Presentation is loading. Please wait.

Presentation is loading. Please wait.

Philips PCCI EPIS, Boca Raton, FL, USA IHE Patient Care Devices Domain

Similar presentations


Presentation on theme: "Philips PCCI EPIS, Boca Raton, FL, USA IHE Patient Care Devices Domain"— Presentation transcript:

1 Philips PCCI EPIS, Boca Raton, FL, USA IHE Patient Care Devices Domain
Integrating the Healthcare Enterprise (IHE) Patient Care Devices Domain (PCD) Alarm Communication Management (ACM) For 2011 Connectathon/Showcase IHE PCD 2010 Fall F2F, Plymouth Meeting, PA Monroe Pattillo Philips PCCI EPIS, Boca Raton, FL, USA IHE Patient Care Devices Domain ACM Working Group © 2010 ACCE, HIMSS, IHE

2 ACM Supplement Update New ACM Supplement TI for 2011 Connectathon (not in TF) HL7 now at 2.6 (2.7’s PRT segment imported into TI) PCD-04 – changed from ORU R01 to ORU R40 PCD-05 – changed from ORU R01 to ORA R41 or R42 PCD-04 – PCD-05 fire hose avoided by default, filter not required Re-use of PCD-04 for AM to AA, now PCD-11, still PCD-04 based PCD-06/PCD-07 - WCTP only, SMTP will not be in final TI or TF New Options in Gazelle (uptake status) AR AM – support for PCD-05 (no ARs, only 1 AM) AR AM – support for indicated AC destinations in PCD-04 (no takers) AM AC – WCTP or SMTP (all ACs this cycle indicate WCTP, SMTP out) AR AM AC – WCM within PCD-04 (no set of end to end takers this cycle) 2

3 ACM Transactions and Protocols
Disseminate Alarm PCD-06 → Report Alarm PCD-04 → Alarm Reporter AR Alarm Manager AM Alarm Communicator AC HL7 2.6 WCTP ← PCD-05 Report Alarm Status ← PCD-07 Report Dissemination Alarm Status PCD-08 ↑ Subscribe To Alarm HL7 2.6 PCD-11 ↓ Report Alarm Alarm Archiver AA 3

4 Areas of Improved Validation Scrutiny
Affects all HL7 msgs in ACM PCD 04, 05, 08, 11 MSH Segment for messages & ACKs Sending & Receiving Application Sending & Receiving Facility H7 version identification 2.6 Profile identification IHE ACM PCD-nn Equipment IDs (strong preference EUI-64) 4

5 WCTP for AM PCD-06 & AC PCD-07
WCTP wctp.org constrained by ACM 2010 TI Only a subset of WCTP required Single message (no multi-/group msgs) Multiple Choice Response (MCR) Accept/Reject device response abstraction – AC soft keys AM to map PCD-04 indicated priority to PCD-06 WCTP Permits AC to implement priority ring tones Phone directory not required Presumes AM & AC pre-config of destination devices PCD-07 – web push response, not polled 5

6 ACM WG Co-Chairs Google Group (IHE-PCD-ACM) Weekly WebEx
Monroe Pattillo John Rhoads Google Group (IHE-PCD-ACM) Weekly WebEx 4 PM ET (Manny’s WebEx schedule)