1
Breathing systems and the anaesthetic machine Zsolt Molnár SZTE, AITI
2
Introduction „Auld” definitions „New”: Mapleson systems 1954
Open-, semi open- Semi-closed, closed-systems „New”: Mapleson systems 1954 Rebreathing Non-rebreathing 10 10
3
Physiolgy refreshment course
VT 4-7 ml/kg VD 2-3 ml/kg Peak inspiratory flow (PIF) At rest: l/min Breathing cycle Inspiration – expiration – expiratory pause 10 10
4
Mapleson systems Parts Efficacy FG inlet Rezervoire balloon
Pop-off valve Efficacy FGF – to prevent CO2 rebreathing 10 10
5
Pressure relieving valve
Parts Disc Sharp edge Spring Safety opening P 60-70 cmH2O 10 10
6
Mapleson A – spont. breathing
„Magill attachment” >110 cm FGF = 1 x VA (70ml/kg/m) VA = ? PIF = ? Monitoring: capnograph 10 10
7
Capnography A: normal B: rebreathing C: diaphragm movements
pCO2 A: normal B: rebreathing C: diaphragm movements D: cardiac oscillation E: sampling error F: bronchospasm A B C D Sebészeink…, anamnesis... Általános szabály, hogy a viszgán legyen egy algoritmususnk, ami alapján válaszolunk a kérdésekre. Tehát egy ilyen beteg ellátása preop…..problémát egyaránt jelent. Mielőttt azonban ennek nekiállnánk, frissítsünk fel néhány elengedhetetenül fontos élettani fogalmat. Molnár ‘99 E F
8
Mapleson A – IPPV „Close” pop-off valve Rebreathing may occur
FGF = 2.5 x VA 10 10
9
Coaxial systems Mapleson D or Bain Mapleson B or Lack
IPPV: FGF = 1 x VA Spont: FGF = 2.5 x VA Mapleson B or Lack „A” better 10 10
10
Mapleson C Water’s circiut „To-and-fro”: IPPV: FGF >3 x VA
Spont: FGF > 2 x VA „To-and-fro”: 10 10
11
Water’s circuit O2 Rezervoire balloon Mask, ET tube Valve Molnár ‘99
A hipovolémia során a keringő vérmennyiség csökkenése a … Valve Molnár ‘99
12
Mapleson E and F E: Ayre’s T-piece F: Jackson-Rees
Spont: FGF = 1.5 x VA IPPV: FGF = 3 x VA F: Jackson-Rees Monitoring PEEP 10 10
13
T-piece T-piece Outlet Inlet Molnár ‘99
14
Bernoulli’s law Molnár ‘99 8 LPM 50% O2 100*8 + X*21 = (8+X)*50
FGF = = 21 L/p Molnár ‘99
15
T-piece parts: Venturi injector + humidification
Physiology PIF~20-30 l/min Bernoulli’s effect FiO2 60-30 L/min Active humidification Hot water bath Heating coil Levegő Molnár ‘99
16
A happy chapy on T-piece
Molnár ‘99
17
The anaesthetic machine
Cylinders (O2, N2O, Air) Rotameters Vaporizers Breathing circuit CO2 absorber Safety: FiO2 + alarm O2 bypass Pressure relieving valve 10 10
18
Rebreathing system (circuit)
10 10
19
CO2 absorber CaOH2, NaOH, sylicate CO2 + 2NaOH Na2CO3 + H2O + heat
Na2CO3 + CaOH2 2NaOH + CaCO3 10 10
20
Rotameters 10 10
21
Rotameters Gas specific Safety features: Min FiO2 = 0.3 FiO2 alarm 10
22
Vaporisers ages ago… Inside the circuit Small vaporizing surface
Quick cooling 10 10
23
…and today Outside the circiut Well insulated containers 10 10
24
…and today Outside the circiut Well insulated containers Agent safe 10
25
…and today Outside the circiut Well insulated containers Agent safe
Heat compensation 10 10
26
Summary Mapleson systems: advantages Mapleson systems: disadvantages
Simple, cheap, safe Low resistance Good for practice Mapleson systems: disadvantages Non-economic Not environment friendly High heat and humid loss 10 10