1 the RASNIK Alignment System

2 Particle Physics CERN, Geneva, Swiss

3 pp collisions 2) heavy collisions: A proton is a bag filled with quarks en gluonen

4 The ATLAS Experiment CERN, Geneva, Switzerland ‘Tracking’ of charged particles Measurement of position of tracks Track curvature: measure for momentum & energy

5 Momentum Measurement of charged particles in the L3 experiment: Chamber Position Monitoring Lorentz Force: Track Curvature measurement Detector Muon Particle Track Detector

6 Principle of CCD-RASNIK Coded Mask Light SourceLens CCD

7

8 Title: The RASNIK opto-electronic alignment system: a high-precision, large range, fast and zero-drift monitor for displacements or deformations. In RASNIK, the image of a (back-illuminated) coded mask is projected, by means of a lens, onto a (pixel) image sensor. A displacement of the mask, or lens, or sensor, relative to the other two components, results in a displacement of the mask’ image on the sensor. This can be registered accurately by means of a processor connected to the (USB) sensor. Displacements in the two transversal directions cause an image shift, and a displacement in the direction of the optical axis results in a change of the image scale. In addition, the relative rotation around the optical axis of mask and sensor can be recorded, making RASNIK a 4D measurement device. With image frame rates up to 100 Hz, vibrations can be measured as well. With RASNIK, the bending of a (roof) bar can be monitored accurately. When the lens and sensor are coupled on one (CAM) base plate, the displacement of the mask with respect to this base plate is measured; this ‘proximity’ RASNIK is applied as displacement monitor for adjacent tunnel sections. The deformation of a complete tunnel could be measured by mounting a series of identical plates, each carrying a mask, lens and sensor, forming a chain of coupled RASNIK systems.

9 dX, dY: Image Displacement

10 dX LED = -2 dX LEN = dX SEN Rasnik 3-point alignment system Alternatives: - Taylor Hobson telescope - Stretched wire system: electronic version after 1985

11 Measurement & Precision - Translation (X, Y): 50 nm per image - Scale: 1.0000 +/- 0.00001 - Rotation around Z-axis: 0.1 mrad Number of images: depending on pixel sensor: - webcam: 30 – 60 images per second: measurement of vibrations! - special graphic image sensor: 10.000 images/s Practical limitation: Temperature gradient in air dT Image info ~ 1Mb is converted into only 4 parameters

12 ATLAS Muon Chambers Image Sensors Lenses Light Sources RASNIK ‘In-Plane’ systems Measures: Chamber sag Chamber torque Temperature gradients

13

14 RASNIK systems in the ATLAS Muon Spectrometer Projective Axial Proximity ‘Praxial’

15 Applications 4 D, no-contact Dial Gauge lensImage sensor mask laptop

16 Weena Rasniks Measurement of relative displacement of adjacent sections

17

18 Measurement of (variation of) pillar height

19 Intergral measurement, in 3D, of deformation of large (long) object: - tunnel - bridge - RasChain plate includes light source, lens and image sensor - mount RasChain plate at ~ 10 m pitch, over 1 km - readout chain at both ends - Deformation is measured with mm precision! RasChain

20 Segment 2 m Laser Diffraction plate (hole) Image sensor RasChain

21

22 Laser Diffraction plate (hole) Image sensor Microcontroller Bus and Power µC Link datapower ID address

23 Level 0 Cameras on a bus link

24 Level 0 Level 1 Chain Chain Control Next chain ~256m ~128m Communication layer

25 RasChain measuring the integral deformation of a long object, i.e. a tunnel

26 Fig. 1 The leap-frog Rasnik system. All plates are identical and each include an illuminated coded mask, a lens and an image sensor.

27 Position resolution with Gaussian noise on Rasnik data The noise per Rasnik system is rather arbitrary. With direct shadow images, 50 nm has been achieved (image position on sensor: X R and Y R ). With RasDif, 20 nm has been reached (over 140 m!). If images of a static system are combined, even lower values are reached. The lower limit is hard to measure due to the presence of systematic image shifts due to non-homogeneity of the ambient medium, causing both a random and a systematical error. Assuming a random Gaussian error in X R of 50 nm in all of the 100 Rasnik systems, the resulting errors in the monitored plate positions is shown in fig.2. As expected, the uncertainty is the largest in the middle of the RasChain. The value, however, is in the order of 10 µm and small enough to be relevant for the presence of long-distance alignment systems such as (long) RasClic or the stretched wire system.

28 Fig. 2. The random error as a function of plate number, due to a Gaussian error of 50 nm on the Rasnik data (common for all 100 systems).

29 Laser zone lens hole dia. 50 mm RasCam 100 m (vacuum tube!) RasDif: replace lens by diffraction plate: just a round hole!

30 Laser expanded beam just monochromatic light source diffraction plate hole dia. 50 mm RasCam RasDif long baseline: lens becomes unpractical Replace lens by ‘diffraction plate’: just a hole!

31 Image position on sensor. Response of earthquake in Mid-Atlantic, 5 Richter Scale, on March 1, 2007 Rasnik as seismic sensor

32

33 RasNap Air-refraction corrected telescope Practical limitation: Temperature gradient in air

34 Rasnik: a new displacement monitor - based on a wide and 27 years long experience - very precise: - high data rate: dynamical measurement - no drift in measurement: monitoring of slow motions - simple, digital, robust & low-cost - a new means of product parameter verification But: - needs 220 V and Ethernet (compare old t, P-sensor: plot) - custom/case-specific application (use ‘standard’ components) - interpretation of data: skilled, educated personel