Download presentation
Presentation is loading. Please wait.
Published byOctavia Copeland Modified over 9 years ago
2
Noise, Reset, EGM, Magnet, Advanced Hysteresis
3
Noise Göran Mathson, 2005
4
for internal use only Noise Electromagnetic Interference (EMI) or Noise At least 400 cycles per minute Is very unlikely to be caused by myopotentials
5
for internal use only Noise EMI in Relative Refractory Period of (RRP) initiates a Noise Sampling Window (NSW) NSW is extension of the RRP Noise detected in the Ventricular Channel during DDD / DDI mode operation extends A and V refractory period
6
for internal use only Noise If the Noise persists there will be asynchroneous pacing at the Base Rate (or Sensor Indicated Rate if the Sensor is On) Pacemaker resumes normal operation at the programmed parameters as soon as the Noise stops
7
for internal use only Noise Mode, Atrial Channel PVARP Programmable PVARP Absolute Relative A-A Interval
8
for internal use only 100 ms175 ms Relative Absolute Sensed Event NSW 150 ms Sensed Event NSW 150 ms Noise sampling window (NSW) Sensed Event NSW 150 ms Sensed Event
9
for internal use only Noise Mode, Atrial Channel A-A Interval
10
Reset Göran Mathson, 2005
11
for internal use only Reset When the device has reverted to backup VVI operation, the programmer will display a pop- up message indicating that the device is in back up VVI values. Under most conditions the previously programmed settings can be restored, by following the on-screen instructions.
12
for internal use only Reset parameters
13
Stored Electrograms Göran Mathson, 2005
14
for internal use only Stored Electrograms Programmability Triggers
15
for internal use only Stored Electrograms Stored electrograms are available in the following devices*: Zephyr ® Victory ® Identity ® Integrity ®
16
Stored Electrograms Programmability
17
for internal use only Programmability The device can store a variety of intracardiac configurations for various time frames The following screen shot demonstrates this programmability
18
for internal use only Programmability
19
for internal use only Programmability Sampling Option This parameter determines how the device treats additional Stored EGMs after the allocated memory has been filled Freeze (nominal) Once the device has stored the programmed number of EGMs no more EGMs will be stored Continuous Allows a new Stored EGM to overwrite the oldest Stored EGM First in, First out
20
for internal use only Programmability Number of Stored EGMs Determines how many EGMs the device can store 1, 2, 4, 8, 12 Maximum Duration Calculated by the device based on programmed values
21
for internal use only Programmability Channel Determines what IEGMs will be recorded by the device Programmed Options Atrial Ventricular Dual Cross-Channel (nominal)
22
for internal use only Programmability Channel Atrial When selected, this value will cause an Atrial IEGM to be stored Ventricular When selected, this value will cause a Ventricular IEGM to be stored
23
for internal use only Programmability Channel Dual Both an Atrial and Ventricular IEGM will be stored Cross-Channel Both the Atrial and Ventricular IEGM will be stored on a single electrogram In some patients, this may resemble a surface ECG
24
for internal use only Programmability
25
for internal use only Programmability
26
for internal use only Programmability Atrial EGM Configuration Determines the IEGM to be recorded for the Atrial channel on the Stored EGMs The various configurations can be viewed on the real-time EGM
27
for internal use only Programmability Atrial Dynamic Range Sets the Stored EGM gain for the Atrial channel Programmable Options 15.0, 7.5, 3.0, 1.5 mV (nominal 3.0 mV) These parameters can be previewed on the real-time electrogram
28
for internal use only Programmability Ventricular EGM Configuration Determines the IEGM to be recorded for the Ventricular channel on the Stored EGMs The various configurations can be viewed on the real-time EGM
29
for internal use only Programmability Ventricular Dynamic Range Programmable Options 15.0, 7.5, 3.0, 1.5 mV (nominal 15.0 mV) These parameters can be previewed on the real-time electrogram
30
for internal use only Programmability If AutoCapture™ Pacing Systems is enabled then the Ventricular Configuration is fixed to Vtip-Vring
31
Stored EGMs Stored Electrogram Triggers
32
for internal use only Stored Electrogram Triggers The device can be programmed to automatically store an EGM when triggered by*: Advanced Hysteresis AT/AF Detection High Atrial Rate High Ventricular Rate AMS Entry AMS Exit PVCs PMT Termination Patient Activated / Magnet application * Available triggers depend on the pacemaker model. See device manual.
33
for internal use only Programmability
34
for internal use only Stored Electrogram Triggers Trigger Options Advanced Hysteresis Triggered when Advanced Hysteresis is activated AT/AF Detection Triggered when AT/AF Detection meets programmed criteria High Atrial Rate Off, 175, 200, 225, 250, 275, 300 › No, of Consecutive Cycles -2, 3, 4, 5, 10, 15, 20
35
for internal use only Stored Electrogram Triggers Trigger Options High Ventricular Rate Off, 175, 200, 225, 250, 275, 300 › No, of Consecutive Cycles -2, 3, 4, 5, 10, 15, 20 AMS Entry Triggered when the device enters mode switches AMS Exit Triggered when the device exits mode switches
36
for internal use only Stored Electrogram Triggers Trigger Options PVCs No. of Consecutive PVCs › 2, 3, 4, 5 PMT Termination Patient Activated Magnet Placement (magnet applied for at least 1 sec)
37
for internal use only Stored Electrogram Triggers Additional Considerations Pre-trigger data is collected, approximately 50% of the EGM Some triggers require an associated feature to be active Advanced Hysteresis, AMS Entry, AMS Exit and PMT Termination EGM Storage for High Atrial Rate and AMS Entry and Exit are mutually exclusive Can’t be programmed on together
38
for internal use only Stored Electrogram Triggers Stored EGMs are disabled: When Telemetry is active When Elective Replacement Indicator (ERI), has been achieved
39
for internal use only Impact of SEGM on Victory VictoryDR - 5816 frozen rolling No egm25% egm50% egm100% egm 2.5 V atrium and ventricle, 100% inhibited 17.814.111.68.6 2.5 V atrium, 1.0 V ventricle, 100 % pacing 1310.99.37.3 2.5 V atrium and ventricle, 100% pacing 11.7108.66.8 SR - 5610 frozen rolling No egm25% egm50% egm100% egm 11.48.87.25.2 10.186.74.9 8.87.26.14.6
40
for internal use only Conclusions & Recommendation Field -> Customer Education SEGM in latest SJM pacemakers always records pre-onset the trigger.This may have a significant effect on PG longevity Use SEGM selectively (for example, if episodes are in the log, patient is symptomatic). Consider using Frozen, not continuous, since it is likely to collect SEGM, then turn off, minimizing current drain.
