Wavefront by Method of Ray Tracing: Because the Eye is Not a Telescope
Joe S. Wakil, M.D. Chairman, Tracey Technologies, LLC and EyeSys Vision, Inc. Speaker has a Financial Interest
2
Founding Technology Developers:
Vasyl Molebny, DSc Kiev, Ukraine Ioannis Pallikaris, MD Crete, Greece Canadian & Swedish Governments
3
Why Dynamic Ray Tracing? Because… the Eye is NOT a Telescope!
Pupil Size: Accommodation: Tear Film: Effects Refraction (ex. Night Myopia) Refraction is NOT a FIXED Number! Effects Refraction (Instrument Myopia) Effects Quality of Vision (Aberrations) Where are the Sources of Aberrations? How do they change with Surgery? Cornea: Lens: Astigmatism (Irreg.), Sph & other HOA Astigmatism, Coma & other HOA
4
The Eye is NOT a Telescope
Off-Axis design No magnification Variable aperture Variable detector res. Accommodation Changing fixation Brain image processing Nature-made TELESCOPE On-Axis design High magnification Fixed aperture Constant detector res. No accommodation Fixed alignment Digital image processing Man-made
5
Aberrometer/Wavefront Technologies
Hartmann-Shack Lenslet Array Tscherning Aberrometer Differential Skiascopy Ray Tracing Features: -Rapid, point by point, IR measurement no data confusion -Pupillometry with auto-tracking/capture -Programmable sampling (256 pts.) in any pupil up to 8mm -Open Field Fixation – avoid instrument myopia and measure Accommodation -Corneal Topography integration – able to measure Lens Aberrations -VISX license for Custom LASIK
6
Hartmann-Shack Wavefront Sensor
H/S Photo of patient with tight eye lid courtesy David Williams
7
Principles of Tracey Programmable thin beam ray tracing measuring forward aberrations of the eye Rapid sequential measurement of data points over entire entrance pupil (<50ms) Localization of each reflected retinal spot Integration of individual retinal spots to form Point Spread Function (PSF) Analysis of PSF for higher order aberrations and other data formats
8
Data Sample Points Data Sample Points
9
Multiplying the Number of Sites
10
Higher Local Density of Sites
11
Overlay of Two Sets of Site Configuration
12
Ray Tracing Objective lens Input Beam Retina Eye optics Position
sensitive detector
13
Ray Tracing Objective lens Retina Eye optics Position sensitive
detector
14
Ray Tracing Objective lens Retina Eye optics Position sensitive
detector
15
Ray Tracing Objective lens Retina Eye optics Position sensitive
detector
16
Ray Tracing Objective lens Retina Eye optics Position sensitive
detector
17
Ray Tracing Objective lens Retina Eye optics Position sensitive
detector
38
Refractive Error Measurements
Myopia Hyperopia
39
Retinal Spot Diagram/Point Spread Function
40
Data Displays Retinal Spot Diagram Refraction Map Ablation Map
X, µm 20 10 -10 -20 Y, µm 40 30 Refraction Map Retinal Spot Diagram Wavefront Map Ablation Map
41
Tracey Analysis Conoid of Sturm Evaluation Contrast Sensitivity Data
Depth of Field Analysis Individual Data Point PSF’s Zernike Calculations
62
Diagnostic Capabilities
Determine Higher Order Aberrations True Point Spread Function Presentation Optical Media Opacity Mapping Integrated Corneal Topography
63
Validation Studies Three independent studies of
Tracey vs. Manifest Refraction Koch et al - 58 eyes Slade et al - 42 eyes Schalhorn et al eyes Results Accuracy to manifest <0.12 D Reproducibility <0.12 D
67
Post LASIK
68
Post LASIK
69
Post LASIK
70
Keratoconus
71
Normal
72
Normal
73
Accommodating IOL Change due to Accommodation Near Fixation
Far Fixation
74
Advantages of Ray Tracing Aberrometry Combined with Corneal Topography - iTrace
Robust dynamic range to measure highly irregular eyes including immediately post-op Multi-Zone Refraction in seconds to measure variations in refraction from normal night myopes to post-op results saving chair time Able to separate Corneal from Lenticular (internal ocular) sources of aberrations with EyeSys Corneal Topography Improved Diagnostic Wavefront to help in patient selection including Custom LASIK and IOL procedures
75
Conclusion The Tracey iTrace brings a new standard of care in completely assessing vision in terms of quality of vision (aberrometry) and corneal topography to progress visual correction diagnosis and treatments to its ultimate goals.