Observations on possible factors affecting LAr purity on the 50kton TPC from the viewpoint of the internal walls Alejandro S. Díaz Chemical Engineering.

Similar presentations


Presentation on theme: "Observations on possible factors affecting LAr purity on the 50kton TPC from the viewpoint of the internal walls Alejandro S. Díaz Chemical Engineering."— Presentation transcript:

1 Observations on possible factors affecting LAr purity on the 50kton TPC from the viewpoint of the internal walls Alejandro S. Díaz Chemical Engineering Department, Surface Analysis and Fluid Interaction Laboratory (ASIF), U of Concepción, Chile First FLARE Collaboration Meeting, Fermilab, Chile, November 4-6, 2003.

2 Summary In the present work, it is presented a first approach to answer some questions about the construction of a big scale LAr detector, under the viewpoint of the internal surfaces’ influence: -What factors may affect scaling up a detector to a very large tank? -What auxiliary experiments/measurements are necessary to provide an ironclad proof that it will have the necessary purity? -How well is the whole process understood?

3 ICARUS standard procedure vs FLARE 50kton ICARUS HNO3 wash Degassing of Assembly Hot air bakeout? N2 Purge Ar Purge LAr filling Water rinse small parts (120 ºC) Drying FLARE HNO3 wash Degassing of Assembly Evacuation to LN2 Cooling LAr filling Water rinse small parts 10-3 mbar Drying

4 Outgassing No vacuum pre-degassing of the detector: - Not evacuating the vessel will influence the residual oxygen in the chamber volume when purging -It has been observed the purification of the liquid argon via sorption of impurities on the walls, under the condition of previous degassing. -It will affect all internal walls' ability to adsorb impurities. One still can expect impurities to adsorb on the walls due to the low temperatures but gas exposed parts will slowly outgas (due to the lower total concentration of the bulk phase)

5 Outgassing, ICARUS data

6 Outgassing, calculated total mass

7 Outgassing, total mass by component -ICARUS run tests have shown that the main adsorbed component on internal surfaces is water (90% mass fraction). Hence, the need of a low temperature bake. -Conservative rough estimates of impurity content of the 50kton calculated extrapolating T600 outgas data (caveant): -If adsorbed water desorbs, it will imply a change in the load of the purifying systems.

8 Outgassing factors - Adsorption enthalphy is negative, enthropy is positive, but depends on chemical potential (which depends on concentration) higher concentration on surface = higher barrier to overcome for adsorption -It is not possible to say (a priori) which factor will be more weighting (P and T vs delta C) (But P and T could win, imho). -Some data on the Ar-H2O-N2-O2 system liquid-adsorption equilibrium in each exposed material should be needed. -The possibility of a low temperature bakeout of the entire detector assembly should be evaluated.Hot air, 120ºC or similar.

9 Fluid dynamics in the vessel: -A smaller vessel (T600) could be easier to homogeneize in the fluid composition. But Residence Times Distribution expected to be close to a countinuously stirred tank (Para, Tang). How close? ----> RTD

10 Fluid dynamics in the vessel: RTD will affect the gas purging time and LAr recirculation time required to obtain a desired purity. -Fluid speed distribution will determine the mass transfer process between the walls and the bulk liquid (boundary layer, eddy currents), given the slow thermal diffusion of oxygen in LAr ( D =2·10 -5 cm 2 s -1 ). CC C*C*  F

11 Fluid dynamics in the vessel Fluid dynamics in the vessel: -More detailed simulations of the inside-tank hydrodynamics could be done to clarify this.

12 What auxiliary experiments/measurements are necessary ? A lab scale experiment, (e.g. 50L) should have as objectives: -Study the new procedure influence on the Argon purity and consequently, on the purification system load -Quantify how the adsorption capacity is affected by the atm O 2 -N 2 exposure stages for a given material. Strong proof of feasibility will be given by the construction of a pilot scale system, with comparable hydrodynamic and thermal behavior, exposed surfaces S/V ratio and chemical composition.

13 Experiments A small volume cryostat, ~1 m3, with a purity monitor, purification system (filter beds, cryopumps, piping). Vessel with similar material surface proportions and similar (geometrical) surface to volume ratio to the projected detector.

14 Do we know well what is happening? It is necessary to analyze three aspects: CC C*C*  F C*=kC C*? Transport Interphase SLE Surface Adsorption Isotherms Data needed -Surface Area -Composition Hydrodynamics

15 Conclusions -The main difference between FLARE and ICARUS will be the (non) outgassing of the walls (from the viewpoint of walls and LAr purity ) -The amount of outgassables can be quantified extrapolating ICARUS data. -Equilibrium data of the H20-Ar-N2-O2 system is needed. (Thermo or real- world) -A “low temp bake” of the detector assembly would help to create a high purity environment -Hydrodynamics determine walls’ mass transfer. -Internal walls behavior is an important factor for the working of the detector

16 LAr purification via adsorption on the walls Argon purity behavior in a lab scale 10-4 mbar outgassed vessel, without purification systems (Bettini et al, 1991).

17 Outgassing, total mass by component Assuming that liquid submerged surfaces will not outgas, an estimate of the outgassable material can be obtained as: