FLASH. Free-Electron Laser in Hamburg Overview and Status of the Standard Electron Beam Diagnostics at FLASH N. Baboi, DESY for the Diagnostics Team
2
Nicoleta Baboi | Electron Beam Diagnostics at FLASH | FLASH Seminar | DESY, Hamburg | Nov. 11, 2014 | Slide 2 FLASH. Free-Electron Laser in Hamburg Contents > Overview of standard beam diagnostics in FLASH1 and FLASH2 General characteristics > Beam monitors Charge Monitors Position Monitors Size Monitors Loss Monitors Not all monitor types covered by this talk
3
Nicoleta Baboi | Electron Beam Diagnostics at FLASH | FLASH Seminar | DESY, Hamburg | Nov. 11, 2014 | Slide 3 FLASH. Free-Electron Laser in Hamburg Overview of Electron Beam Diagnostics in FLASH FLASH1FLASH2 Charge Toroids125 Dark Current Monitor1 Faraday Cups3 BPMs633 Transverse Size OTR-Screens~30 Scintillating-Screens17 Wire scanners (MDI)10 Wire scanners (Zeuthen)9 Transverse Position Button-BPMs2612 Stripline-BPMs334 Cold Cavity BPMs6 Cavity BPMs17 HOM-based monitors39 Beam Loss BLMs>70~55 Cherenkov Fibers21 Beam Halo Monitors1x4 Ionization Chambers44
4
Nicoleta Baboi | Electron Beam Diagnostics at FLASH | FLASH Seminar | DESY, Hamburg | Nov. 11, 2014 | Slide 4 FLASH. Free-Electron Laser in Hamburg General Characteristics > General Deliver bunch by bunch measurements, with few exceptions > FLASH1 Most designed for TTF2, for charges of 1nC and higher Electronics: VME-based, ADCs with 1MHz > FLASH2 Most designed for the E-XFEL Make use of synergies given by the similar time line Some have been designed specially for FLASH(2) or have a temporary solution Generally based on MTCA-technology 9 MTCA crates for diagnostics installed at FLASH2 Designed for lower charge N. Baboi, IBIC’14, THIXB1 J. Feldhaus, D. Nölle, EPAC2004, p. 262
5
Nicoleta Baboi | Electron Beam Diagnostics at FLASH | FLASH Seminar | DESY, Hamburg | Nov. 11, 2014 | Slide 5 FLASH. Free-Electron Laser in Hamburg Charge Monitors / Toroids FLASH1 > All signals go to VME and uTCA- system 3GUN and 10DBC2 have new front-end (amplifier & filter, like in FLASH2) FLASH2 > Temporary solution Low-gain front-end channel CAT cables (E-XFEL type under tests M. Werner, IBIC’14, WEPF02) Charge Monitors: Toroids W W W W + + + Front-end Adapter RTM RTM SIS8900 AMC SIS8300 AMP-200 ADC peak ADC base VME (Fl1) uTCA (Fl1&2)
6
Nicoleta Baboi | Electron Beam Diagnostics at FLASH | FLASH Seminar | DESY, Hamburg | Nov. 11, 2014 | Slide 6 FLASH. Free-Electron Laser in Hamburg ► Charge Monitors: Toroids (2) Resolution of charge monitors along FLASH(1) (~ 0.4 nC) o FLASH-type (VME) o FLASH2-type (MTCA) old amplifier O FLASH2-type with new amplifier ► Dark current monitor
7
Nicoleta Baboi | Electron Beam Diagnostics at FLASH | FLASH Seminar | DESY, Hamburg | Nov. 11, 2014 | Slide 7 FLASH. Free-Electron Laser in Hamburg Toroid Protection System (TPS) 5 MeV 150 MeV1250 MeV Bunch Compressors 450 MeV Accelerating Structures RF Stations Lasers RF Gun Soft X-ray Undulators sFLASH FEL Experiments Photon Diagnostics Beam Dump THz FLASH1 FLASH2 TPS 0 TPS 1 TPS 2 A. Hamdi, DIPAC 2007, p. 349 > Temporary solution using 3 TPS modules designed for FLASH (not yet commissioned)
