Download presentation
Published byIsaac Sullivan Modified over 9 years ago
1
CENTRAL GAS SYSTEMS Without reference, identify principles relating to central gas systems as they pertain to facility and equipment interface with at least 70 percent accuracy.
2
CENTRAL GAS SYSTEMS Primary objective of suppliers of medical gas distribution systems is to provide Creative technical service Quality products that keep pace with changing needs Central Gas Systems Supply medical gases in life support situations It is essential that the reliability and integrity of the system be the responsibility of all professions Manufacturing, Design, Installation, Testing and the Operation
3
CENTRAL GAS SYSTEMS Medical Gas Systems
The main features of medical gas systems are the types of gases and the means of their delivery Types of medical gases Oxygen Oxygen is stored as a compressed gas at room temperature or refrigerated as a liquid Primarily used for respiratory therapy and anesthesia Constant supply and immediate availability through out medical facility is essential Small hospitals often store oxygen in banks of cylinders connected by a manifold to provide a “central gas supply” Should be designed to provide 60 psig at the outlets Most anesthesia machines accommodate one or two cylinders to be used in emergencies
4
CENTRAL GAS SYSTEMS
5
CENTRAL GAS SYSTEMS Nitrous oxide Primarily used for OR area
Is almost always stored by hospitals in large high- pressure cylinders connected by a manifold Should be designed to provide 60 psig at the outlets Bulk liquid storage of nitrous oxide is economical only in large institutions Most anesthesia machines accommodate one or two cylinders to be used in emergencies
6
CENTRAL GAS SYSTEMS Medical air (compressed air)
Primarily used for respiratory therapy in areas like nurseries, emergency rooms, intensive care areas and OR’s
7
CENTRAL GAS SYSTEMS The use of air is becoming more frequent in anesthesiology due to the hazards of high concentrations of oxygen Dehumidified but non sterile air is provided to the hospital pipeline by compression pumps Compression pumps should be designed to provide 60 psig at the outlets Must be free from oil and moisture The compressor will not add contaminants in the form of particulate matter, odor or other gases
8
CENTRAL GAS SYSTEMS The medical air compressors intake will be
Located outdoors Above roof level A minimum distance of 10 feet from any door, window, exhaust, other intake, or opening in the building A minimum distance of 20 feet above the ground The intake will be turned down and screened or otherwise be protected against the entry of vermin or water
9
CENTRAL GAS SYSTEMS Compressor alternation
Compressors will automatically alternate in operation to allow division of operating time If automatic alternation is not provided, the facility will arrange a proper schedule for manual alternation A backup medical compressor will be available to automatically activate if the unit in operation is: Incapable of adequately maintaining pressure Inoperative
10
CENTRAL GAS SYSTEMS Vacuum
Compressors will be alarmed appropriately for Conditions that affect air quality Failure Vacuum Not a medical gas, but an essential part of medical gas system Widely used throughout medical facilities in Lab, patient treatment areas, surgical recovery and intensive care areas
11
CENTRAL GAS SYSTEMS Uses
To remove fluid from incisions and assist in vital post- operative drainage In active scavenging systems Used to remove excess waste gases from the patient circuit of anesthesia units Routes gases out of the operating room and towards a safe area outside of medical facility
12
CENTRAL GAS SYSTEMS Vacuum system
Normally designed to provide in Hg of negative pressure at the furthest point from central vacuum Types of systems Wet systems will remove all waste through the service inlet Dry systems will only remove air-gas through the service inlet Types of waste removed Liquid, Air-gas and/or Solids
13
CENTRAL GAS SYSTEMS Vacuum pumps
Two or more pumps will be installed that serve the system alternately or simultaneously on demand Each pump will be large enough to serve peak demand with the other pump inoperative Station inlets for vacuums will be legibly labeled “Suction” or “Vacuum”
14
CENTRAL GAS SYSTEMS Common Medical Gases Labels Air – air
Carbon dioxide – CO2 Carbon monoxide – CO Helium – He Hydrogen – H2 Nitrogen – N2 Nitrous oxide – N2O Oxygen – O2
15
CENTRAL GAS SYSTEMS Delivery of Medical Gases
Gases are delivered from their central supply source to the operating room through a piping network. Usually constructed of seamless copper tubing Internal contamination of the pipelines with dust, grease, or water must be avoided. The delivery system terminates in the operating room as Hose drops Gas columns Elaborate articulating arms
16
CENTRAL GAS SYSTEMS Medical equipment is connected to the gas system through quick coupler mechanisms Each coupler is color coded O2 – green Air – yellow N2O – blue Vacuum – white
17
CENTRAL GAS SYSTEMS Designed to fit a specific gas
PISS – pin index safety systems This system is used with small gas cylinders Accomplished by 2 metal pins on the yoke that correspond to two holes in the cylinder head There are six pin positions, each of which are specific to a particular gas Some gas mixtures have 2 pin positions assigned to differentiate different gas proportions One cylinder gasket must be used Multiple gaskets can defeat the PISS
18
CENTRAL GAS SYSTEMS
19
CENTRAL GAS SYSTEMS DISS – diameter indexing safety system
This system is used with central pipeline hoses and large gas cylinders Accomplished by using various diameter male to female connectors specific to each gas
20
CENTRAL GAS SYSTEMS Factors in Supplying Medical Gases Location
Number of outlets required Pipeline routing and size of pipes Medical gas supply source Intermediate pipeline controls Gas service outlets Critical care area and equipment System specifications
21
CENTRAL GAS SYSTEMS Medical Gas Alarms Master alarm Area alarm
Monitors the operation and condition of the source of supply, the reserves and the pressure of the main lines of all medical gas piping systems The master alarm system consists of two or more alarm panels located in two separate locations Area alarm Provided for each medical gas piping system supplying anesthetizing locations and other vital life support and critical care areas Usually located at the nurse’s station
22
CENTRAL GAS SYSTEMS Standards Purpose Safety Low infection rate
Low post operative care Assurance of the functionality of the system Confidence to medical workers Economical Avoid medical legal complications
Similar presentations
© 2024 SlidePlayer.com. Inc.
All rights reserved.