HARPS... North Geneva Observatory, Switzerland Francesco Pepe et al.
2
What’s HARPS? Fiber fed, cross-disperser echelle spectrograph Spectral resolution: geometrical 84’000, optical 115’000 Field: 1 arcsec on the sky (HARPS-N: 0.9 arcsec!) Wavelength range: 383 nm - 690 nm Sampling: 4 px per geometrical SE (3.3 real) Environmental control Drift measurement via simultaneous thorium
4
The Doppler measurement cross-correlation mask
5
Error sources Stellar noise (or any other object) Contaminants (Earth’s atmosphere, moon, etc.) Instrumental noise ✴ Calibration accuracy (any technique) ✴ Instrumental stability (from calibration to measurement) Photon noise
6
Stellar “noise”: p-modes
8
Stellar “noise”: Activity
9
Contaminants: Atmosphere
10
Photon “noise” Is NOT only SNR !!!! Spectral resolution Spectral type Stellar rotation
11
Contaminants: Close-by objects
12
Flux Photon “noise”: Spectral information
13
Photon “noise”: Spectral resolution
14
Photon “noise”: Stellar rotation
15
Instrumental errors External ✴ Illumination of the spectrograph Internal ✴ “Motion” of the spectrum on the detector
16
Limitations: Telescope centering and guiding Slit spectrograph Δ RV 1 arcsec Stored guiding image for QC
17
Limitations: Light-feeding Fiber-fed spectrograph Fiber entrance Fiber exit Image scrambler Guiding error: 0.5’’ → 2-3 m/s for a fiber-fed spectrograph
18
ΔRV = 1 m/s Δ = 0.00001 A 15 nm 1/1000 pixel ΔRV =1 m/s ΔT = 0.01 K Δp = 0.01 mBar Vacuum operation Temperature control Instrumental stability
19
Design Elements Fiber feed (mandatory for this techniques) Stable enviroment (gravity, vibrations, etc.) Image Scrambling No moving or sensitive parts after fiber SIMPLE and ROBUST optomechanics “Best” (reasonably) achievable env. control ✴ Vacuum operation ✴ Thermal control High spectral resolution
20
Instrumental stability
21
Line (and Instrumental) stability Absolute position on the CCD of a Th line over one month
22
Object ThAr Simultaneous reference
23
Object fiber RV 0 ThAr reference Object spectrumThAr spectrum RV 0 Wavelength calibration
24
Object fiber RV 0 ThAr referenc e Object spectrum ThAr spectrum RV 0 Measurement RV (object) =- RV (measured) RV(drift)
25
Simultaneous reference
26
The wavelength calibration px
27
Instrumental errors: Calibration pixel-position precision ✴ photon noise ✴ blends ✴ pixel inhomogeneities, block stitching errors accuracy of the wavelength standard ✴ systematic errors, Atlas, RSF ✴ instabilities (time, physical conditions: T, p, I) accuracy of the fit algorithm
28
Calibration: The problem of blends Isolated lines are very rare! Fit neighbouring lines simultaneously with multiple Gaussians
29
But HARPS-N is also...... a software concept delivering full precision observables: Scheduling many observations efficiently Full quality pipeline available at the telescope Fully automatic, in “near” realtime, RV computation Link to data analysis Continuous improvements and follow-up
30
Limiting factors and possible improvements New calibration (and sim. reference) source Perfect guiding and/or scrambling, good IQ needed Improve detector stability (mounting, thermal control)
31
Subsystem break-down Isolation box Services Fiber run Detector Spectrograph room Adapter LCUs WS CfA OG ESO/OG Spectrograph Vacuum system
32
Subsystem: Opto-mechanics
33
Subsystem: Detector
34
Subsystem: Exposure meter
35
Exposure meter
36
Subsystem: Vacuum System
37
Subsystem: Fiber run
38
Subsystems: Front end, HW, SW Calibration fibers (0.3mm dia.) CfA
39
Interfaces CfA - OG I. Detector - Spectrograph II. Fiber run - Front end III. Vacuum System - HARPS Room/Enclosure IV. Electronic components
40
Detector - Spectrograph ✓ Chip position and tilt ✓ Field-lens tilt ✓ Electrical connectors and cables ✓ Front-amplifier size and location -> ICD between SP and DU
41
Fiber run - Front end ✓ Fiber-hole position(s) ✓ Mirror position and tilt ✓ Mirror shape (possibly flat !) -> ICD between FR and FE
42
Vacuum system - Spectrograph Room ✓ Heat load on spectrgraph room ✓ Rail-fixation plate ✓ Location of services ✓ Feed-through window through SR wall ✓ Hoist > 2500 kg -> ICD between VS and SR
43
Spectrograph electronics Elements to be integrated in SW: ✓ F-200 Temperature controller (conf., read) ✓ Agilent pulse counter (conf., read) ✓ Pfeiffer Digiline P-sensors (read) ✓ Uniblitz shutter controller (read/write) ✓ Lakeshore T-controller for CCD (conf., read) ✓ Lakeshore T-controller for Isolation Box (conf., read) ✓ I-Omega T-controllers for CFC -> temperatures and alarms (read) ✓ LN2-level gauge (read)
44
Best wishes to HARPS-N
45
3-level concept Spectrograph room: +- 0.2 K Isolation Box: +- 0.01 K Spectrograph: +- 0.001 K 15°C 17°C
46
Spectrograph room Model : YORK YEB 3S Serial Nr. : 135.157.DN003
47
Room thermal control
48
Temperature control ✓ Lakeshore 331S T-controller + diode sensors + heaters ✓ 80 mm polysterene panels ✓ Thermal load on Room: 10 W/K
49
Performances, but...
50
Leassons learned Concept works well and is simple Changing thermal load through feet produces gradient and seasonal effects ➡ Thermal isolation of feet ➡ Heater below feet, Tref = vacuum vessel
51
Project schedule OG 2008: Procurement of components 04/2008 - 04/2009: Manufacturing of mechanical parts for vacuum and optics 01/2009: Start assembly 03/2009: Delivery of FA, DU and Control HW and SW by CfA to OG 04/2009 - 07/2009: Integration and tests OG