41
Magnet Response Göran Mathson, 2005
42
for internal use only Settings: Off; Battery Test (Nominal: Battery Test) How will the PM respond to the magnet placement Telemetry is disabled when magnet is held over device. Temporarily suspended functions during magnet is held over or Cancelled The Magnet Response
43
for internal use only Terminologie: BOL= Beginning of life EOL = End of life ERT = Elective replacement time ERI = Elective replacement indicator: indication ERT is reached TMT = Threshold margin test: automatic action PM Response: VOO DOO AOO Modus Freq. = indication of batt. Condition Use: Freq. = indication of batt. Condition Troubleshooting – deactivation of algorithms Inhibition of sensing The Magnet Response
44
for internal use only Pacemaker reference guide
45
for internal use only Pacemaker reference guide WWW.SJM.COM Devices For Physicians & Health Care Professionals Implantable Device Reference Guide
46
Neurocardiogenic Syncope Göran Mathson, 2005
47
for internal use only Neurocardiogenic Syncope Common faint also called “Neurally mediated syncope” or “Vasovagal syncope” Young adult, structurally normal heart Malignant vasovagal syncope pacing First line therapy: medication Ineffective and bradycardic component 3 types: Cardioinhibitory, vasodepression and mixed
48
for internal use only Neurocardiogenic Syncope 3 Major Objectives Provide high rate pacing only during the spells Otherwise inhibit the pacemaker Maintain AV Synchrony
49
for internal use only Auto Rate Hysteresis With Search Most patients with Neurocardiogenic Syncope are in sinus rhythm with intact conduction the majority of the time. However, if they have an abrupt drop in rate, they may benefit from an elevated rate of pacing.
50
for internal use only Programmable parameters Hysteresis Rate Programmable Search Interval Programmable Cycle Count Intervention Rate Intervention Duration Recovery Time Advanced Hysteresis Response
51
for internal use only Advanced Hysteresis Response Hysteresis Rate Allows intrinsic activity to be tracked below the base rate Programmable Values: OFF, 30 - 130 in 5 min -1, 140, 150
52
for internal use only Advanced Hysteresis Response Search Interval Allows the device to periodically extend the pacing interval to search for intrinsic activity Programmable Values: OFF, 5, 10, 15, or 30 min.
53
for internal use only Cycle Count The number of intervals the device will extend the pacing interval to search for intrinsic activity. The Cycle Count also serves as the number of cycles the patient’s intrinsic rate will be allowed to drop below the hysteresis rate before pacing at the Intervention Rate begins. Programmable Values: 1-16 cycles Advanced Hysteresis Response
54
for internal use only Advanced Hysteresis Response The rate the pacemaker will operate at if the patients intrinsic rate falls below the Hysteresis Rate for the programmed Cycle Count. It is programmable as a fixed value or as a function of the Intrinsic rate.** Programmable Values: OFF, Base Rate, 80 - 120 (steps of 10 min -1 ), Intrinsic + 0, + 10, +20, +30 min -1 ** The Intrinsic rate is calculated as the average Ventricular Rate over the previous 64 cycles Intervention Rate
55
for internal use only Advanced Hysteresis Response The time in minutes that the pacemaker will operate at the Intervention Rate. Following the Intervention Duration the rate will be decreased, according to the programmed Recovery Time, until either the Base Rate or Sensor Indicated Rate is reached. Programmable Parameters: 1-10 min (steps of 1 min) Intervention Duration
56
for internal use only Advanced Hysteresis Response Programmable Values Fast Medium Slow Very Slow The length of time (in min) that it will take for the pacemaker to reduce the pacing rate from the Intervention Rate to the Base Rate NOTE: This parameter also serves as the setting for the sensor Recovery Time Recovery Time
57
for internal use only Parameter settings for Neurocardiogenic syncope ParameterValueRationale Pacing modeDDD DDI Excer. AV block Rate modulationDISABLEEven no mild levels Base Rate60 -100 bpmSympt. dependent Intervention rate 10 – 20 bpm Above av. rate Av.= intrins.rate before drop Hysteresis rate≤ 45 bpmAsist. >3 sec = 20bpm
58
for internal use only Parameter settings for Neurocardiogenic syncope ParametersValueRationale Rest Rate45 – 50 bpmLower rate at rest Escape Cycles2Persistent slow Cycle Number3Sinus warm up Intervention Duration 2 to 5 minutesThe spells are very brief Recovery TimeMediumSame as sensor AV delaySufficient / VIPVP not needed AutocaptureEnableThreshold rise
59
for internal use only Parameter settings for Neurocardiogenic syncope OFF
60
for internal use only Parameter settings for Neurocardiogenic syncope
61
for internal use only Parameter settings for Neurocardiogenic syncope
Similar presentations
© 2025 SlidePlayer.com. Inc.
All rights reserved.