Presentation is loading. Please wait.

Presentation is loading. Please wait.

Novel high resolution detectors for Positron Emission Tomography (PET)

Similar presentations


Presentation on theme: "Novel high resolution detectors for Positron Emission Tomography (PET)"— Presentation transcript:

1 Novel high resolution detectors for Positron Emission Tomography (PET)
Siena 2002 8th Topical Seminar on Innovative Particle and Radiation Detectors October 2002 Siena, Italy Novel high resolution detectors for Positron Emission Tomography (PET) N. Belcari, M. Camarda, A. Del Guerra, A. Motta, S. Righi, A. Vaiano Department of Physics, University of Pisa and INFN Sezione di Pisa, Via F. Buonarroti 2, I Pisa (Italy) G. Di Domenico, G. Zavattini Department of Physics, University of Ferrara and INFN Sezione di Ferrara, Via Paradiso 12, I Ferrara (Italy)

2 Positron Emission Tomography (PET)
Outline of the talk Positron Emission Tomography (PET) Clinical PET Dedicated PET systems Novel detectors for High Resolution PET Requirements PS-PMT Multi-Pixel HPD

3 PET principle A b+ emitting radiotracer is injected to a “patient”
The radiotracer marks a specific function (e.g. glucose metabolism) - Uptake process The positron annihilates with an electron and one pair of nearly collinear 511keV photons are emitted in opposite direction. 511keV photon A set of detectors surrounding the “patient” detects the pair of photons (time coincidence) (Q,Z or X,Y coordinates of the point of interaction) An algorithm performs the 3-D reconstruction of the activity density within the body. Ring geometry Parallel plane geometry

4 Physical problem: Position sensitive detection of both the 511 keV annihilation photons Matrices of scintillators are usually used for the photon detection Fast, good energy resolution, high Z For the position determination we have to determine the crystal where the interaction occurs  Position Sensitive Photodetector YAP:Ce matrix 4cm  4cm  3cm (400 elements, 2  2  30 mm3 each) PIXEL IDENTIFICATION!

5 Whole body PET scanners
Needs of dedicated scanners Whole body PET scanners 2% Solid Angle Coverage 4-8 mm spatial resolution High cost for routine scanning For specific application such as “small animal functional imaging” or “Positron Emission Mammography (PEM)” dedicated scanners are needed so as to offer: Higher spatial resolution ( <3mm FWHM for PEM) to reliably detect small tumors (< 5 mm Ø) ( <2mm FWHM for Small Animal PET) to detect very small structures (hot spots) Higher sensitivity to reliably detect very low specific activity (1mCi/cc) with a low uptake ratio (hot spot/Background < 10) in a short time (10-20 min.) Better handling (easier positioning of the mouse or of the breast within the FoV) Application specific design (Rat or Mice for Small Animal PET), (Fit the breast for Axillary lymph nodes scanning or compression mode PEM) Multimodality scanning (PET-SPECT, RX-PEM) Use of a reduced number of detectors  reduction of cost

6 Spatial Resolution Requirements
Human body: ~70 kg Heart mass: ~300 g Aorthic cannula Ø : ~ 30 mm Rat body: ~200 g Heart mass: ~1 g Aorthic cannula Ø : mm Mouse body: ~20 g Heart mass: ~0.1 g Aorthic cannula Ø : mm Relative heart size Required spatial resolution 4-8 mm FWHM ( mm3) 2 mm FWHM (8 mm3)  1 mm FWHM ( 1 mm3) How to achieve this ? Small size detectors (high pixellization) Individual detectors or “perfect” coding

7 Sensitivity Requirements
Imaging of low activity sources low uptake processes such as in small cancers Imaging of small hot spots discrimination of small regions with a low spot/background ratio How to achieve this ? High geometry efficiency (large solid angle covered by detectors) High detection efficiency (e.g. for crystals: high Z, high density)

8 Detectors for High Resolution PET

9 Detector configuration- Light sharing
X Center of gravity calculation for X and Y B A D C Identification from a LUT Y The light distribution is measured by 4 PMTs A BGO block is saw at different depth. The cuts are filled with a reflective material and provide a light guide to the PMTs. CTI Exact HR detector block* Advantages Reduction of the number of PMT Cheap Easy to pack Drawbacks Loss of resolution - Statistical light sharing - Distortion - Pileup at high count rate *Casey & Nutt, IEEE TNS 33 (1986)

10 Detector configuration- Individual coupling
Early PET detector element Novel configuration Single PMT Array or independent photodiodes (APD) Scintillator block (BGO) Matrix of scintillators (BGO/LSO) + Simple coding (“parallel” operation) + High spatial resolution with no distortion + Small size + High count rate - Quite difficult to pack - Tricky tuning - Gain dependence on bias & temperature - Expensive + Simple coding (“parallel” operation) - Difficult to pack - Expensive Si APD array (8×4) Hamamatsu S8550 A detector module made of a 8×4 LSO matrix couped to a Hamamatsu S8550 (Pichler B., IEEE TNS 45 (1998) )

11 Electronic coding - PS-PMT readout (Single tube)
A single position sensitive PMT is used for the identification of the hit crystal in a scintillator matrix Advantages Large active area (up to 100 mm Ø) Good spatial resolution (up to 0.5 mm FWHM) Good uniformity Limited number of channels (up to 28 x + 28 y crossed wire anodes) Easy to use with a resistive chain (4 channels) Drawbacks Distortion at the borders Low packing fraction (limited active area) Round shape (difficult to pack) The PS-PMT R2486 by Hamamatsu. Active area 50 mm Ø 16 x + 16 y anodes Flood field irradiation(122 keV) of a 20 × 20 crystals(2 mm × 2 mm each) YAP:Ce matrix read by a Hamamatsu R2486 (resistive readout)

12 Electronic coding - PS-PMT readout (Multi tube)
An array of densely packed small and square position sensitive PMT is used to build a large area photodetector for the readout of a scintillator matrix Advantages Higher packing fraction with respect to round tubes (up to 73%) Good spatial resolution Good uniformity Easy to use with a resistive chain (4 channels) YAP:Ce matrix, 11 × 11 crystals (2mm × 2mm each) Drawbacks Dead area between adjacent PMTs Small active area for each tube CsI:Tl matrix, 8 × 8 crystals (2.8 mm × 2.8 mm each) The PS-PMT R8520-C12 by Hamamatsu. Active area mm × 22 mm 6 x + 6 y anodes Flood field irradiation(122 keV) of a matrix of scintillators read by a Hamamatsu R8520-C (single tube, resistive readout)

13 PS-PMT readout - Multi tube recovery of dead area - light sharing
Matrix of scintillators Flood field (122 keV) image of a matrix of a NaI(Tl) crystals (2 × 2 × 6 mm3 each) coupled to an array of four R7600-C12 via the transparent window of the NaI matrix* Transparent window + Simple and cheap - Resolution worsening - Requires enough light yield *Courtesy of R. Pani, University of Rome, La Sapienza

14 Effect of: Quartz window - White reflector
Hamamatsu R8520-C12 readout Effect of: Quartz window - White reflector YAP:Ce Matrix 25 crystals 2230 mm3 each Hamamatsu R8520-C12 (resistive readout) YAP:Ce Matrix 121 crystals 2230 mm3 each YAP:Ce matrix White reflector FWHM  0.8 mm Direct contact (optical grease) 511 keV source Quartz window Direct contact (optical grease) FWHM 2.6 pixel (1.0 mm) FWHM  0.7 mm FWHM 1.4 mm 3mm quartz window (optical grease) FWHM  1.1 mm NO YES NO 3mm quartz window (optical grease) White reflector (about 15% more light with a white reflector)

