1
Clinical Research: Sample Measure (Intervene) Analyze Infer
2
A study can only be as good as the data... -J.M. Bland
3
Understanding Measurement: Aspects of Reproducibility and Validity Reproducibility vs validity Impact of reproducibility on validity & statistical precision Assessing reproducibility of interval scale measurements –within-subject standard deviation –coefficient of variation (Section: assessing validity of interval scale measurements)
4
Measurement Scales
5
Reproducibility vs Validity Reproducibility –the degree to which a measurement provides the same result each time it is performed on a given subject or specimen –less than perfect reproducibility typically caused by random error Validity –from the Latin validus - strong –the degree to which a measurement truly measures (represents) what it purports to measure (represent) –less than perfect validity is fault of systematic error
6
Reproducibility vs Validity Reproducibility –aka: reliability, repeatability, precision, variability, dependability, consistency, stability Validity –aka: accuracy
7
Vocabulary for Error Overall Inferences from Studies Individual Measurements Systematic Error Validity (aka accuracy) Random Error PrecisionReproducibility
8
Reproducibility and Validity Good Reproducibility Poor Validity Poor Reproducibility Good Validity
9
Reproducibility and Validity Good Reproducibility Good Validity Poor Reproducibility Poor Validity
10
Why Care About Reproducibility? Impact on Validity Mathematically, the upper limit of a measurement’s validity is a function of its reproducibility Consider a study of height and basketball ability: –Assume height measurement: imperfect reproducibility –If we had measured height twice on a given person, most of the time we get two different values; at least 1 of the 2 values must be wrong (imperfect validity) –If study measures everyone only once, errors, despite being random, will lead to biased inferences when using these measurements (i.e. lack validity)
12
Impact of Reproducibility on Statistical Precision Classical Measurement Theory: –observed value (O) = true value (T) + measurement error (E) –If we assume E is random and normally distributed: E ~ N (0, 2 E ) Fraction error -3 0.02.04.06 Error -20123
13
Impact of Reproducibility on Statistical Precision Assume: –observed value (O) = true value (T) + measurement error (E) –E is random and ~ N (0, 2 E ) Then, when measuring a group of subjects, the variability of observed values ( 2 O ) is a combination of: the variability in their true values ( 2 T ) and the variability in the measurement error ( 2 E ) 2 O = 2 T + 2 E
14
Why Care About Reproducibility? 2 O = 2 T + 2 E More measurement error means more variability in observed measurements –e.g. measure height in a group of subjects. –If no measurement error –If measurement error Height Frequency
15
More variability of observed measurements has profound influences on statistical precision/power 2 O = 2 T + 2 E Descriptive studies: wider confidence intervals Analytic studies (Observational/RCT’s): power to detect a exposure (treatment) difference is reduced truth truth + error truthtruth + error
16
Mathematical Definition of Reproducibility Reproducibility Varies from 0 (poor) to 1 (optimal) As 2 E approaches 0 (no error), reproducibility approaches 1 Note: we can never directly measure this
17
Simulation study looking at the association of a given risk factor and a certain disease. Truth is a risk ratio= 2.0 R= reproducibility Power: probability of estimating a risk ratio within 15% of 2.0 Phillips and Smith, J Clin Epi 1993 Power
18
Sources of Random Measurement Error: What contributes to 2 E ? Observer (the person who performs the measurement) within-observer (intrarater) between-observer (interrater) Instrument within-instrument between-instrument Importance of each varies by study
19
Sources of Measurement Error e.g., plasma HIV viral load –observer: measurement to measurement differences in tube filling, time before processing –instrument: run to run differences in reagent concentration, PCR cycle times, enzymatic efficiency
20