8
Nicoleta Baboi | Electron Beam Diagnostics at FLASH | FLASH Seminar | DESY, Hamburg | Nov. 11, 2014 | Slide 8 FLASH. Free-Electron Laser in Hamburg Charge Monitors: Dark Current Monitor Dark current Bunch 7DBC2, 32m
9
Nicoleta Baboi | Electron Beam Diagnostics at FLASH | FLASH Seminar | DESY, Hamburg | Nov. 11, 2014 | Slide 9 FLASH. Free-Electron Laser in Hamburg > Designed by INFN > Some camera-PCs have to be reset/restarted from time to time > Recently had to readjust 3 stations in DBC2 Difficult due to missing fine adjustment 2 are adjusted only for a magnification of 1 > Emittance tool Screens at FLASH1 A. Cianchi, EPAC2004, p. 2619 23.09.2014, 13:42
10
Nicoleta Baboi | Electron Beam Diagnostics at FLASH | FLASH Seminar | DESY, Hamburg | Nov. 11, 2014 | Slide 10 FLASH. Free-Electron Laser in Hamburg Screens at FLASH2 BEAM Optics Axis > Scintillator based (LYSO:Ce), spatial suppression of COTR > Optics: Look under 45 deg Uses “Scheimpflug´s” principle to extend depth of field > 10 µm resolution required single bunch, streaked > Demonstrated at FLASH1 200µm thick LYSO Screen (ON-Axis ) 2 half 200µm thick LYSO Screens (OFF-Axis) Dot Grid Target(spot Ø.50mm) C. Wiebers, IBIC2013, WEPF03
11
Nicoleta Baboi | Electron Beam Diagnostics at FLASH | FLASH Seminar | DESY, Hamburg | Nov. 11, 2014 | Slide 11 FLASH. Free-Electron Laser in Hamburg Screens at FLASH1 (2) > Fully commissioned FLASH2 FLASH1 1FL2SEED4 1FL2SEED61FL2SEED7 1FL2SEED5
12
Nicoleta Baboi | Electron Beam Diagnostics at FLASH | FLASH Seminar | DESY, Hamburg | Nov. 11, 2014 | Slide 12 FLASH. Free-Electron Laser in Hamburg > One horizontal and one vertical fork, each with 3 wires Used also for cross-calibration of BPMs > Tool for profile scans and emittance Wire Scanners: Zeuthen Type P. Castro, DIPAC 2005, p. 205
13
Nicoleta Baboi | Electron Beam Diagnostics at FLASH | FLASH Seminar | DESY, Hamburg | Nov. 11, 2014 | Slide 13 FLASH. Free-Electron Laser in Hamburg > One fork placed at 45deg with horizontal, vertical and diagonal wire Several WS use one detector Also possible to read signals from BLMs, with some change in the beam profile Some have some readout/timing problems, now under work Jddd panels to be updated Wire Scanners: MDI/CERN Type M. Werner, DIPAC 2001, p. 139
14
Nicoleta Baboi | Electron Beam Diagnostics at FLASH | FLASH Seminar | DESY, Hamburg | Nov. 11, 2014 | Slide 14 FLASH. Free-Electron Laser in Hamburg > Button BPMs, stripline BPMs, cold cavity BPMs, button BPMs inside undulators BPMs in FLASH1 ~0.35 nC
15
Nicoleta Baboi | Electron Beam Diagnostics at FLASH | FLASH Seminar | DESY, Hamburg | Nov. 11, 2014 | Slide 15 FLASH. Free-Electron Laser in Hamburg BPMs in FLASH2: Cavity BPMs > Cavities Dipole + reference resonator 3.3 GHz, low Q (about 70) 17 installed between undulators – 10 mm pipe No individual tuning > Readout (PSI) MBU (Modular BPM Unit) with 2 RFFE for 2 cavity BPMs Up to 4 MBUs connected to 1 DAMC2 interface board Timing: decoding timing protocol via fiber in MBU > Resolution requirement 2 m rms for 0.1-1 nC and within ± 0.5 mm (1 m at XFEL, but longer cables at FLASH2) M. Stadler, IBIC’14, WEPD12 B. Keil, IBIC’14, WEPD11
