IPC Friedrich-Schiller-Universität Jena 1 7. Fluorescence microscopy 7.3 FRET microscopy - The different Dipoles 1. static electric Dipole: Far away: E ~ r -3 along all directions 2. radiating (Hertz's) Dipole: (VERY DIFFERENT!) far away: in line: zero field (n x p = 0) orthogonal: E ~ r -1 r = = distance from middle to position n = direction unit vector p = dipole vector
2
IPC Friedrich-Schiller-Universität Jena 2 7. Fluorescence microscopy 7.3 FRET microscopy - The different Dipoles 3. transition Dipole moment – an analogy : - + "molecular" conductor E = electric Field Induced Transition Dipole Transition Dipole Moment : a vector along who's direction the dipole will be induced
3
IPC Friedrich-Schiller-Universität Jena 3 Resolution of a light microscope is limited to several hundred nanometers (< organelles) FRET allows detection of molecule-molecule interactions on a nanometer scale by means of a light microscope Decrease of donor- emission Increase of acceptor emission Reduction of donor fluorescence life-time Energy transfer (FRET-efficiency) depends strongly on donor-acceptor distance R 0 = Förster-radius (distance for which energy transfer is half maximal) 7. Fluorescence microscopy 7.3 FRET microscopy sensitized emission
4
IPC Friedrich-Schiller-Universität Jena 4 FRET ratio imaging = acceptor emission at donor excitation (sensitized emission S Akzeptor ) divided by donor emission at donor excitation (S Donor ) Advantages: Since both donor decrease as well as acceptor increase contribute to the signal the signal-to-noise ratio is better than for solely recording the acceptor fluorescence S Akzeptor S Donor 7. Fluorescence microscopy 7.3 FRET microscopy
5
IPC Friedrich-Schiller-Universität Jena 5 FRET ratio imaging – problems: Correction for direct excitation of the acceptor when exciting donor (control measurement with YFP only) = correction factor r DE Excitation wavelength Correction for bleedthrough : Portion of CFP in yellow channel for blue excitation in absence of FRET (acceptor) = bleedthrough of CFP in YFP-channel (r BT,CY ) or bleedthrough of YFP in CFP-channel (r BT,YC ) FRET-detection channel 7. Fluorescence microscopy 7.3 FRET microscopy
6
IPC Friedrich-Schiller-Universität Jena 6 FRET ratio imaging – 3-filter-set: 1.Donor excitation and emission (I CFP,430 ) 2.Acceptor excitation and emission (I YFP,514 ) 3.Donor excitation and acceptor emission (I YFP,430 ) 7. Fluorescence microscopy 7.3 FRET microscopy
7
IPC Friedrich-Schiller-Universität Jena 7 FRET ratio imaging – 3-filter-set: Model of FRET-detection of Src- Csk protein interaction (Src = protein tyrosine kinase Csk = C-terminal Src kinase) Important signal transduction step during blood coagulation 7. Fluorescence microscopy 7.3 FRET microscopy No FRET
8
IPC Friedrich-Schiller-Universität Jena 8 FRET ratio imaging – 3-filter-set: Visualization of Src-Csk-interaction during a IIb ß 3 -induced fibrinogen adhesion in a thrombocyte model cell line (A5-CHO) by means of FRET 7. Fluorescence microscopy 7.3 FRET microscopy superposition
9
IPC Friedrich-Schiller-Universität Jena 9 FRET ratio imaging – 3-filter-set: FRET for displaying Ca 2+ in living cells via Yellow-Cameleon-2 (YC2) sensor FRET-ratio image of HeLa-cells, expressing the YC2-sensor before and after adding ionomycin FRET response of HEK/293 cells expressing YC2-seonsor after adding 1nM ionomycin and additional extracellular Ca21 (30 mM) 7. Fluorescence microscopy 7.3 FRET microscopy
10
Acceptor Bleaching Free Donor + 1 Donor (Pairs) Acceptor Removed Free Donor + Donor (Pairs) Donor excitation, Donor emission detection 1 can be calibrated De-Quenching
11
Principle of Acceptor-bleaching- FRET microscopy Donor fluorescence should increase after acceptor bleaching Principle Donor fluorescence should increase (dequenching) after “removal” of the acceptor Acceptor depletion FRET
12
CFP Exc 457 Em 470-500 ‘Sens-YFP’ Exc 457 Em 535-570 YFP Exc 514 Em 535-570 Combined 100 µm Data of Dorus Gadella Donor Emission channel Acceptor Emission channel
13
CFP Exc 457 Em 470-500 ‘Sens-YFP’ Exc 457 Em 535-570 YFP Exc 514 Em 535-570 Combined 100 µm Data of Dorus Gadella BUT is it: High concentration, low efficiency? Low concentration, high efficiency? Donor Emission channel Acceptor Emission channel
