Presentation is loading. Please wait.

Presentation is loading. Please wait.

Basic fMRI Physics In BOLD fMRI, we are measuring:

Similar presentations


Presentation on theme: "Basic fMRI Physics In BOLD fMRI, we are measuring:"— Presentation transcript:

1 Basic fMRI Physics In BOLD fMRI, we are measuring:
the inhomogeneities introduced into the magnetic field of the scanner… as a result of the changing ratio of oxygenated:deoxygenated blood… via their effect on the rates of dephasing of hydrogen nuclei. Ehhh???

2 History of MRI NMR = nuclear magnetic resonance
nuclear: properties of nuclei of atoms magnetic: magnetic field required resonance: magnetic field x radio frequency 1946: Block and Purcell atomic nuclei absorb and re-emit radio frequency energy 1992: Ogawa and colleagues first functional images using BOLD signal NMR  MRI: Why the name change? most likely explanation: nuclear has bad connotations less likely explanation: NMR means Nouveau Mouvement Religieux Bloch Purcell Ogawa

3 Necessary Equipment Magnet Gradient Coil RF Coil 3T magnet RF Coil
(inside) Magnet Gradient Coil RF Coil Source: Joe Gati, photos

4 Recipe for MRI 1) Put subject in big magnetic field (leave him there)
2) Transmit radio waves into subject [about 3 ms] 3) Turn off radio wave transmitter 4) Receive radio waves re-transmitted by subject Manipulate re-transmission with magnetic fields during this readout interval [ ms] 5) Store measured radio wave data vs. time Now go back to 2) to get some more data 6) Process raw data to reconstruct images 7) Allow subject to leave scanner (this is optional) Source: Robert Cox’s web slides

5 The Big Magnet B0 x 60,000 = Main field = B0 Continuously on
Very strong : Earth’s magnetic field = 0.5 Gauss / 1 Tesla (T) = 10,000 Gauss 3 Tesla = 3 x 10,000  0.5 = 60,000 x Earth’s magnetic field B0 x 60,000 = Source: Robarts Research Institute 3T

6 Source: http://www.simplyphysics.com/flying_objects.html
Safety The strength of the magnet makes safety essential : things fly – even big things! Source: Source: Anyone entering the magnet must be metal free This subject was wearing a hair band with a ~2 mm copper clamp. Left: with hair band. Right: without. Source: Jorge Jovicich Develop screening strategies : do the metal macarena!

7 1H aligns with B0 randomly oriented
Protons are abundant: high concentration in human body have high sensitivity: yields large signals longitudinal axis Outside magnetic field randomly oriented M = 0 transverse plane Inside magnetic field spins tend to align parallel or anti-parallel to B0 net magnetization (M) along B0 spins precess with random phase no net magnetization in transverse plane only % of protons/T align with field Source: Mark Cohen’s web slides M Source: Robert Cox’s web slides Longitudinal magnetization

8 Larmor Frequency Larmor equation : resonance frequency f = B0/2π
for 1H = MHz/T Frequency (MHz) 127.7 63.8 1.5 3.0 Field Strength (Tesla) Turn your dial to 3T fMRI … … broadcasting at a frequency of Mhz

9 Radio-Frequency Excitation
transmission coil: apply magnetic field along B1 (perpendicular to B0 for ~3 ms) oscillating field at Larmor frequency frequencies in range of radio transmissions B1 is small: ~1/10,000 T tips M to transverse plane – spirals down analogies: guitar string (Noll), swing (Cox) final angle between B0 and B1 is the flip angle Transverse magnetization longitudinal axis Source: Robert Cox’s web slides

10 Relaxation and Receiving
Receive Radio Frequency Field receiving coil: measure net magnetization (M) readout interval (~ ms) relaxation: after RF field turned on and off, magnetization returns to normal longitudinal magnetization   T1 signal recovers (realignment) transverse magnetization   T2 signal decays (dephasing) Source: Robert Cox’s web slides