15 Hamamatsu R8520-C12 readout - Multi tube
recovery of dead area - Quartz window Two Hamamatsu R8520-C12 (resistive readout) YAP matrix 1124 crystals (2mm  2mm  30mm each) White reflector 22Na source BGO crystal + PMT for selecting 511keV annihilation photons Quartz window 3 mm thick Two rows of crystals (2 mm +2 mm) outside the active area 22mm Flood field irradiation of the YAP matrix 1124 crystals (511 keV). All crystals can be identified 22mm

16 recovery of dead area - fibers
PS-PMT readout - Multi tube recovery of dead area - fibers Matrices of scintillators Matrix of scintillators Fiber optic bundle (individual coupling) FOP (Fiber Optic Plate) Example: the detector element of the MicroPET® + Linear demagnification of the image - Expensive + No demagnification - Only for ring configuration PS-PMT Hamamatsu R7600-C8 8 × 8 array of 2 mm square fibres by Kuraray 8×8 LSO matrix

17 Hamamatsu Flat Panel Multi Anode PMT
See next talk at this conference “Flat Panel PMT: advances in position sensitive photodetection” R. Pani, University of Rome Hamamatsu R8500

18 (better than 92% with a quartz window)
Comparison Application: readout of a 5 cm × 5 cm matrix of scintillating crystals R2486 Array of 4 R8520 R8500 Flat Panel See next talk 78% 77% (better than 92% with a quartz window) 96% Active area

19 What is an HPD (Hybrid Photo Diode)* ?
*R. de Salvo et al. NIM A315 (1992) Main characteristics Good quantum efficiency Low noise Single photon counting capabilities Single or Multi - pixel structure Very low output signal (Low noise and high gain readout electronics is crucial) HV (up to tens kV required) Available HPD structures: Proximity focused: Electrons reach the Si sensor by straight lines (no demagnification) (multi-pixel: up to 73 pixel, up to ~1” Ø, good spatial resolution) Fountain focused: Electrons are focused on the Si sensor (linear demagnification) (multi pixel: up to 73 pixel, up to ~ 3” Ø, poorer spatial resolution, spherical entrance window) Picture from:

20 Multi Anode HPD - WLS Fibres readout
Large number of crystals and large area crystal read out Separation of scintillating crystal from detector and read out electronics No dead zone around the detector Possibility of Z coordinate measurement To 61 Pixel HPD Scintillator matrix To 61 Pixel HPD WLS fibres 2.5 p.e. for 122keV (57Co source)* ~10 p.e. for 511keV (22Na source) (photopeak event)* YAP:Ce matrix Not enough light yield! 61 pixel HPD proximity focusing by DEP ~1” WLS fibres by Kuraray *N. Belcari et al. NIM A461 (2001)

21 Multi Anode HPD - Direct readout
Multi anode HPDs can be used as a PS-PMT for the direct readout of a matrix of scintillating crystals and readout by electronic coding. Four peaks of the second row: the peak-to-valley ratio is about 4. 3-D profiles of 4  2 crystals of a YAP:Ce matrix (2 mm side each ). Pixels are clearly separated.* *N. Belcari et al. Presented at the conference “New Developments in Photodetection” Beaune, June 17-21, 2002 (NIM 2002) HAMAMATSU PMT R-5900 4  4 YAP:Ce MATRIX 2 mm 2 mm 2 mm 12 mm DEP HPD 61 PIXEL Larger HPD with higher pixellization will be soon available. A 5” HPD is produced at CERN* *A. Braem et al., NIM A (2002)

22 Multi Anode Large area HPD - Direct readout
Near ?? future Further developments are needed in order to make this technology available. New devices and new front-end electronics (low noise-high gain-fast triggering preamplifiers) are necessary. Still a dream* 5” Pad HPD CERN Optimized for medical applications Spherical window + bundled small fibers or Fiber Optic Plate (FOS) K2CsSb photocatode *C. Joram et al. Presented at the conference “New Developments in Photodetection” Beaune, June 17-21, 2002 (NIM 2002)

23 Dedicated Scanners for
Small Animal PET To fulfil the spatial resolution and sensitivity requirements dedicated instruments are needed. Many different technologies are actually used in these scanners. Scintillator crystals (BGO, LSO, YAP) Position sensitive PMT (electronic coding) Avalanche Photo Diode (individual coding) Multi Wire Proportional Chambers (charge sharing) Review of Small Animal PET Scanners* *(A. Del Guerra, N. Belcari. Q. J. Nucl. Med. 46(1) (2002) 35-47).

