Colorimetry & Spectrophotometry
2
Useful Terminology Colorimetry
is the use of the human eye to determine the concentration of colored species. Spectrophotometry is the use of instruments to make the same measurements. It extends the range of possible measurements beyond those that can be determined by the eye alone.
3
Colorimetry Visual Observations – Because colorimetry is based on inspection of materials with the human eye, it is necessary to review aspects of visible light. Visible light is the narrow range of electromagnetic waves with the wavelength of nm. ROYGBIV = the mnemonic used to remember the colors of the visible spectrum.
4
COLOR Interaction between LIGHT and MATTER
5
COLOR RESULTS WHEN RADIATION IS ABSORBED BY AN ELEMENT OR
BY A COMPOUND FORMED THROUGH A REACTION RED YELLOW RED ABSORBED W H I T E YELLOW GREEN BLUE BLUE
6
Measurement of the amount of
COLORIMETRY Measurement of the amount of LIGHT ABSORBED by the COLOR DEVELOPED in a sample
7
Color Wheel (ROYGBIV) Complementary colors lie across the diameter on the color wheel and combine to form “white light”, so the color of a compound seen by the eye is the complement of the color of light absorbed by a colored compound; thus it completes the color.
8
Observed Color of Compound Color of Light Absorbed
Observed Color of Compound Color of Light Absorbed Approximate Wavelength of Light Absorbed Green 700 nm Blue-green 600 nm Violet 550 nm Red-violet 530 nm Red 500 nm Orange 450 nm Yellow 400 nm
9
Observed Color of Compound Color of Light Absorbed
Observed Color of Compound Color of Light Absorbed Approximate Wavelength of Light Absorbed Green Red 700 nm Blue-green Orange-red 600 nm Violet Yellow 550 nm Red-violet Yellow-green 530 nm 500 nm Orange Blue 450 nm 400 nm
10
Simple Spectrophotometer Schematic
The lamp emits all colors of light (i.e., white light). The monochromator selects one wavelength and that wavelength is sent through the sample. The detector detects the wavelength of light that has passed through the sample. The amplifier increases the signal so that it is easier to read against the background noise.
11
Spectrophotometer Sample Cell Detector Monochromator Light Source
12
RATIO TRANSMITTANCE (T) OF THE INTENSITY OF LIGHT LEAVING SOLUTION (I)
TO THE INTENSITY OF LIGHT ENTERING SOLUTION (IO)
13
TRANSMITTANCE IO I I T = IO %T = T x 100
14
LAMBERT’S LAW Relates the absorption of light to the depth or
thickness of the colored liquid Each layer of equal thickness will absorb the same fraction of light which passes through it An arithmetic increase in thickness gives a geometric decrease in light intensity transmitted
15
Transmittance Comparing Light Transmittance to Cell Length I0 I1 I2 I3
1.0 .9 .8 .7 .6 .5 .4 .3 .2 .1 Transmittance I1 I2 I3 I5 In I4 Units of Optical Path
16
BEER’S LAW Relates the absorption of light to the concentration
of the absorbing substance in the solution The fraction of light absorbed is directly proportional to the concentration of the absorbing substance An arithmetic increase in concentration gives a geometric decrease in light intensity transmitted
17
Transmittance Comparing Light Transmittance to Concentration I0 I1 I2
1.0 .9 .8 .7 .6 .5 .4 .3 .2 .1 Transmittance I1 I2 I3 I5 In I4 Units of Concentration
18
COLORIMETRY The Amount of Light Absorbed Is Related To:
The Chemistry Involved. 2. The Length of Light Travel. 3. The Amount (Concentration) of Absorbing Material.
