GCMS: gas chromatography and mass spectrometry

Similar presentations


Presentation on theme: "GCMS: gas chromatography and mass spectrometry"— Presentation transcript:

1 GCMS: gas chromatography and mass spectrometry
in Forensic Science Applications

2 What does the GCMS do? Separates complex mixtures of chemicals found in substances found at crime scene Quantifies the amount of each chemical present in a sample, compared to a standard part of database Identifies chemicals – determines the unique chemical composition and structure-its fingerprint Compared to a standard fingerprint

3 volatile liquids (~1L)
Gas Chromatography Gas Chromatograph Sample: mixture of volatile liquids (~1L) Gas Chromatogram B E C A Overview of what a GC can do. ~ microliter injection volume is example only - sould not be construed as definitive Abundance D 5 10 15 20 Time (minutes)

4 Gas Chromatograph Injection Port Detector Capillary Column Data System
or Recorder Block diagram of GC. Carrier Gas Supply Oven

5 Typical sample: isolated
Mass Spectrometry O C H 3 Mass Spectrometer H C C N 3 N C C H C C O N N H Typical sample: isolated compound (~1 nanogram) Mass Spectrum 194 67 109 Horizontal axis (mis)labeled as mass rather than m/z so as to not to confuse them at first. Special molecule for Seattle. nanogram amount is an example only - should not be construed as definitive. Abundance 55 82 42 94 136 165 40 60 80 100 120 140 160 180 200 Mass (amu)

6 Mass Spectrometer Sample Introduction Data Output Inlet Data System
Source Mass Analyzer Ion Detector Block diagram of mass spectrometer. Vacuum Pumps

7 Electron Impact Ionization Source + _ _ + e- e- e- _ ~70 Volts
Electron Collector (Trap) Positive Ions + Repeller Neutral Molecules Inlet _ _ + to Analyzer + + + + + + EI source e- e- e- _ Electrons Filament Extraction Plate

8 Magnetic Sector Mass Analyzer
ion trajectory in register ion trajectory not in register (too light) S Detector Ion Source N ion trajectory not in register (too heavy) Electromagnet

9 Quadrupole Ion Filter resonant ion non-resonant ion _ Detector + + _
Source DC and AC Voltages