Preprocessing John VanMeter, Ph.D. Center for Functional and Molecular Imaging Georgetown University Medical Center.

Similar presentations


Presentation on theme: "Preprocessing John VanMeter, Ph.D. Center for Functional and Molecular Imaging Georgetown University Medical Center."— Presentation transcript:

1 Preprocessing John VanMeter, Ph.D. Center for Functional and Molecular Imaging Georgetown University Medical Center

2 FixationThumb movement time Recall Analysis of fMRI Data is Based on Examining Changes in Voxel Across Time

3 Basic Preprocessing Chain

4 Slice Timing Correction fMRI data typically acquired using 2D acquisitions (i.e. one slice at a time) Order of slices can be acquired in different ways –Ascending or descending (slice order is sequential) –Interleave: even numbered slices acquired then odd numbered slices or vice versa Order of slice acquisition determines when in the hemodynamic response the signal for a slice is acquired

5

6 Slice Timing Correction Uses temporal interpolation to make it appear as though all of the slices were acquired at the same time Thus, HRF across slices are aligned Interpolation will be most effective if the sampling rate (time between slices) is higher than the changes in the data Generally more effective for short TR (1-2 secs) than larger TR (>3 secs)

7 Order of Slice Timing Correction and Motion Correction Slice timing first if data acquired interleaved –Small changes in head position between slices will cause timing differences to be off by 1/2 a TR Motion correct first if ascending or descending or short TR Slice timing correction generally unnecessary for block designs regardless of slice acquisition order and TR

8 Effect of Head Motion

9 Example of Ringing

10 Minimizing Head Motion Use of various types of restraints (padding, bite bar, vacuum cushion, face mask) Mock scanner training –Accustomization to environment –Reinforcement systems Impress importance of not moving on subjects

11 Motion Correction (aka Realignment) Coregister each volume in the run to a reference volume Possible reference volumes –First volume in the run –Middle volume in the run –Average of all the volumes in the run before motion correction No consensus on which choice of reference volume is best but typically first volume is used

12 Target Reslice

13 Surface Registration

14 Coregistration Algorithm Components Four components are needed for any coregistration algorithm –Model that defines the set of transformations (eg. translations, rotations, scaling, etc) used –Search algorithm to test different combinations of the model parameters to find the “best” set –Cost Function that provides estimate of how well registered the two volumes are –Interpolation algorithm to reslice/reformat and create a new volume based on the best model parameters

15 Transformation Described by number of ways that one volume can be changed to match another (degrees of freedom) Rigid-body transform allows for translation in all 3 directions and rotation about all 3 axes (6 degrees of freedom) Higher transforms could be used but for motion correction rigid-body transform is most common and appropriate

16 Search Algorithm Systematic method for searching for the best combination of transformation parameters that aligns the reslice volume to the target volume Cost function – provides measure of how closely aligned two images are

17 Standard Deviation of the Ratio = ? Standard Deviation of ratio of all brain voxels = 0 ResliceTargetRatio Image

18 Standard Deviation of the Ratio = ResliceTargetRatio Image Standard Deviation of ratio of all brain voxels >> 0 or

19 Reslice Target y x z 10 o

20 Image Interpolation Reslice

21

22

23 Interpolation Methods Nearest Neighbor –voxel intensity in new volume equal to the closest voxel in the original volume Trilinear –voxel intensity in new volume is based on weighted average of surrounding voxels from the original volume –weights based on distance from the new voxel

24 Reslice (Rz = -10 o ) Target Nearest Neighbor Trilinear

25 Original 90 o 180 o 360 o 10 o 20 o

26 SincWindowed Sinc

27 Original 10 o 20 o 90 o 180 o 360 o

28 Quality Assurance: Motion Correction Plots Examine the amount of motion in each run by looking at plots of motion Big spikes in motion plots problematic Run motion correction twice to see how much residual motion there was from first application of motion correction

