1 New stent technology Magnesium-alloy The Progress-AMS Study
Raimund Erbel, M Haude, Th Konorza, D Boese Department of Cardiology West-German Heart Center Essen University Duisburg-Essen Erbel et al., Lancet 2007;369(9576):

2 Hypotheses Purpose Design
Clinical Performance and Angiographic Results of the Coronary Stenting with Absorbable Metal Stents The PROGRESS-AMS Study To evaluate the clinical feasibility of an absorbable metal stent in the treatment of a single de novo lesion in a native coronary artery Purpose Prospective, multi-center, consecutive, non-randomized FIM (First In Man – coronary) study Design MACE rate after 4 months <30 % comparable to BMS Hypotheses Erbel et al., Lancet 2007;369(9576):

3 PROGRESS STUDY Principal Investigator Raimund Erbel, MD, Essen, Germany Co-Chairman Ron Waksman, MD, Washington, USA Raimund Erbel, MD, Essen, Germany Steering Committee Ron Waksman, MD, Washington, USA Bernd Heublein † , MD, Hannover, Germany CEC & DSMB Jan Bart Hak, PhD, Groningen, NL Martial Hamon, MD, Caen, France Rafael Beyar, MD, Haifa, Israel IVUS Core laboratory Neil J. Weissman, MD, Washington, USA QCA Core laboratory Cardialysis, Rotterdam, The Netherlands Data Coordinating Ron Waksman, MD, Washington, USA Study Coordination Stefan Wagner, PhD, Erlangen, Germany Erbel et al., Lancet 2007;369(9576):

4 PROGRESS STUDY Australia M Horrigan, Melbourne, AUS
Belgium B de Bruyne & W Wijns, Aalst, BE Germany M Haude, S Sack, D Boese, R Erbel, DE Netherlands JJRM Bonnier & J Koolen, Eindhoven Switzerland F Eberli & T Lüscher, Zurich, CH P Erne, Luzern, CH UK C Di Mario & C Ilsley, London, UK USA R Waksman, Washington, USA Erbel et al., Lancet 2007;369(9576):

5 PROGRESS STUDY Procedure Details
2.5 mm x 15 mm pre PTCA < 16 atm AMS implantation 3.0 mm mm AMS size 10 mm mm < 16 atm post dilatation if necessary double marker balloon angiogram/IVUS before and after implantation Erbel et al., Lancet 2007;369(9576):

6 PROGRESS Study Protocol
Screening Treat- ment 1 d post 1 m post ± 7 d 4 m post ± 1 w 6 m post ± 2 w 12 m post ± 4 w Clinical follow-up X CK (CK-MB)/ Troponin I QCA IVUS MRI (subgr.)* *MRI for analysis of degradation kinetics Erbel et al., Lancet 2007;369(9576):

7 PROGRESS STUDY PCI Procedure Characteristics n - pre dilatation 100 %
63/63 pressure (8 atm, 20 sec) 9 ± 2.1 - AMS pressure, atm 16 ± 0.9 - post dilatation 67 % 42/63 - post dilatation pressure, atm 16 ± 3.9 2nd stent - average stent number 13 % 1.1 ± 0.3 8/63 Erbel et al., Lancet 2007;369(9576):

8 PROGRESS Study MLD: MLD stent: 1,05 ± 0,38 2,47 ± 0,37
100 80 60 Gain: 1,41 ± 0,46 Cumultive distribution % 40 20 0,0 0,5 1,0 1,5 2,0 2,5 3,0 3,5 4,0 MLD [mm] Erbel et al., Lancet 2007;369(9576):

9 Magnet Resonance Imaging of AMS The MRI compatible Stent
Magnetom, (Sonata, 1.5 T, Siemens) optimal vessel imaging no stent artefacts, AMS not visible Eggebrecht et al Circulation 112, 303 – 4, 2005

10 Computed Tomography 16 MSCT: AMS Stent
B C D bms- stent AMS Stent AMS Stent BMS Stent Lind et al Heart 91:1604, 2005

11 Micro CT of AMS and BMS Mg-Stent bare metal stent Malyar et al 2006
1 mm Malyar et al

12 GIRO 065-001 C-R Acute result After AMS Stent implantation
after 18 days Erbel et al JACC 2005

13 Absorbable Metal Stent (AMS)
MLDf/u: 1,34 ± 0,49 MLDstent: 2,47 ± 0,37 100 80 60 Loss: 1,08 ± 0,49 cumulative distribution % 40 20 0,0 0,5 1,0 1,5 2,0 2,5 3,0 3,5 4,0 MLD [mm] Erbel et al., Lancet 2007;369(9576):

14 QCA Analysis Parameter before after 4 - Months Ref MLD/mm 2.76 2.67
± ± 0.46 MLD/mm ± ± ± 0.49 Acute gain/mm ± 0.46 Late loss/mm ± 0.49 D % Stenosis ± ± ± 17 Erbel et al., Lancet 2007;369(9576):

15 PROGRESS STUDY 4 Months Angiography
Erbel et al., Lancet 2007;369(9576):

16 PROGRESS STUDY % N MACE Mortality Q- MI (Q- mit CK or CK-MB)
Hospital 30-days 4-Months % N MACE Mortality Q- MI (Q- mit CK or CK-MB) Non Q - MI (CK 2 x UNL) Ischemia driven TLR 23.8 15 Erbel et al., Lancet 2007;369(9576):

17 Vasomotion Testing after Magnesium Stent
Baseline values Stenosis diameteer (%) 39.1 Stenosis area ( %) 62.9 MLD (mm) 1.67 MLA (mm²) 2.19 Reference diameter (mm) 2.74 Reference area (mm²) 5.90 Vessel segment length(mm) 27.1 Stenosis length (mm) 13.5 Erbel et al., Lancet 2007;369(9576):

18 Vasomotion Testing after Magnesium Stent
Acetylcholine testing Stenosis diameter (%) 56.4 Stenosis area (%) 81.0 MLD (mm) 1.16 MLA (mm²) 1.06 Reference diameter (mm) 2.67 Reference area (mm²) 5.59 Vessel segment length (mm) 25.3 Stenosis length (mm) 17.9 Erbel et al., Lancet 2007;369(9576):

19 Magnesium Stent Conclusion AMS realized with low recoil
High technical sucess AMS permits MRT and CT based imaging No acute or subacute stent thrombosis iTLR rate comparable to BMS IVUS detected degradation within 4 M Vasomotion reactivation Drug elution and AMS delayed degradation – Dream concept* * Please listen to R Waksman in the next session, room 6