CLINITEK Status® Plus UACR Operator Training Program

Similar presentations


Presentation on theme: "CLINITEK Status® Plus UACR Operator Training Program"— Presentation transcript:

1 CLINITEK Status® Plus UACR Operator Training Program

2 Learning Objectives This training module provides a comprehensive overview of the Clinitek Status® Plus meter features and operating procedures including: System Overview System Setup Quality Control Participant Testing Cleaning M Investigator Training | | Company Confidential © 2013

3 System Overview Microalbumin product is useful to test for microalbuminuria in patients with diabetes or hypertension in order to detect early kidney disease Albumin, creatinine, and albumin-to-creatinine ratio results to help identify clinically relevant microalbuminuria Results reported in 60 seconds Reportable range: Microalbuminuria is indicated at a ratio result of 30–300 mg/g Touch screen A/C Power or Rechargeable battery (Sonar study will use A/C power only) M Investigator Training | | Company Confidential © 2013

4 The CLINITEK Status® Plus Device
System Overview The CLINITEK Status® Plus Device 1. Analyzer 2. Power supply adaptor and AC power cord (Figure shows US version) 3. Test table with calibration bar 4. Test table insert 5. Paper roll M Investigator Training | | Company Confidential © 2013

5 Getting Started LOCATION
The Analyzer should be on a level work surface in a temperature and humidity controlled environment Ideal temperature: 22o–26oC (72o–79oF) POWER SUPPLY Connect the power supply to the power inlet located on the back left hand side of the Analyzer and connect the other end into an AC electrical wall outlet Country specific adapter not included, available separately M Investigator Training | | Company Confidential © 2013

6 Getting Started INSERT TEST STRIP TABLE
Hold table by the end opposite of the white calibration bar. DO NOT TOUCH the white calibration strip Insert with the white bar facing UP into the port on the front Push into analyzer JUST OVER HALF-WAY. DO NOT PUSH TABLE FULLY INTO ANALYZER INSERT TEST TABLE INSERT Place appropriate side of insert into strip table M Investigator Training | | Company Confidential © 2013

7 Getting Started PRINTER PAPER Turn analyzer to back
Open the printer compartment. This reveals the paper feeder. Pull up on the grey lever on the LEFT side of the printer compartment Push down on the grey tab to expose the paper holder Place the paper into the grey paper holder so the paper is fed out the bottom of the roll, and away from you. (Paper should be up the back wall of compartment). Feed the paper through to expose about 4 inches Push the grey lever on the left down Feed the paper though the printer compartment door and close the printer compartment door making sure it clicks back into place Paper holding arm Printer paper M Investigator Training | | Company Confidential © 2013

8 Getting Started Video M Investigator Training | | Company Confidential © 2013

9 Initial Setup POWERING ON
Press the Power button located on the front of the Analyzer to turn on your Analyzer. The Analyzer will perform a System Test each time it is turned on. During the initial start-up, the Analyzer will prompt you to select a language. Select the language and press Next. This will bring you to the Start-up Wizard. This screen will show you what the Wizard will help you set up. Click Next to begin the Set-up. Select proper Date: Day/Month/Year Time: 24 hours Preferred Test Sequence: Select Quick Test Type of Strip Select “Multistix 10 SG” Results Format: Conventional Done will bring you to a Confirmation Screen where you can verify your selections Press Next and it will bring you to your main <Select> screen where QC/Patient tests can be run, system settings can be accessed, and results can be recalled M Investigator Training | | Company Confidential © 2013

10 Initial Setup Regular Start up: For all future startup…
The Analyzer will perform a system test each time it is powered up Upon successful start-up, the analyzer will take you directly to the main screen. Title Bar will say <Select> & Ready along with a Time/Date Stamp. The main <Select> screen displays with options: Instrument Setup Recall Results Cassette Test (Not applicable for Sonar Study) Strip Test: Select this option to run Liquid QC and UACR tests M Investigator Training | | Company Confidential © 2013