16
Nicoleta Baboi | Electron Beam Diagnostics at FLASH | FLASH Seminar | DESY, Hamburg | Nov. 11, 2014 | Slide 16 FLASH. Free-Electron Laser in Hamburg > BPMs in FLASH2 Difference between expected and measured position with Gaussian fit BPMs in FLASH2: Cavity BPMs: Resolution > Test BPMs in FLASH1 0.5 m rms for 0.24 nC bunch charge and 0.5 mm bunch offset Resolution of various FLASH1-BPMs (designed for 1 nC) ~0.35 nC preliminary D. Lipka, IBIC’14, TUPF07
17
Nicoleta Baboi | Electron Beam Diagnostics at FLASH | FLASH Seminar | DESY, Hamburg | Nov. 11, 2014 | Slide 17 FLASH. Free-Electron Laser in Hamburg BPMs in FLASH2: Cavity BPMs: Charge > Charge resolution for 100 pC bunch charge < 0.2 pC rms D. Lipka, IBIC’14, TUPF07
18
Nicoleta Baboi | Electron Beam Diagnostics at FLASH | FLASH Seminar | DESY, Hamburg | Nov. 11, 2014 | Slide 18 FLASH. Free-Electron Laser in Hamburg BPMs in FLASH2: Button and Stripline BPMs > Low Charge BPMs (LCBPM) electronics Name chosen in contrast to system designed for FLASH for charges of 1nC and higher Designed for FLASH2 (and to be installed in all FLASH) Button or Strip-line BPM Delay ~100ns Broadband RF Combiner RF cable 3/8“ Length < 80m Signal of horizontal plane, delay line: ~100ns TYPICAL SIGNALS FROM BUTTON/STRIP-LINE BPM B. Lorbeer, IBIC’14, TUPF08 F. Schmidt-Föhre, IPAC’14, THPME117 S. Vilcins, IBIC’14, TUPF11
19
Nicoleta Baboi | Electron Beam Diagnostics at FLASH | FLASH Seminar | DESY, Hamburg | Nov. 11, 2014 | Slide 19 FLASH. Free-Electron Laser in Hamburg BPMs in FLASH2: Button and Stripline BPMs: LCBPM Rear Transition Module (RTM) 180mmx162mm Compatible with the MTCA.4 for physics standard RF Front-End MicroTCA crate X-Channel Y-Channel 100ns Crate used for the tests includes: management carrier hub (MCH) hard disk (HD), processor (CPU) Digitizer SIS8300 timing module x2timer RTM Analog Output Signals G=38dB Att.>60dB Peak Detector 5 BPMs/crate Front Side Back Side
20
Nicoleta Baboi | Electron Beam Diagnostics at FLASH | FLASH Seminar | DESY, Hamburg | Nov. 11, 2014 | Slide 20 FLASH. Free-Electron Laser in Hamburg BPMs in FLASH2: Button &Stripline BPMs: Operation RMS at q=36 pC, 200 pulses (incl. beam jitter) Server Panel for Firmware Adjustment BPM Expert Panel Strip-Line BPM Button BPM y = 22 m y = 39 m bunch charge0.1-1nC bunch spacing≥222ns maximum macro-pulse repetition rate25Hz beam pipe diameter40.5mm single bunch resolution50μm Averaged RMS resolution over 1000 bunches of identical train10μm Range for max. resolution+/- 3mm Operation range with reasonable signal+/-10 mm Requirements (from XFEL requirements)
21
Nicoleta Baboi | Electron Beam Diagnostics at FLASH | FLASH Seminar | DESY, Hamburg | Nov. 11, 2014 | Slide 21 FLASH. Free-Electron Laser in Hamburg HOM-based Position Monitors > Monitor dipole, beam position correlated, signals excited in the accelerating cavities Amplitude shows the beam alignment in the cavities, which influences the (transverse) beam quality Available in ACC1 (3 cavities), ACC39 and ACC2-5 Display only signal envelope
22
