DMC Calibration Klaus NeumannProduct Manager
2
Page 2 DMC - Camera Head Production Raw Die Pan CCD Certification - Burn In & Test Flight DMC Assembly Lens assembly Assemble Optics, Sensor & Electronics Validation Adjustment of CCD and Optics Radiometric calibration Geometric calibration Calibrated Camera Heads Delivery
3
Page 3 Zeiss Jena Clean Room Facilities
4
Page 4 Internal Calibration DataCalibration CD homogeneous virtual image radiometry mosaicking geometric quality assessment Initial EO for all 8 camera heads Adjusted, homogeneous pixel sensitivity Color Balance Interior Orientation Data PRC PPS S/W Component DMCORI PPS S/W Component DMC Post Processing Ulbricht SphereCollimator RadiometryOnline PlatformGeometryRadiometryGeometry PlatformSingle Head DMC Calibration - Overview
5
Page 5 Geometric Calibration Designed for DMC Rotationtable with 0.3 mgon accuracy Circular Targets Fully automatic Zeiss Jena
6
Page 6 Radiometry Calibration with Ulbricht Sphere
7
Page 7 DMC Environmental Tests RTCA / DO 160 E (= ISO 7137) EN 61000-6-2:2001Electromagnetic compatibility (EMC) Part 6-4: Generic standards Section 2: Immunity for industrial environments Section 4: Emission standard for industrial environments Part 4: Testing and measurement techniques Section 2: Electrostatic discharge immunity test Section 5 : Surge immunity test EN55011:1998 Industrial, scientific and medical (ISM) radio-frequency equipment -Radio disturbance characteristics -Limits and methods of measurement
8
Page 8 DMC Environmental Tests Vibration -Sinus:DO 160 E – Section 8 (Vibration) Figure 8-3 Category M (5 -500 Hz) (Standard Sinusidal Vibration Test curves for equipment installed in fixed wing aircraft with reciprocating or turbo propeller engines) -Random:ISO 9022-37 Category 21 (1,4 g rms @ 20-2000 Hz) Shock 15 g @ 11ms according ISO 9022-30 Category 01 Temperature -Temp-Burn In according own specification: Operating Temperature: 0 – 40°C T1 = 30min T2 = 4h T3 = 30min (In minimum 4 cycles) -Storage Temperature (according own specification): DMC Camera -20°C – 70°C (T1 = 1h T2 = 48h T3 = 1h) All other Equipment -40°C – 70°C (T1 = 1h T2 = 48h T3 = 1h) Pressure -Under pressureRTCA / DO 160 E – Section 4 (Temperature & Altitude) Category C1 10700 m (guaranteed for customers only : 8000m) T0 – T1: 30min T1–T2: 0 T2–T3: 4h Temperature 20°C
9
Page 9 Factory Acceptance and Certification –Factory acceptance Test 1st images taken and processed Test procedure includes GPS simulated flight – Test Flight Full Mission -> Test field “Elchingen” 500m flying height –5cm GSD 8 control points –AT is run to determine platform calibration
10
Page 10 Standard DIN 18740-4 Geometric Calibration Protocol Geometric Calibration Protocol for each camera module or spectral channel Alignment of the IMU to the focal plane For mosaicked images the relative orientation of the cameras for each exposure
11
Page 11 Noch Fragen ? Klaus Neumann Z/I - Earth Imaging Solution Center www.intergraph.com Thank you !