29 Pre-motion correction Post-motion correction

30 How Much Motion is Acceptable? Examine residual motion correction plot for the maximal amount of motion in each dimension Ideal –Less than 0.5-0.75 mm –Less than 0.5 o Patient population dependent –Children, rare diseases, hard to recruit subject, etc –Go with what you got

31 Distortion Correction Geometric distortion occurs in EPI sequences arising from inhomogeneity in static magnetic field Shimming adjusts for distortions due to head and body being in the field Automatic shimming generally available though not always sufficient to prevent distortion On our scanner geometric distortion is very minimal

32 Susceptibility Artifacts Ojemann, NeuroImage 1997

33 Examples of Geometric Distortion sinuses ear canals Poor ShimmingOur Scanner With Automatic Shimming

34 Geometric Distortion in Our Scanner Distortions

35 Field Mapping Collect two short scans at two different TE’s to determine differences in the phase component of the signal (difference is the field map) Used to correct distortion

36 Coregistration: Registering Functional Data to Structural Scan Registering functional and structural data useful for both looking at single subject results (e.g. overlaying statistical results onto that subjects structural scan) Also used as precursor to warping subject into standard or atlas space for multi-subject analyses

37 Spatial Normalization Variations between individual brains is large: –Overall size varies from 1100 cc to 1500 cc (30% variation) –Major landmarks such as central sulcus will vary considerable in position, branching, and length Spatial normalization warps individual brains into a common reference space Allows for examination of fMRI signal changes across individuals within a group or between groups of subjects Most commonly used reference space is based on the Talairach atlas

38 Talairach Atlas Atlas of brain anatomy Developed by two French neurosurgeons Jean Talairach and Pierre Tournoux Defines a coordinate system used for stereotaxic reference Stereotaxic space is a “precise” mapping of a brain into 3D coordinate system The Talairach atlas has become the de facto standard in Neuroimaging

39 Talairach Atlas Originally used to for surgical planning for deep subcortical structures in the brain (e.g. basal ganglion, thalamus, etc.) Based on the brain of 69 YO French female Numerous stories about the preparation of this woman’s brain for sectioning

40 Coronal Atlas Sections

41 Sagittal Atlas Sections

42 Disadvantages of Talairach Based on a single cadaver brain and not representative of younger living brains Left-Right hemispheric differences are ignored (assumed symmetric for spatial normalization) Cerebellum completely ignored Notorious problems (e.g. occipital lobe is much smaller in atlas than most brains) Non-European brains (e.g. Asian) do not fit well within the Talairach space

43 Talairach Coordinate System Origin located at the Anterior Commissure (AC) Three axes define coordinate system: –Axial plane parallel to the line passing through the AC – PC –Sagittal plane parallel to the AC – PC line –Coronal plane perpendicular to other two

44 Talairach Coordinate System X-axis goes from the Left Hemisphere to the Right Hemisphere. Positive X coordinate is Right Hemisphere. Range: -75 to +75mm Y-axis goes from the Anterior extent to the Posterior extent. Positive Y coordinate is anterior of the AC. Range: -105 to +70mm Z-axis goes from the Superior extent to the Inferior extent. Positive Z coordinate is above the plane passing through the AC-PC. Range: -40 to +65mm

45 Automated Talairach Registration Transform individual brains into a standard space – typically Talairach Space Colloquially referred to as Talairach’ing Formally referred to as Spatial Normalization Same components as motion correction – mathematical model, search algorithm, cost function, image interpolation

46 Spatial Normalization Register image of an individual brains to match a template image in a standard space, typically the Talairach atlas Linear models incorporate parameters for scale and skew (12 degrees of freedom) Nonlinear models use either polynomials or cosine functions to model warping Higher orders provide greater accuracy of fit

47 Templates As with the French woman’s brain, there are numerous stories about the various templates in use Several of the templates have been developed in an ad hoc and poorly documented fashion