Nicoleta Baboi | Electron Beam Diagnostics at FLASH | FLASH Seminar | DESY, Hamburg | Nov. 11, 2014 | Slide 22 FLASH. Free-Electron Laser in Hamburg Beam Loss Monitors in FLASH1 Cannot distinguish between losses from FLASH1 and FLASH2 pulses
23
Nicoleta Baboi | Electron Beam Diagnostics at FLASH | FLASH Seminar | DESY, Hamburg | Nov. 11, 2014 | Slide 23 FLASH. Free-Electron Laser in Hamburg Beam Loss Monitors in FLASH2 > Detector Scintillator or fused silica (rod or fiber) PMT: HV voltage (~500 V) generated by Cockcroft-Walton multiplier on board mounted at PMT Scintillators and mechanics built by IHEP, Protvino (60 for FLASH2) > Electronics MTCA-based with low latency interface with MPS > Alarms algorithms Single bunch Multi-bunch Integration OR of comparator and FPGA alarms Readout Electronics Power control PMT & HV Scintillator Beam Loss Monitor e, γ,n 0 Alarm (to MPS) A. Kaukher, BIW 2012, p. 35
24
Nicoleta Baboi | Electron Beam Diagnostics at FLASH | FLASH Seminar | DESY, Hamburg | Nov. 11, 2014 | Slide 24 FLASH. Free-Electron Laser in Hamburg Beam Loss Monitors in FLASH2: Status > Calibrated single and multi-bunch alarm > Integration alarms to be calibrated
25
Nicoleta Baboi | Electron Beam Diagnostics at FLASH | FLASH Seminar | DESY, Hamburg | Nov. 11, 2014 | Slide 25 FLASH. Free-Electron Laser in Hamburg Dump Loss Monitors in FLASH1 & FLASH2: BLMs and Ionization Chambers > BLMs based on quarz-fiber + Ionization Chambers (air filled cables) along last 2m of beam line > VME/uTCA electronics beam FLASH1 N. Baboi, BIW 2010, p. 420
26
Nicoleta Baboi | Electron Beam Diagnostics at FLASH | FLASH Seminar | DESY, Hamburg | Nov. 11, 2014 | Slide 26 FLASH. Free-Electron Laser in Hamburg Dump Loss Monitors: Beam Halo Monitors (FLASH1 & FLASH2) > Installed behind dump vacuum window 8 sensors: 4 Diamond + 4 Sapphire > VME-based in FLASH1 > uTCA-based in FLASH2 Basically same electronics and firmware as FLASH2 BLMs Under commissioning FLASH1 A. Ignatenko, IPAC 2012, p. 816
27
Nicoleta Baboi | Electron Beam Diagnostics at FLASH | FLASH Seminar | DESY, Hamburg | Nov. 11, 2014 | Slide 27 FLASH. Free-Electron Laser in Hamburg Acknowledgement > Many people contributed to the work and this talk R. Neumann, D. Lipka, N. Wentowski, Z. Pisarov, M. Werner, J. Lund-Nielsen, B. Lorbeer, A. Kaukher, T. Wamsat, A. Ignatenko, F. Schmidt-Föhre, K. Honkavaara, H. Tiessen, B. Michalek, R. Zahn, B. Keil, D. Nölle, O. Hensler, G. Marinkovic, K. Knaack, J. Schwarz, J. Liebing, H. Sokolinski, G. Kube, C. Wiebers, M. Felber, R. Zahn, S. Vilcins, J. Kruse, D. Renner, M. Felber, S. Schreiber, M. Holz, M. Pelzer, A. Ziegler, M. Hoeptner, H. Sokolinski, M. Stadler, M. Wendt, I. Krouptchenkov and many others Thank you !
28
Nicoleta Baboi | Electron Beam Diagnostics at FLASH | FLASH Seminar | DESY, Hamburg | Nov. 11, 2014 | Slide 28 FLASH. Free-Electron Laser in Hamburg Other Electron Beam Monitors in FLASH2 > Loss monitors Cherenkov fibers along undulators (not for XFEL) > Beam Compression Monitors > Beam Arrival Monitors > Coherent Transition Radiation monitors C. Behrens, IPAC’10, p.912 A. Angelowski, IBIC’14, MOPD25 S. Wesch, BIW 2012, THAP01