Physiologic Control Algorithms for Rotary Blood Pumps using Pressure Sensor Input Edward Bullister, Ph.D. Sanford Reich, Ph.D. APEX Medical, Inc. ISRP 2001 18 August 2001
2
Why Use Pressure Inputs? Provides physiologic feedback for pump control. Provides added-value pump diagnostic and monitoring functions. Increases capability for patient monitoring. Potentially increases patient quality of life.
3
How to Implement? Control Algorithm Development Design Strategy to Mimic Patients’ Physiologic Control Control Algorithm Schematic Control Algorithm Detail Control Algorithm Results Added-Value Diagnostic and Monitoring Functions Patient Monitoring Hardware Considerations Summary
4
Control Algorithm
5
Level 1: Basic Control Algorithm Level 1 Control Input: LVDFP - Left Ventricular Diastolic “Filling Pressure” Level 1 Control Output: Pump Flow Rate Proportional Integral Control Algorithm d/dt(Flow) = K * (LVDFP - P desired ) K = 0.1 L/min/mmHg Flow Pressure Simple Robust
6
Level 1 Results
7
Level 2: Exercise Control Algorithm Level 2 Control Inputs: Arterial Pressure Pulse Rate Increase (e.g., during exercise) Level 2 Control Output: Desired LVDFP Level 2 Limits: Max/min LVDFP Max/min Arterial Pressure
8
Level 2 Results
9
Flow Rate Monitor using Pressure Pressure Calculated from Pump Speed and Pressure Difference Independent of Motor Current Includes High Frequency Content Flow (L/min) Calculated From Pressure Flow meter Measurement Time (sec)
10
Hydraulic Power Monitor Hydraulic Power (HP) into Blood Pump: HP pump = P pump * PumpFlow (continuous) Heart: HP heart ~ P heart * PumpFlow (measured during systole)
11
Hardware Considerations Pressure Sensor Technology Thin-Film Based MEMS Based Any Rotary VAD Pressure Sensor Placement
12
Component Analysis Computational Fluid Dynamics (CFD) Example - Inlet Cannula Establish optimal location for pressure sensor Calculate pressure coefficient K for nonlinear relationship: P = K*V 2
13
Summary An initial control algorithm has been implemented to auto- regulate rotary blood pumps using physiological pressure inputs. Two levels of control for a rotary pump have been tested in a mock loop setup. The pressure signals produce added-value information. Additional monitoring and control levels have been conceived. Goal is to contribute to patient quality of life.