24 microPET®4 - Images FDG in rat myocardium
Courtesy of S. Cherry, UCLA, 2002 FDG in rat myocardium Imaging dopamine receptors in the rat FDG in rat brain Gene expression in mouse liver

25 In vivo dynamic imaging of rat brain
YAP-(S)PET University of Ferrara Ex vivo images of a rat brain injected with 18F-FDG (left) and 18F-FESP (right) A. Del Guerra et al. IEEE-MIC Seattle, 1999, M10-45 In vivo dynamic imaging of rat brain with 11C-Flumazenil The rat was injected with 500 µCi of 11C-Flumazenil. Sagittal section through the center of the brain. (Nose towards the left. Upper skull towards top)

26 Dedicated scanners for Positron Emission Mammography (PEM)

27 Pisa YAP-PEM prototype*
YAP:Ce crystal matrix 6cm × 6cm × 3cm 2mm × 2mm × 3cm each (900 elements) Monte Carlo Simulation PS-PMT R8520-C12 (Hamamatsu) 2.2cm × 2.2 cm active area The tumor is embedded in an active breast tissue, with specific activity ratio 10:1 between cancer (1.0 mCi/cc) and tissue (0.1 mCi/cc). The distance between detectors is 5 cm. Variable distance for compression ( cm) Optical diffuser (Quartz) Reconstructed images of simulated tumors in active breast tissue. Tumors have different size (diameter/volume): (a) 12.4 mm / 1.0 cc; (b) 9.8 mm / 0.5 cc; (c) 5.8 mm / 0.1 cc; (d) 5.0 mm / cc Profiles of the images a,b,c SNR values as a function of tumor size. A. Del Guerra et al. To be presented at IEEE-MIC, Norfolk, November 2002

28 Scintillator: YAP:Ce 30  30 crystals 2  2  30 mm3 each
YAP-PEM simulation Results with YAP-PET (2 heads) Tumor diameter 5.0 mm (in breast tissue) Tumor / background activity: 1 mCi /cc / 0.1 mCi /cc (10:1) Moderate compression (5 cm) Events 3×106 (~10 min.) - Threshold 50keV The simulated breast is a specially designed 68Ge solid phantom. A tumor is simulated by a cylinder 0.5 cm Ø and 0.5 cm height with an activity of 44 nCi. The breast tissue is simulated by an active cylinder with diameter of 10.8 cm and height of 6 cm with a distributed 68Ge planar source of 25 mCi . The specific activity tumor/breast ratio is about 10:1 for all sources. Scintillator: YAP:Ce 30  30 crystals 2  2  30 mm3 each 1 detector 6.0  6.0 cm2 4 detectors 2.2  2.2 cm2 Experimental Monte Carlo   2.3% 9.3  106 counts SNR = 10.7  1.0   0.8% 3.0  106 counts SNR = 6.0  0.9 The image of the 5.0 mm source after the subtraction of the backprojection of the planar source.

29 Conclusions Dedicated scanners for small animals
Needs well established Commercially available Further developments required Dedicated scanners for Positron Emission Mammography Needs under scrutiny Few research prototypes available Similarities between these two types of scanners A high density / medium Z scintillator (such as YAP) coupled to novel PS-PMTs could be a very successful detector for these apparatus (e.g. YAP-(S)PET and YAP-PEM)


Download ppt "Novel high resolution detectors for Positron Emission Tomography (PET)"

Similar presentations


Ads by Google