48 T1-MRI Template (MNI 305)

49 EPI Template (FIL)

50 Methods of Template Creation Manually transform each individual brain into Talairach space using linear stretching then average - MNI 305 done this way with 305 young volunteer’s T1 MRI scans after Spatially normalize each subject to match an existing template then average Round-Robin registration of each subject to every other subject over multiple iterations, averaged then transformed into atlas space

51 Comparison of Template Creation Methods Round Robin Register to Existing

52 6-8 yrs 14-16 yrs 32 yrs Pediatric Templates

53 Nonconformity of Templates with Talairach Atlas: MNI vs Talairach MNI 305 does not exactly match the atlas –MNI 305 is bigger in all three dimensions –Near center of the brain MNI 305 matches the atlas best –Towards the extents larger differences –Variations are not uniform, overall MNI 305 is ~5mm taller in the superior-inferior dimension however parts of the temporal lobes extend ~10mm lower

54 Accuracy of Talairach Registration Methods Classical Talairach method –Good alignment of subcortical structures –Cerebral structures substantially less well aligned –Subjective since landmarks especially PC and to lesser extent AC are difficult to identify Automated Spatial Normalization –Cerebral structures are aligned better –Overall alignment is better –But MNI and ICBM templates don’t conform to Talairach -> MNI vs Talairach space

55 Classical Talairach Method vs Nonlinear Spatial Normalization Classical Talairach Automated Nonlinear

56 Classical Talairach Method vs Nonlinear Spatial Normalization Classical Talairach T1 TemplateNonlinear Transformation

57 Converting MNI to Talairach: Piecewise Linear Above the AC (Z >= 0): –X'= 0.9900X –Y'= 0.9688Y +0.0460Z –Z'= -0.0485Y +0.9189Z Below the AC (Z < 0): –X'= 0.9900X –Y'= 0.9688Y +0.0420Z –Z'= -0.0485Y +0.8390Z

58

59 Converting MNI to Talairach: Nonlinear http://www.bioimagesuite.org/Mni2Tal/index.html Atlas Piecewise-Linear correction Non-linear Correction

60 Comparison of Different Linear Models

61 SPM5 Spatial Normalization Method Co-register mean EPI image to subject’s MPRAGE (or other high resolution structural) Segment MPRAGE by tissue type (i.e. CSF, gray, and white matter) Spatially normalize subject’s gray matter map to template gray matter map Apply both transforms to each EPI volume in the run

62 SPM5 Spatial Normalization - Good Idea But… Structures of interest in fMRI are in gray matter Matching gray matter to gray matter bound to be an improvement But it relies on accuracy of co- register step

63 Coregistering EPI to MPRAGE Can be problematic Ratio Image EPI acquired with a FatSat pulse - thus scalp eliminated (good for fMRI; reduces ghosting) MPRAGE has bright fat signal from scalp Coregistration can get thrown off and try to fit edges of EPI to bright scalp instead of bran

64 Smoothing Form of a more general image processing technique called convolution Neighborhood operation New pixel value based on weighted sum of a pixel and its neighbors A convolution operator is referred to as a kernel

65 Average Kernel 111 111 111

66 111 111 111 32 * = ?

67 1 * = ? 111 111 111 28

68 A kernel is defined by its weights and size 111 111 111 111 111 111 111 111 111 111 111 111 111 111 111 3x3 5x5

69 Gaussian Kernel Weights in Gaussian kernel defined by normal distribution Gaussian kernel specified by Full-Width at Half Maximum (FWHM)

70 2D Gaussian Kernel

71 1.071.141.07 1.161.14 1.07 FWHM ~ 12mm 2

72 Example Reduced Noise Using Smoothing

73 Original PET Image 12mm 3 6mm 3 24mm 3

74 Gaussian Smoothing Advantage – Removes isolated points of noise in an image – Compensates for imperfect spatial normalization – Helps to prepare data to meet statistical analysis requirements – Reduces false positives in statistical analysis Disadvantage – Reduces the spatial resolution of the image; looks blurrier – Peaks are squashed Ideal filter size is one that matches the size of the areas of activation (Matched Filter Theorem) - generally unknown and varies across brain General rule of thumb for fMRI data is to use a filter width that is twice the voxel size

