Download presentation
Presentation is loading. Please wait.
Published byRhoda Grant Modified over 9 years ago
1
Technical Update Detection / Quantitation and PBMS Richard Burrows STL ACIL Mid-Winter Meeting, February 2006
2
ACIL objectives Technically reliable detection and quantitation procedures that are: Reasonably straightforward to perform Implemented widely through replacement of the MDL and ML procedures at 40CFR Part 136 Appendix B www.epa.gov/waterscience/methods/det
3
D/Q FACA People Lab Group Richard Burrows (STL) Steve Bonde (Batelle) Nan Thomey (Environmental Chemistry Inc) Cary Jackson (Hach Instruments) State Group Bob Avery, Michigan DEQ Timothy Fitzpatrick, Florida DEP Thomas Mugan Wisconsin DNR Dave Akers Colorado DEP Industry Group John Phillips, Ford Larry LaFleur, NCASI Roger Claff, API David Pilar, Exelon Power Environmentalists Richard Rediske, Michael Murray Barry Sulkin
4
People Wastewater treatment group David Kimborough Chris Hornback, National Association of Clean Water Agencies James Pletl, Hampton Road Sanitation District Zonetta English, Louisville and Jefferson County Metrepolitan Sewer District
5
Technical Workgroup Labs Richard Burrows Steve Bonde States Cliff Kirchmer, Washington State Dept of Ecology Tim Fitzpatrick Industry John Phillips Larry LaFleur Treatment plants Jim Pletl Ken Osborn Environmentalists Richard Rediske David Rocke, UC Davis Federal Bill Foreman, USGS Bill Ingersoll, Navy
6
Multi lab / Multi concentration procedures IDE / IQE ASTM Procedures LCMRL Drinking Water Hubaux Vos
7
Single lab procedures Detection ACIL procedure Consensus group procedure MDL Osborn
8
Single Lab procedures Quantitation ML MRL Extension of ACIL procedure?
9
Single lab procedures Quantitation MRL 7 replicates at proposed quantitation limit Mean recovery plus 3.96 times RSD must be < 150% Mean recovery minus 3.96 times RSD must be > 50% ML 3.18 times L C (MDL) For example, if mean recovery = 75%, then RSD must be < 6%
10
Definitions L C The smallest result that can be distinguished from a blank L D The smallest true concentration that gives a result greater than L C L Q The smallest concentration where a specified degree of uncertainty is met
11
Distribution of results from true zero concentration LCLC Distribution of true concentration at zero 0
12
Distribution for true concentration = L C 0 L C or MDL Distribution of true concentration at L C Note that if the true concentration is L C, 50% of the results cannot be distinguished from a blank False negative rate = 50%
13
Distribution for true concentration = L D 0LCLC Distribution of true concentration at L D LDLD False negative rate for a true concentration at LD is 1% (or 5%)
14
Controversies Is L D needed? In theory, LD is the concentration for which 1% false negatives are observed It can be statistically predicted, but requires assumptions regarding qualitative identification criteria, constant variance, spike recovery, normal distribution, etc. We do not use L D now (MDL = L C )
15
Controversies Is L D needed? L D is very difficult to demonstrate, especially for multi analyte methods A large number of spikes at concentrations very close to L D would be needed
16
Controversies Is L D needed? If L D is not used then we report values between L Q and L C as estimated If L D is used, we do the same! What should the non detect value be? <L Q, L D, <L C, something else?
17
What is “Quantitative”? Above low calibration standard? Multiple of detection limit? Better than some defined value of precision and accuracy? Above low standard and demonstrate detection capability?
18
Low calibration standard Pros i. Very simple, no additional work ii. In line with some statements from OSW Cons i. Very weak technically – some analytes can be calibrated at much lower levels than they can be prepared and analyzed, and very large error (100% or more) in the low point of the calibration has minimal effect on the correlation coefficient. ii. Will be strongly opposed by industry, and probably Office of Water as well
19
Some multiple of Lc (eg 3.18) Pros i. Simple ii. Same as the MDL/ML relationship Cons i. For some censored tests such as GC/MS, may be below the detectable threshold ii. Will be strongly opposed by industry
20
Level at which some specified degree of precision and accuracy can be demonstrated Pros i. Strongest technically ii. Will be supported by industry and the drinking water contingent Cons i. Hardest to implement – who decides what level of precision and accuracy is acceptable, same criteria cannot be used for all analytes. ii. Requires a lot of low level spikes to develop the statistics
21
Combo approach Low calibration standard plus at least 3X L C plus periodic (eg once per quarter) demonstration of capability with a single L Q level spike extracted analyzed on all instruments Pros i. Compromise approach, hopefully would satisfy the most people. ii. Reasonably strong technically iii. Reasonably easy to implement Cons i. Probably will still be opposed by industry ii. More complex than options 1 or 2
22
Extension of ACIL Detection Limit procedure? Replace spikes used to determine L D with spikes used to determine L Q Should be ongoing, not one time or every year Provides a measure of variance at the quantitation limit that can also be used to meet uncertainty requirements Frequency? Level?
23
Controversies What is L C for a “censored” method? We do not observe results for blanks, so the level below which 99% of blank results fall is meaningless
24
Current activities Collection and evaluation of existing data Uncensored methods, L C A considerable volume of data has been collected, from commercial, state and wastewater treatment plant labs In general initial assessment seems to indicate that the ACIL and Consensus group procedure work well i.e., the number of blanks above the calculated L C is in the region of 1%
25
Current activities Collection and evaluation of existing data L D, L Q and censored methods Much less data available because low level spikes are needed. PT data may help, at least for L Q
26
Next steps Design and perform a pilot study Evaluate procedures for performance and usability Next meetings, March 28-29 and June 28- 29
27
PBMS
28
PBMS New interest at the agency, driven from the highest levels Office of Water proposes restating their “Streamlining” proposal
29
Streamlining Based on demonstrating equivalent or superior performance to a reference method Determine and compare to reference method MDL Calibration linearity Calibration verification IPR OPR Matrix spike precision and recovery Blanks
30
Streamlining Tiers Different requirements for Tier 1 Single lab, multi matrix Tier 2 Multi lab, single matrix from one industry type Tier 3 Nationwide, multi lab, multi matrix
31
Problems with “Streamlining” State audition procedures (both NELAC and non-NELAC) Legal issues Problems with marketing non-EPA methods Some limits would be difficult to meet, Eg: IPR limits X +/- 5.3s OPR limits X +/- 6s
32
An alternative Remove all possible method details from EPA methods A detailed procedure could still be developed and published (ES&T, Anal. Chem) but would not be in regulation Control the performance of the analytical technique for the analyte through QC criteria.
33
Perchlorate example EPA has developed 5 new methods for perchlorate, all have some prescriptive details and are not consistent with each other. Total 169 pages – one analyte
34
PBMS perchlorate method Perchlorate is determined by LC/MS/MS or IC/MS/MS. Mass 83 is used for quantitation and mass 85 for confirmation. In the absence of interferences, single stage mass spectrometry may be used, in which case use mass 99 for quantitation and 101 for confirmation. 18 O labeled perchlorate is used as an internal standard. Aqueous samples may be analyzed directly or following clean up on solid phase columns. Solid samples are extracted using an aqueous leach. Identification of perchlorate requires peaks for the quantitation and qualifier ions to maximize within one scan, and within 0.1 minutes of the labeled perchlorate internal standard. In the absence of project specific quality control requirements the following must be used:
35
Questions?
Similar presentations
© 2025 SlidePlayer.com. Inc.
All rights reserved.