07-13-2001 TIME 01 : 31 PM Equipment Overview

24 Unit is ready 30 minutes after a cold start. Unit should be powered-on 24/7. Default 5 minute warm-up time. Default 15 minute warm-up time when changing the desorber. Power down only as directed by pressing the switch to the center position Equipment Overview

25 Where is the sampling trap placed? Which button takes you to other screens? How long is a cold-start warm up? When do you shut the unit down? Equipment Overview

26 Place the sample trap into the desorber. Press the Start button to begin analysis. When in Vapor mode, hold the Start button for as long as necessary. Analysis begins when the Start button is released. Maintain control of the item you are sampling. Observe Ready or Alarm on the display. Remove the sampling trap. If Ready, reuse the trap. If Alarm, discard trap and gloves. Analysis Procedures

27 Resolve the alarm in accordance with current SOP. Examples: Contact your supervisor. Maintain control of item. Let the owner know that further samples may be necessary. Decontaminate the work area. Analysis Alarm Response ALARM S-CLR E-SV 10 100 1k 10k COC l 4.13 HER l _ _ THC l _ _ MET l _ _ EDM l _ _ MDM l _ _

28 Save the alarm by pressing ENTER. Clear the unit after saving by pressing START. Decontaminate the work area with alcohol only The unit must be cleared before additional samples can be taken. Analysis Saving Alarms

29 Make sure the alarm was saved. Press the START button when you see Ready on the display. Insert a new sampling trap and press START. If the new trap alarms, inform the supervisor. In most cases the unit will clear after an alarm. Otherwise, maintenance may be necessary. Analysis Clearing the Unit

30 What is the indication of an alarm? No alarm? What button do you press to save an alarm? What button do you press to clear the unit? What do you do with the trap after an alarm? Analysis

31 Calibration introduces a known substance into the detector for identification. After the VaporTracer 2 identifies the calibrant, the software will adjust the positions of all substances. The VaporTracer 2 must be calibrated for accurate detection. Why Calibrate? Calibration

32 After powering up After changing modes At the beginning of each shift When the unit is moved At higher elevations When atmospheric conditions change (humidity, pressure) When Do You Calibrate? Calibration

33 Press to begin calibration. Procedure Calibration CAL REQUIRED MENU TO CONTINUE COC l _ _ HER l _ _ THC l _ _ MET l _ _ EDM l _ _ MDM l _ _ P+

34 ,. Use to highlight CAL EXT, then press. MENU 1 SELECT SUBSTANCE BEEPER OFF LIGHT ON METHOD VAP INT ON OFF CAL EXT PART Procedure Calibration

35 INSERT CAL TRAP TIME STD LAST 8.469 8.534 -MENU- TO EXIT -START- TO CAL Insert the calibration trap into the desorber and then press. Procedure Calibration

36 Press to accept the highlighted peak. STD 8.46 LAST 8.53 HEIGHT TIME -ENTER- TO ACCEPT GET EXPANDED LIST 5892 8.454 Procedure Calibration

37 If the proper peak is not highlighted, use to select it. The correct peak is the peak closest to the STD peak (usually between 6 and 8 ms). STD 8.46 LAST 8.53 HEIGHT TIME -ENTER- TO ACCEPT GET EXPANDED LIST 6328 8.459 234 3.289 567 9.583 Procedure Calibration

38 CALIBRATION DONE -ENTER- TO CLEAR Procedure Calibration READY 10 100 1k 10k COC l _ _ HER l _ _ THC l _ _ MET l _ _ EDM l _ _ MDM l _ _ P+

39 Why is calibration necessary? Which menu is used to access calibration? If a calibration peak is not highlighted, which arrow do you use to select the proper peak? When do you calibrate? Calibration