19
THE COMBINED LAMBERT’S LAW
AND BEER’S LAW T = I I o T = abc Where: a = constant for particular solution b = length of absorbing layer (light path length) c = concentration of absorbing substance {- Sign Indicates an Inverse Relation}
20
Absorbance = A = - log T A =
TRANSMITTANCE T = I Io Absorbance = A = - log T T = 10 -abc log T = log (10 -abc) log T = -abc -log T = -(-abc) = abc -log T = abc A =
21
ABSORBANCE (A) A = abc A = - log T Where:
a = constant for particular solution b = length of absorbing layer (light path length) c = concentration of absorbing substance
22
ABSORBANCE (A) A = abc A = - log T If: Then:
a = held constant by carefully performing the analysis b = held constant by controlling the light path length Then: A is Directly Related to c (conc. of absorbing substance) If we can measure A, then we can determine c
23
CONCENTRATION CAN BE COLORIMETRICALLY
DETERMINED IF: 1. Able to chemically develop a color with that substance and only that substance 2. The developed color obeys (follows) Beer’s Law over a reasonable range of concentrations 3. The developed color must be stable for reasonable length of time, reproducible, and sensitive to small changes in concentration 4. All loss of transmitted light must be from absorbance by substance measured (developed color) 5. All of substance present in sample must be available for reaction with color developing agent 6. Able to measure amount of light absorbed
24
Sample Preparation Dilution Solids Removal --- Filter pH Adjustment
--- Coagulation --- Centrifuge --- Filter pH Adjustment Digestion
25
Release Combined Constituent Change Form of Constituent
DIGESTION Destroy Organics Release Combined Constituent Change Form of Constituent
26
Colorimetry Sensitive Color Development Stable Reproducible
Color Must Be: Stable Reproducible Sensitive
27
Color Development Must Control : pH Time Temperature Ionic Strength
28
Color Measurement Compare Sample Color to Known Standards
“Color Comparators” O.K. For Control – Not For Reporting
29
Compare Sample Color to Known Standards “Calibration Curve” (verified)
Color Measurement Compare Sample Color to Known Standards Spectrophotometer “Calibration Curve” (verified)
30
Colorimetric Instruments
31
Light Source Constant Controllable WHITE LIGHT Voltage Regulation
Diaphragm Fatigue Voltage Adjustment
32
Color (wavelength) Band
33
Monochromator Must be CAREFULLY Adjusted APERATURE OR SLIT PRISM OR
DIFFRACTION GRATING Must be CAREFULLY Adjusted
34
The Light Path is affected by the Cuvette
Sample Cell The Light Path is affected by the Cuvette Cuvette
35
Sample Cell Must be CAREFULLY Aligned Cuvette
36
PHOTOELECTRIC TUBE Differing Response for Various Wavelengths
“DETECTOR” Differing Response for Various Wavelengths Bausch & Lomb nm (w / filter) nm (w / filter) nm
37
of Filter and Phototube
PHOTOELECTRIC TUBE “DETECTOR” Differing Response for Various Wavelengths Must Use the Correct Combination of Filter and Phototube For Wavelength Of Analysis
38
Gives the Readout in Transmittance or Absorbance
INDICATING METER Gives the Readout in Transmittance or Absorbance
39
Some Meters Give Readout Directly in Concentration
INDICATING METER Some Meters Give Readout Directly in Concentration Use Only those Readings Between the Lowest and Highest Standard of Calibration
40
Some Meters Have “Built-in” Calibration
INDICATING METER Some Meters Have “Built-in” Calibration These Calibrations Should Be Verified Periodically Using a Series of Standards and Only those Readings Between the Lowest and Highest Standard of Calibration Should be Used
41
Optical System Lenses Mirrors Apertures Occluders
42
The Instrument Must be Carefully Handled,
Optical System The Instrument Must be Carefully Handled, Protected From Dust and Vapors, and Serviced Only By Qualified Technicians
43
Spectrophotometer Sample Cell Monochromator Detector Light Source
44
Instrument Operation: Set Zero Absorbance w/Blank
COLORIMETRY Instrument Operation: Warm-up Set Monochromator Set ∞ Absorbance Set Zero Absorbance w/Blank Re-adjust as Needed
45
Instrument Operation: Watch for Irregularities
COLORIMETRY Instrument Operation: General Rule – Absorbance Between and 0.700 Some Analyses More Restrictive Best Readings – Between Lowest and Highest Standards Used In Calibration Watch for Irregularities
46
COLORIMETER CALIBRATION
Calibration or Standardized By Measuring Absorbance Readings of a Series of Known Standards Comparison of These Readings to the Reading for a Sample 1. Computer Spreadsheet 2. Instrument with Internal Microprocessor 3. “Plotting” a Graph
47
COLORIMETER CALIBRATION
Calibration or Standardized By Measuring Absorbance Readings of a Series of Known Standards Comparison of These Readings to the Reading for a Sample Verified Frequently At Least One Standard In Acceptable Range Each Time Samples Are Analyzed
48
COLORIMETER CALIBRATION Repeat Calibration:
1. Significant Change In Procedure, Equipment, or Reagents Determined Length of Time (Max. Six Months) 3. Verification Standard Not In Acceptable Range
49
Calibration Steps: 1. Prepare Stock Solution
2. Prepare a Series of Dilutions 3. Same Preparation Steps as Sample 4. Develop Color 5. Measure Absorbance of Each 6. Prepare Calibration “Curve”
50
(Using Phosphorus Analysis Example)
Calibration Curve (Using Phosphorus Analysis Example)
51
How Do We Use This Principle?
COLORIMETRY How Do We Use This Principle? Perform a Chemical Reaction with the Element to be Analyzed that Results in a Compound of that Element that Absorbs Light. Measure the Amount of Light Absorbed.