11 Initial Setup Video M Investigator Training | | Company Confidential © 2013

12 Consumable – Microalbumin 2 Test Strip
25 strips per vial Expiration date is printed on the vial of test strips Store strips in their original vial at room temperature between 15o–30oC (59o–86oF) Keep tightly closed when not in use Do not use product after expiration date Do not store the bottle in direct sunlight M Investigator Training | | Company Confidential © 2013

13 Consumable – Quality Control Solutions
Two levels of Quality Control: Level 1 and Level 2 IMPORTANT: Daily routine when conducting patient testing Conduct two levels of Quality Control testing (requires 2 strips) once the patient visit is confirmed for that day Patient testing is valid for 24 hours following Quality Control Repeat Quality Control on next confirmed patient visit day 18-month shelf life and open vial stability from date of manufacture when stored at 2o– 8oC Keep the vials tightly closed when not in use Expiration date is printed on the control vials M Investigator Training | | Company Confidential © 2013

14 Conducting Quality Control
Collect materials needed to conduct Quality Control Two (2) Test strips (1 for each control level) Level 1 and Level 2 Quality Control solutions (Take out of refrigerator 15 minutes before use) Paper towel for blotting Touch Strip Test on the main <Select> screen Select “Quick Test” The Analyzer will proceed to the <Prepare Test> Screen Click Start on the <Prepare Test> screen. The Analyzer begins an eight (8) second countdown for you to: Apply Level 1 Quality Control solution to the strip pads Blot excess liquid from strip Insert strip on tray with test pads facing up NOTE: DO NOT lay the pads on the paper towel or cover the pads with the paper towel Step 1 Step 2 M Investigator Training | | Company Confidential © 2013

15 Conducting Quality Control
The test table and strip will automatically be pulled into the analyzer at the end of the 8 second countdown <Analyzing in progress> screen will automatically begin at the end of calibration and a countdown until result will be displayed At the conclusion of the testing countdown, <Results> will be displayed on the screen Select Print to print the results Click Done to return to the main <Select> screen and remove used urinalysis strip and discard in a biohazard container Repeat steps two (2) through six (6) for Level 2 Quality Control testing Confirm that the QC results are acceptable by comparing with range on QC vials M Investigator Training | | Company Confidential © 2013

16 Conducting Patient Test
Collect materials needed to conduct Quality Control Test strip Patient urine sample Collect specimen in a dry, covered container free of detergents or contaminating substances and urine should be at least 3 inches in depth Specimen should be well mixed and un-centrifuged Paper towel for blotting Touch Strip Test on the main <Select> screen Select “Quick Test” The Analyzer will proceed to the <Prepare Test> Screen Click Start on the <Prepare Test> screen. The Analyzer begins an eight (8) second countdown for you to: Dip the reagent strip into the urine sample making sure to wet all the pads Blot excess liquid from strip Insert strip on tray with test pads facing up NOTE: DO NOT lay the pads on the paper towel or cover the pads with the paper towel Step 1 Step 2 M Investigator Training | | Company Confidential © 2013

17 Conducting Patient Test
The test table and strip will automatically be pulled into the analyzer at the end of the 8 second countdown <Analyzing in progress> screen will automatically begin at the end of calibration and a countdown until result will be displayed At the conclusion of the testing countdown, <Results> will be displayed on the screen Select Print to print the results Click Done to return to the main <Select> screen and remove used urinalysis strip and discard in a biohazard container Record patient results per protocol M Investigator Training | | Company Confidential © 2013

18 Conducting a Test Video
M Investigator Training | | Company Confidential © 2013

19 Cleaning Clean the meter on a routine basis Cleaning Test Table Insert
Remove the Test Table Insert and rinse both sides under running water Dry and replace insert Cleaning Exterior of device Turn Analyzer off by pressing on on/off button for 2 seconds Wipe the outside of the unit with a cloth dampened (not wet) with a mild detergent Allow unit to air dry and turn on M Investigator Training | | Company Confidential © 2013

20 Complete the QC Confirmation Step
Go to Enter your Information on the form Submit You will receive an confirmation You will only have to complete this form once! M Investigator Training | | Company Confidential © 2013

21 Questions Thank You for Your Participation!
M Investigator Training | | Company Confidential © 2013