Within-Subject Biologic Variability Although not the fault of the measurement process, moment-to-moment biological variability can have the same effect as errors in the measurement process Recall that: –observed value (O) = true value (T) + measurement error (E) –T = the average of measurements taken over time –E is always in reference to T –Therefore, lots of moment-to-moment within-subject biologic variability will serve to increase the variability in the error term and thus increase overall variability because 2 O = 2 T + 2 E
21
error
22
Assessing Reproducibility Depends on measurement scale Interval Scale –within-subject standard deviation and derivatives –coefficient of variation Categorical Scale –Kappa (see Clinical Epidemiology course) –(can be used for both predictors and outcomes)
23
Reproducibility of an Interval Scale Measurement: Peak Flow Assessment requires >1 measurement per subject Peak Flow Rate in 17 adults (Bland & Altman)
24
Assessment by Simple Correlation
25
Pearson Product-Moment Correlation Coefficient r (rho) ranges from -1 to +1 r r describes the strength of linear association r 2 = proportion of variance (variability) of one variable accounted for by the other variable
26
r = -1.0 r = 0.8 r = 0.0 r = 1.0 r = -1.0 r = 0.8r = 0.0
27
Correlation Coefficient for Peak Flow Data r ( meas.1, meas. 2) = 0.98
28
Limitations of Simple Correlation for Assessment of Reproducibility Depends upon range of data –e.g. Peak Flow r (full range of data) = 0.98 r (peak flow <450) = 0.97 r (peak flow >450) = 0.94
30
Additional Limitations of Simple Correlation for Assessment of Reproducibility Depends upon ordering of data –get different rho depending upon classification of meas 1 vs 2 Measures linear association only –it would be amazing if the replicates weren’t related
31
Meas. 2 Meas 1 1003005007009001100130015001700 100 300 500 700 900 1100 1300 1500 1700
32
Final Limitation of Simple Correlation for Assessment of Reproducibility Gives no meaningful parameter using the same scale as the original measurement –What does rho = 0.7 vs 0.8 vs 0.9 mean in the context of peak flow data which ranges from 200 to 600? –(Note: rho is not “R” from prior slide)
33
Within-Subject Standard Deviation Common (or mean) within-subject standard deviation (s w ) = 15.3 l/min
34
s w : Further Interpretation If assume that replicate results: – are normally distributed – mean of replicates estimates true value 95% of replicates are within (1.96)(s w ) of true value x true value swsw (1.96) (s w )
35
Interpretation of s w : Peak Flow Data If assume that replicate results: – are normally distributed – mean of replicates estimates true value 95% of replicates within (1.96)(15.3) = 30 l/min of true value x true value s w = 15.3 l/min (1.96) (s w ) = (1.96) (15.3) = 30
36
s w : Further Interpretation Difference between any 2 replicates for same person = diff = meas 1 - meas 2 Because var(diff) = var(meas 1 ) + var(meas 2 ), therefore, s 2 diff = s w 2 + s w 2 = 2s w 2 s diff
37
Interpreting s w : Difference Between Two Replicates If assume that differences: – are normally distributed and mean of differences is 0 – s diff estimates standard deviation The difference between 2 measurements for the same subject is expected to be less than (1.96)(s diff ) = (1.96)(1.41)s w = 2.77s w for 95% of all pairs of measurements x diff 0 s diff (1.96) (s diff )
38
s w : Further Interpretation: The Repeatability Value For Peak Flow data: The difference between 2 measurements for the same subject is expected to be less than 2.77s w =(2.77)(15.3) = 42.4 l/min for 95% of all pairs i.e. the difference between 2 replicates may be as much as 42.4 liters just by random measurement error alone. 42.4 liters termed (by Bland-Altman): “repeatability” or “repeatability coefficient” of measurement
39
Interpreting the “Repeatability” Value: Is 42.4 liters a lot? Depends upon the context Clinical management If other gold standards exist that are more reproducible, and: –differences < 42.4 are clinically relevant, then 42.4 is bad –differences < 42.4 not clinically relevant, then 42.4 not bad If no gold standards, probably unwise to consider differences as much as 42.4 to represent clinically important changes –would be valuable to know “repeatability” for all clinical tests Research Depends upon the differences in peak flow you hope to detect –If ~40, you’re in trouble –If several hundred, then not bad