11 Why the dephasing ? protons precess at slightly different frequencies because of random fluctuations in the local field at the molecular level that affect both T2 and T2*; larger scale variations in the magnetic field that affect T2* only. over time, the frequency differences lead to different phases between the molecules (clock analogy) as the protons get out of phase, the transverse magnetization decays this decay occurs at different rates in different tissues Source: Mark Cohen’s web slides

12 T1 and TR T1 = recovery of longitudinal magnetization (B0)
due to realignment of spins TR (time to repetition) = time to wait after excitation before sampling T1 = time before next RF excitation ≈ M0(1-exp(-t/T1)) Source: Mark Cohen’s web slides

13 T2 and TE T2 = decay of transverse (B1) magnetization
due to dephasing of spins TE (time to echo) = time to wait before sampling T2 (after refocusing of signal) ≈ exp(-t/T2)) Source: Mark Cohen’s web slides

14 T1 and T2 contrasts TISSUE T1(s) T2(s) grey matter 1.0 0.10
white matter 0.7 0.08 CSF 2.0 0.25 blood 1.2 water 4.7 3.50 Source: Mark Cohen’s web slides

15 T2* relaxation dephasing of transverse magnetization due to both:
- microscopic molecular interactions (as for T2) - spatial variations of the external main field B (tissue/air, tissue/bone interfaces) exponential decay (T2*  ms, shorter for higher Bo) Mxy Mo sin T2 T2* time Source: Jorge Jovicich

16 Spatial Coding: Gradients
How can we encode spatial position? Add a gradient to the main magnetic field Excite only frequencies corresponding to slice plane Use other tricks to get other two dimensions left-right: frequency encode top-bottom: phase encode Frequency Field Strength ~ z position Gradient switching – that’s what makes all the beeping & buzzing noises during imaging => EAR PLUGS !

17 Echos pulse sequence: series of excitations, gradient triggers and readouts Echos = refocusing of signal Spin echo (not shown) – measure T2 Gradient echo (shown) - measure T2* flip the gradient at t=TE/2 measure after refocusing at t=TE (left-right) (top-bottom) t = TE/2 A gradient reversal at this point will lead to a recovery of transverse magnetization TE = time to wait to measure refocused spins Source: Mark Cohen’s web slides

18 A walk through the K-space
(inverse Fourier transform) Source: Traveler’s Guide to K-space (C.A. Mistretta)

19 Susceptibility Adding a nonuniform object (like a person) to B0 will make the total magnetic field nonuniform This is due to susceptibility: generation of extra magnetic fields in materials that are immersed in an external field Susceptibility Artifact - occurs near junctions between air and tissue sinuses, ear canals - spins become dephased so quickly (quick T2*), no signal can be measured sinuses Source: Robert Cox’s web slides ear canals Aha! Susceptibility variations can also be seen around blood vessels where deoxyhemoglobin affects T2* in nearby tissue

20 Hemoglobin A molecule to breathe with: - four globin chains
- each globin chain contains a heme group - at center of each heme group is an iron atom (Fe) - each iron ion Fe2+ can attach an oxygen molecule (O2) - oxy-Hemoglobin (four O2) is diamagnetic  no B effects - deoxy-Hemoglobin is paramagnetic  if [deoxy-Hgb]   local B  Source: Jorge Jovicich

21 BOLD signal Blood Oxygen Level Dependent signal
neural activity   blood flow   oxyhemoglobin   T2*   MR signal time Mxy Signal Mo sin T2* task T2* control TEoptimum Stask Scontrol S Source: Brief Introduction to fMRI by Irene Tracey Source: Jorge Jovicich

22 BOLD signal Source: Doug Noll’s primer

23 (return to equilibrium state)
To take away Tissue protons align with magnetic field (equilibrium state) RF pulses Protons absorb RF energy (excited state) Relaxation processes Protons emit RF energy (return to equilibrium state) Spatial encoding using magnetic field gradients NMR signal detection Repeat RAW DATA MATRIX Fourier transform IMAGE Magnetic field Kwong et al., 1992 Source: Jorge Jovicich


Download ppt "Basic fMRI Physics In BOLD fMRI, we are measuring:"

Similar presentations


Ads by Google