Presentation is loading. Please wait.

Presentation is loading. Please wait.

Cleaning, Packaging & Sterilization of Dental Instruments

Similar presentations


Presentation on theme: "Cleaning, Packaging & Sterilization of Dental Instruments"— Presentation transcript:

1 Cleaning, Packaging & Sterilization of Dental Instruments

2 Presented by SPSmedical
Largest sterilizer testing Lab in North America with over 50 sterilizers Develop and market sterility assurance products that offer advanced technologies Provide full day sterilization Seminars and Facility audits for compliance with standards and recommended practices Corporate member: CSA and AAMI, serving on numerous sterilization committees

3 Continuing Education Program
Following established protocols for instrument processing is an important aspect of modern dental care as it helps to minimize the patient’s risk for infection. This program is intended to provide an “overview” of current recommended CDC guidelines and is approved for 1.0 credit (category 4) by DANB (Dental Assisting National Board).

4 INSTRUMENT PROCESSING
There are 6 recommended steps for Instrument Processing: Cleaning Inspection Packaging Sterilization Storage & Delivery 6. Quality Assurance

5 CLEANING Cleaning is the first and most important step which should be
performed in a designated area, ASAP wearing PPE. Quick cleaning removes blood much easier, and can minimize instrument staining, corrosion and/or pitting.

6 CLEANING Mechanical cleaning is a safer practice for staff and is much
more effective over manual cleaning. Therefore, whenever possible, mechanically clean instruments, using warm water and a neutral pH detergent.

7 CLEANING After cleaning, thoroughly rinse
instruments with tap water and ensure all debris and detergent residue is removed. If the tap water is of poor quality, consider using treated water as a final rinse to avoid instrument staining.

8 INSPECTION Each instrument should be critically inspected after
each cleaning for residual debris or damage. Replace instruments as needed and never sterilize a “dirty” instrument.

9 INSPECTION In addition, check each instrument for proper
function and lubricate as required by the instrument manufacturer. Hinged instruments with stiff joints may be a sign of inadequate cleaning..

10 PACKAGING Instrument packaging should be done in a clean and low
contamination area, using FDA approved products: Sterilization pouches Sterilization wrappers Sterilization containers

11 PACKAGING Sterilization pouches are for packaging single instruments
and small, light weight items. Paper/plastic pouches allow you to see the contents and come with a build in adhesive strip for seal sealing. It is important to remove all excess air, prior to sealing the sterilization pouch.

12 PACKAGING For quality assurance, be sure
to include a chemical indicator or integrator inside each pouch. Per CDC guidelines, this will verify sterilization parameters were met inside as well as outside the pouch (time, temperature, and sterilant contact). Indicator Integrator

13 PACKAGING To assist sterilization and aid drying, place pouches facing each other and on edge using a divider. Pouch dividers are part of the SPSmedical product line.

14 PACKAGING Sterilization wrap is ideal for packaging dental cassettes.
For sterility maintenance, be sure to use two layers of wrap per industry standards, and the wrap mfg’s FDA clearance.

15 PACKAGING For quality assurance, use indicator tape to seal and be
sure to include a chemical indicator or integrator inside, to verify sterilization parameters were met. For staff and patient safety, consider using a latex free indicator tape.

16 STERILIZATION Heavier wraps should be placed on the lower shelf of
the sterilizer and pouches (being lighter) on the top, on edge. This enables condensate to drain out without wetting other items in the load. To maintain sterility, packages should be visibly dry before transporting to a sterile storage area

17 STERILIZATION Sterilizers are Class 2 medical
devices requiring FDA clearance. They are available in a variety of sizes with the following processes most common to dental: Steam Chemical vapor Dry heat / Rapid heat transfer

18 Chamberland autoclave built in 1880
STERILIZATION Saturated steam under pressure is one of the oldest methods used to sterilize dental instruments. The CDC recommends steam sterilization as the process of choice, because it is efficient, fast, and inexpensive. Chamberland autoclave built in 1880

19 STERILIZATION By heating distilled water under
pressure, moist heat is created and rapidly kills microorganisms. Some common steam sterilizer cycle parameters are: 250°F/121°C for 30 minutes 270°F/132°C for 10 minutes 273°F/135°C for 4 minutes Dry times are additional and can be from 20 to 40 minutes, depending on load.

20 STERILIZATION Chemical vapor sterilizers are available with a pre-set
cycle of 270°F/132°C for 20 minutes exposure. This proprietary process uses a chemical solution, instead of distilled water, that rapidly kills micro- organisms. OSHA requires a Material Safety Data Sheet on the chemical solution used for this proprietary sterilization process

21 STERILIZATION Dry heat sterilizers are common
in dentistry for instruments and materials that can withstand very high temperatures. Static air dry heat sterilizers are similar to an oven, as they utilize heating coils to transfer heat to the instruments over long exposure times. Static air dry heat sterilizers typically operate at 320°F/160°C for 1 to hours exposure time

22 STERILIZATION Forced air dry heat sterilizers are
also called “rapid heat transfer” sterilizers and they circulate hot air during the cycle reducing the time needed for sterilization. This type of dry heat sterilizer is capable of sterilizing unwrapped instruments in as little as 6 mins. Forced air dry heat sterilizers operate at a range of 375°F to 420°F depending on the manufacturer

23 Storage & Delivery Sterile items should be stored
in a manner that reduces the potential for contamination. The shelf-life of sterile items is event related and depends on the quality of the packaging material, storage conditions and amount of handling.

24 ____________ Storage & Delivery Sterile packages should always
be handled with care. Avoid dragging, crushing, bending, compressing or puncturing, as this can compromise sterility. Be sure to inspect sterile packages before distributing. Do not use any package that is damaged, wet or opened. Sterilant Microbes ____________

25 Storage & Delivery Special Note: Opening sterile packages
at point of use, increases patients’ confidence.

26 Quality Assurance Sterility assurance of processed instruments should be routinely verified using three (3) types of indicators: 1) Physical 2) Chemical 3) Biological

27 Quality Assurance Physical indicators are the
time, temperature and pressure gauges built into sterilizers. These readings should be recorded for every cycle and verified prior to unloading the sterilizer. Hospital sterilizers are required to have a chart or printout, whereas this is optional for sterilizers located in private offices or clinics.

28 Quality Assurance 2) Chemical indicators should
be on the outside and inside of all packages to visibly show they have been processed. Class 5 chemical indicators FDA cleared as equal in performance to a BI are called “integrators”. Use of steam integrators is common in hospitals an growing in dentistry.

29 Quality Assurance 3) Biological indicators provide
users the highest level of sterility assurance and contain bacterial spores available in plastic vial or paper strip format. The CDC recommends all sterilizers be spore tested at least weekly, and every load that contains an implant.

30 Quality Assurance Biological indicators are processed along with a normal load. After processing, self-contained spore vials are activated by crushing the side of the plastic vial and then incubated. Sterilizer failure is noted if the spores in the vial grow during incubation. Growth is observed by a color change in the media from purple to yellow.

31 Quality Assurance Biological indicators can also
be sent to an outside Lab for 3rd party verification. After processing the BI strip along with a normal load, the processed strip along with an unprocessed strip are sent to an independent lab for incubation. Test results are returned via mail or fax, with some services offering electronic record keeping.

32 Quality Assurance Mail-in spore testing of sterilizers
is very popular with private offices. For user convenience, SPSmedical test reports are also posted online* at

33 Quality Assurance While sterilizers can and do
mechanically fail, operator error is the leading cause of sterilizer failure, e.g. Cold start Wrong cycle Overloading Improper packaging Sterilizers that fail the spore test, should not be used until a passed test is recorded

34 SPSmedical thanks you for taking the time to review…
The six 6 recommended steps for Instrument Processing: Cleaning Inspection Packaging Sterilization Storage & Delivery 6. Quality Assurance

35 Cleaning, Packaging and Sterilization of Dental Instruments
THANK YOU! SPSmedical Supply Corp. Sterilization Products & Services 6789 W. Henrietta Road Rush, NY USA Fax: (585) Ph: (800) Website: © , SPSmedical Supply Corp. Cleaning, Packaging and Sterilization of Dental Instruments Certificate is available after viewing by calling SPSmedical or ing


Download ppt "Cleaning, Packaging & Sterilization of Dental Instruments"

Similar presentations


Ads by Google