40
One Common Underlying s w Appropriate only if there is one s w i.e, s w does not vary with true underlying value Within-Subject Std Deviation Subject Mean Peak Flow 100300500700 0 10 20 30 40 correlation coefficient = 0.17, p = 0.36
41
Another Interval Scale Example Salivary cotinine in children (Bland-Altman) n = 20 participants measured twice
42
Cotinine: Absolute Difference vs. Mean Subject Absolute Difference Subject Mean Cotinine 0246 0 1 2 3 4 correlation = 0.62, p = 0.001
43
Logarithmic (base 10) Transformation
44
Log 10 Transformed: Absolute Difference vs. Mean Subject abs log diff Subject mean log cotinine -.50.51 0.2.4.6 correlation = 0.07 p=0.7
45
s w for log-transformed cotinine data s w because this is on the log scale, it refers to a multiplicative factor and hence is known as the geometric within-subject standard deviation it describes variability in ratio terms (rather than absolute numbers)
46
Interpretation of s w : Cotinine Data If assume that replicate results: – are normally distributed – mean of replicates estimates true value 95% of replicates within a factor of 0.34 log10 of true value x true value s w = 0.175 log10 (1.96) (s w ) = (1.96) (0.175) = 0.34
47
Interpretation of s w : Cotinine Data 95% of replicates are within a factor of 0.34 log10 of true value back-transforming to base10 scale: –antilog(0.34) = 10 0.34 = 2.2 95% of replicates are within a factor of 2.2 of true value An observed cotinine value of 2 ng/ml would tell us that the true value may be: –as little as 2/2.2 = 0.9 –as big as 2*2.2 = 4.4 –just by measurement error alone
48
Interpretation of s w : Cotinine Data Repeatability The difference between 2 measurements for the same subject is expected to be less than a factor of (1.96)(s diff ) = (1.96)(1.41)s w = 2.77s w for 95% of all pairs of measurements For cotinine data, s w = 0.175 log10, therefore: –2.77*0.175 = 0.48 log10 –back-transforming, antilog(0.48) = 10 0.48 = 3.1 For 95% of all pairs of measurements, the ratio between the measurements may be as much as 3.1 fold (this is “repeatability”)
49
Coefficient of Variation For cotinine data, the within-subject standard deviation (on the native scale) varies with the level of the measurement If the within-subject standard deviation is proportional to the level of the measurement, this can be summarized as: coefficient of variation = = 1.49 -1 = 0.49 At any level of cotinine, the within-subject standard deviation of repeated measures is 49% of the level
50
Coefficient of Variation for Peak Flow Data By definition, when the within-subject standard deviation is not proportional to the mean value, as in the Peak Flow data, then there is not a constant ratio between the within-subject standard deviation and the mean. Therefore, there is not one common coefficient of variation Estimating the the “average” coefficient of variation (within-subject sd/overall mean) is not meaningful
51
Peak Flow Data: Use of Coefficient of Variation when s w is Constant Could report a family of CV’s but this is tedious
53
Assessing Validity Measures can be assessed for validity in 3 ways: –Content validity Face Sampling –Construct validity –Empirical validity (aka criterion) Concurrent (i.e. when gold standards are present) –Interval scale measurement: 95% limits of agreement –Categorical scale measurement: sensitivity & specificity Predictive
54
Conclusions Measurement reproducibility plays a key role in determining validity and statistical precision in all different study designs When assessing reproducibility, for interval scale measurements: avoid correlation coefficients use within-subject standard deviation and derivatives like “repeatability” or coefficient of variation if within-subject sd is proportional to the magnitude of measurement (For categorical scale measurements, use Kappa) What is acceptable reproducibility depends upon desired use Assessment of validity depends upon whether or not gold standards are present, and can be a challenge when they are absent