Presentation is loading. Please wait.

Presentation is loading. Please wait.

INSIDE 15/16-09-2014 MECHANICS UPDATES CHARIOT - NEW DESIGN COOLING PET DETECTOR G. GIRAUDOTORINO 15/16 - 09 -2014.

Similar presentations


Presentation on theme: "INSIDE 15/16-09-2014 MECHANICS UPDATES CHARIOT - NEW DESIGN COOLING PET DETECTOR G. GIRAUDOTORINO 15/16 - 09 -2014."— Presentation transcript:

1 INSIDE 15/16-09-2014 MECHANICS UPDATES CHARIOT - NEW DESIGN COOLING PET DETECTOR G. GIRAUDOTORINO 15/16 - 09 -2014

2 G. GIRAUDOTORINO 15/16 - 09 -2014 MECHANICS UPDATES Not approved by CNAO

3 G. GIRAUDOTORINO 15/16 - 09 -2014 MECHANICS UPDATES ALL COMMERCIAL COMPONENTS ARE CERTIFIED FOR MEDICAL ENVIRONMENTS (CE)

4 G. GIRAUDOTORINO 15/16 - 09 -2014 MECHANICS UPDATES PET TRACKER VERTICAL SLEDGES LINEAR GUIDES NADELLA LM65-1200 GAS SPRINGS LIFT-O-MAT WEELS TENTE

5 G. GIRAUDOTORINO 15/16 - 09 -2014 MECHANICS UPDATES STRESS ANALISYS 10 MPa (130 MPa)

6 G. GIRAUDOTORINO 15/16 - 09 -2014 MECHANICS UPDATES

7 G. GIRAUDOTORINO 15/16 - 09 -2014 MECHANICS UPDATES

8 STAUBLI RME09-7102-0 TO INSIDE FROM INSIDE MSC SRH 4110-03 MANIFOLD 1x4 REG 1202 MANIFOLD 1x4 CHILLER SMC HRS012-AF-20-MT G. GIRAUDOTORINO 15/16 - 09 -2014 MECHANICS UPDATES

9 G. GIRAUDOTORINO 15/16 - 09 -2014 MECHANICS UPDATES The INSIDE support mechanics has been redesigned The structural FEM analysis ready/done The cooling system drawing is ready Patch Panel location to be decided Chiller type to be decided Report on mechanics/cooling to be submitted to CNAO ready soon

10 G. GIRAUDOTORINO 15/16 - 09 -2014 PET detector updates

11 G. GIRAUDOTORINO 15/16 - 09 -2014 PET detector updates COOLING PLATE – NEW DESIGN ALUMINIUM ALLOY AlSi7Mg T6  150 W/m∙K AISI 316 L  15 W/m∙K Tube embedded in the melted plate found an art foundry (ITALFOND) plate in order: minimum 8 pieces

12 G. GIRAUDOTORINO 15/16 - 09 -2014 PET detector updates

13 G. GIRAUDOTORINO 15/16 - 09 -2014 PET detector updates

14 G. GIRAUDOTORINO 15/16 - 09 -2014 PET detector updates The INSIDE support mechanics has been redesigned The structural FEM analysis ready/done The cooling system drawing is ready Patch Panel location to be decided Chiller type to be decided Report on mechanics/cooling to be submitted to CNAO ready soon Cooling plates has been redesigned Cooling tube embedded in the melted plate New drawing from Hamamatsu Reference mark locations have to be found