1
Perchlorate April 14, 2004 An Integrated Approach to Ensure Reliable Perchlorate Analytical Results
2
Agenda n Perchlorate Chemistry and Sources n Current Methods for Testing n Ion Chromatography n LC/MS/MS n An Integrated Approach n A Success Story n Summary and Conclusions
3
Perchlorate Ion n Perchlorate is a compound containing one chlorine atom and four oxygen atoms. n Perchlorate is an ion, meaning it carries an electrical charge, as indicated by the negative sign.
4
Natural Sources of Perchlorate n Chilean fertilizer deposits n New Mexican potash n Canadian potash n Californian hanksite n Bolivian playa crusts
5
Anthropogenic Sources of Perchlorate n Rocket Fuel n Fireworks n High explosives n Flares n Herbicides n Automobile airbags n Tracer munitions n Detergents ?
6
How is perchlorate measured? n EPA Method 314.0 Ion chromatography/conductivity detector n Enhanced (Modified) 314.0 Suppressed Conductivity Detector n SW-846 Method 9058 (draft) Ion Chromatography/conductivity detector
7
How is perchlorate measured? n Other analytical methods that can detect perchlorate: IC “Improved” Method 314.1 approved use of 2 nd column confirmation and pre-concentration of samples. IC/MS (ion chromatography/mass spectrometry) Proposed Method 330.0 LC/MS (liquid chromatography/mass spectrometry) LC/MS/MS (liquid chromatography/mass spectrometry/mass spectrometry) Method 8321 Modified, Proposed Method 331.0
8
Ion Chromatography
9
n An aqueous sample, comprised of components A and B, is injected into the ion chromatograph. n The sample enters the analytical column which is packed with a material that can adsorb the components of the sample. n Some components are more strongly adsorbed. Diagram from Fundamentals of Analytical Chemistry, Skoog, 1988.
10
Ion Chromatography n Fresh solvent is continuously injected into the instrument. n The solvent carries the sample components, A and B, through the column. n Component B is more strongly adsorbed onto the column and is moving thought the column at a slower rate. n The boundary between A and B is indistinct at this point. Diagram from Fundamentals of Analytical Chemistry, Skoog, 1988. Detector
11
Ion Chromatography n Component B is almost completely separated from component A. n The boundary between the two components is more distinct. Diagram from Fundamentals of Analytical Chemistry, Skoog, 1988. Detector
12
Ion Chromatography n Component B is completely separated from component A. n Component A exits the column through the detector at time t 3. n The detector records component A as a peak (pictured below the column) in the sample chromatogram (graph). Diagram from Fundamentals of Analytical Chemistry, Skoog, 1988. Detector
13
Ion Chromatography n Component B exits the column through the detector at time t 4. n The detector records component B as a peak (pictured below the column) in the sample chromatogram. Diagram from Fundamentals of Analytical Chemistry, Skoog, 1988. Detector Solvent
14
Diagram from Fundamentals of Analytical Chemistry, Skoog, 1988.
15
Identification of Perchlorate by Ion Chromatography 11.2.4The width of the retention time window used to make identifications should be based upon measurements of actual retention time variations of standards measured over several days. Three times the standard deviation of retention time may be used as a suggested window size but the retention time window should not extend beyond ± 5% of the retention time for perchlorate. The experience of the analyst should weigh heavily in the interpretation of these chromatograms. EPA Method 314.0
16
IC Perchlorate Standard
17
High Conductivity
18
Carry Over
19
Matrix Spike Confirmation
20
Matrix Spike Non- Confirmation
22
Liquid Chromatography/Mass Spectrometry/Mass Spectrometry
23
LC/MS/MS n Leading edge technology n DoD Recommended n Pending Regulatory Approval
25
API MS/MS
26
Bowling Balls and Feathers: a simplification of mass spec separations
27
Identification By LC/MS/MS n Liquid chromatography separates perchlorate from other sample components in the same manner that ion chromatography does n Mass spectrometers (MS) are used as detectors in place of the conductivity detector used in IC. n Unlike conductivity detectors, an MS is a specific detector
28
Detection By LC/MS/MS n The MS counts the number of ions that have a specific molecular weight. Perchlorate has a characteristic molecular weight (MW). n The first MS counts the number of ions that have the MW of perchlorate: 1 Chlorine (35) + 4 Oxygens (16) = 99 n The second MS removes one of the oxygen atoms and then counts the number of “daughter” ions that have that MW: 1 Chlorine (35) + 3 Oxygens (16) = 83 n Chlorine has a less abundant isotope (MW 37), so ions with MW 101 are also counted by the first MS and ions with MW 85 are counted by the second MS. The retention time of the peak for MW 99 must be the same as the retention time of the peak for MW 83 for perchlorate to be confirmed. If the retention times are different, then the two molecular weight were detected from two different components and perchlorate is not confirmed.
29
Non-detect Analysis by LC/MS/MS Analysis by LC/MS/MS n Reporting limit is 2 g/L. n Bottom chromatogram shows the counts of ions with molecular weight 83 (1 chlorine + 3 oxygens). n Top chromatogram shows the counts of ions with molecular weight (the isotope of chlorine). n Perchlorate is not detected.
30
Comparison of LC/MS/MS Non-DetectDetect
31
n Reporting limit 2 g/L. n Bottom left shows MW 83 (1 chlorine + 3 oxygens). n Top left shows MW 85 (the isotope of chlorine). n Right chromatogram show the 18 O internal standard (recovery 88%). n Perchlorate is detected. Detect Analysis by LC/MS/MS
32
Perchlorate Decision Tree
33
Success Using the Perchlorate Decision Tree n Approximately 1000 samples were collected and analyzed for perchlorate from August 2003 through January 2004. n 89 samples (about 9% of total) were pretreated and reanalyzed. n 39 samples (about 3% of total) were reanalyzed by LC/MS/MS - 24 detects and 15 nondetects. n 10 perchlorate detects (about 0.4% of total) were confirmed by LC/MS/MS – 58% false positive rate and a 0% false negative rate.
34
Summary n Why use 314.0? Cost Capacity Availability Promulgated Methods n When to use LC/MS/MS? Definitive Data Critical Samples Small Projects
35
Thank You Contact Elizabeth Wessling at 303.935.6505 elizabeth.wessling@amec.com Questions and Comments