75 Smoothing & Statistical False Positives Random dataset left is unsmoothed and right is smoothed

76 Smoothing and SNR Smoothing actually increases the ability to detect task-related signal changes (i.e. functional SNR increases) Cost is of course reduced spatial resolution Relies on spatial correlation of fMRI data (i.e. neighboring voxels have similar intensity and activation characteristics)

77 Temporal Filtering Similar to spatial filtering expect the data are processed across time Some sources of temporal noise –Transient jumps and drops in the global signal from timepoint to timepoint (ie. overall brightness fluxuates) –Drift in MRI signal over time –Head movement

78 Global Normalization Global signal is defined as the average intensity of all the voxels in the brain Variations in the global signal can swamp the regional changes in the MRI signal that we are trying to measure Ratio correction divides the intensity at each voxel by the global signal in that scan and multiples by a constant ANCOVA is an alternative that allows for differences in variations based on trial type or condition

79 Ratio Normalization

80 Signal Drift Across Time Gradually, over time the MRI signal in a voxel drifts This drift can vary from one voxel to the next A simple model of this drift is as a linear rise or fall in the signal

81 Linear Detrending

82 High-Pass Filtering Similar to linear detrending but higher order variations are also removed Types of signal variations that can be corrected are more complex All of the signal below a certain frequency is removed.

83 Non-linear Drift

84 Effect of Slow Non-linear Drifts

85 High-Pass Filtering Nyquist Sampling Theorem – removing signal fluxuations that are greater than half the frequency of the task potentially removes signal of interest –Thus, cut-off frequency should be half the task frequency For block designs the cut frequency can be determined from the periodicity of your paradigm The cut-off is specified in time in SPM –Thus, combined time of the on- and off-blocks times 2 is the cut-off for a block design For event related designs there is no simple formula; using a cut-off that is a ~2 minutes long is a safe bet

86 High-Pass Filtering

87 Band-Pass Filter Remove signal changes that are both higher and lower than task frequency in a block design Cut-offs would be set to 2x and 1/2x the task frequency Advantage is removal of both signal drift (high- pass) and cardiac, respiration signal changes (low-pass)

88 Low-Pass Filtering Typical Task Frequency

89 Quality Control in fMRI Very important to examine your data after each stage of preprocessing and statistics –Look at raw data for artifacts –Examine realignment plots –Examine mean & variance images before and after realignment (ana4dto3d toolbox for SPM) –Look at movies of time series before and after realignment (MEDx or AQuA toolbox for SPM) –Examine plots of image-to-image signal variance (tsdiffana toolbox for SPM) –Examine how well spatial normalization worked (use Check Reg in SPM) –Look at single subject statistics for ringing (best with task vs. lowest control condition) –Do statistics make sense (eg. if motor response used is the appropriate motor cortex activated?)

90 fMRI Raw Data Artifacts Spikes - caused by transient electrical discharges during scanning; something is leaking RF noise into the scanner room Ghosting - Nyquist reconstruction artifacts can sometimes be limited by choice of certain acquisition parameters (eg. echo spacing)

91 Ghosting on Our Scanner On our scanner adjusting echo spacing to be outside of 0.7-0.8ms minimizes ghosting Ghosting Artifact

92 BIG Technique Experimental Design Behavior Interleaved Gradient (BIG) technique (aka Sparse Sampling) puts a fixed delay between each EPI acquisition During this break auditory stimuli can be presented without noise from gradients Subjects can verbally respond with minimal effect on motion in the EPI image Relies on delay in hemodynamic response Also called Sparse Sampling

93 BIG Acquisition Requires accurate timing of the task relative to the the scan acquisition to capture peak BOLD response in the volume Greatly increases scan time


Download ppt "Preprocessing John VanMeter, Ph.D. Center for Functional and Molecular Imaging Georgetown University Medical Center."