14
IPC Friedrich-Schiller-Universität Jena 14 FRET fluorescence life-time microscopy: In case of FRET the donor fluorescence life-time is reduced. Determination of this donor life time reduction yields a quantitative FRET measurement which is independent of dye concentration or spectral contamination (crosstalk, bleedthrough). Dimerization of C/EBP® – proteins in GHFT1-5 cell nuclei (donor/acceptor CFP/YFP-C/EBP®) 7. Fluorescence microscopy 7.3 FRET microscopy
15
IPC Friedrich-Schiller-Universität Jena 15 7.4 Fluorescence Life-Time Imaging Microscopy (FLIM) 7. Fluorescence microscopy Laser pulse Longer fluorescence life-time Shorter fluorescence life-time Methods of time-resolved fluorescence diagnostics Time-resolved measurements Sample is excited by a short laser pulse Sample molecules relax individually according to the transition probability of the different relaxation pathways to the ground state Fluorescence intensity exhibits mono-, multi or non-exponential decay depending on nature and number of fluorescence contributions.
16
IPC Friedrich-Schiller-Universität Jena 16 Time-resolved measurements Intensity integrating measurements The determination of the fluorescence decay time ¿ or times ¿ i and relative amplitudes ® i in case of multiple contributions is possible by recording the fluorescence signal for several measurement points after the excitation pulse. For a mono-exponential decay behavior or to determine the average decay time ¿ two sampling points are sufficient For two times t 1 and t 2 after the excitation pulse the detector signal is integrated for a sampling window ¢ T. The ratio of the measurement signals D 1 und D 2 can be used to calculate the decay-time ¿ or the average decay-time ¿ : Methods of time-resolved fluorescence diagnostics 7. Fluorescence microscopy 7.4 Fluorescence Life-Time Imaging Microscopy (FLIM)
17
IPC Friedrich-Schiller-Universität Jena 17 Methods of time-resolved fluorescence diagnostics Time-resolved measurements Gated fluorescence detection Gated optical image intensifiers (GOI) are capable of taking pictures with high (sub- nanosecond) time resolution i.e. camera with ultrafast shutter (gate < 100 ps) which can be opened and closed for different delay-times after the sample has been excited with an ultrashort laser-pulse. By collecting a series of time-scanned fluorescence intensity images for different delay-times after excitation the fluorescence decay profile for every pixel in the field of view can be accessed and displayed as false color plot = fluorescence life-time image 7. Fluorescence microscopy 7.4 Fluorescence Life-Time Imaging Microscopy (FLIM)
18
IPC Friedrich-Schiller-Universität Jena 18 Methods of time-resolved fluorescence diagnostics Time-resolved measurements Gated fluorescence detection Tissue section of a rat ear: (a) Brightfield microscopy image stained with orcein (b) Fluorescence intensity- and (c) FLIM images of an unstained parallel sample (tissue autofluorescence) (excitation 410 nm; FLIM false color plot from 200 ps (blue) to 1800 ps (red) (a) (b) (c) (Top) FLIM image of an unstained human pancreas section (tissue autofluorescence) with an endocrine tumor (below) Brightfield image of the same section after conventional histopathological staining 7.4 Fluorescence Life-Time Imaging Microscopy (FLIM) 7. Fluorescence microscopy
19
IPC Friedrich-Schiller-Universität Jena 19 Methods of time-resolved fluorescence diagnostics Time-resolved measurements TCSPC = time-correlated single photon counting In case of intensive excitation light many electrons of the dye are getting excited for every laser pulse i.e. the average life-time can be deduced from the fluorescence decay-time after every pulse (multiple photon emission). A common FLIM method is the measurement of the life-time for single fluorescence photons. In doing so the dye is excited by light pulses of extremely low intensity in a way that at most one electron per pulse gets excited. The individual life- time of every photon is measured and the average life-time is determined staistically. 7.4 Fluorescence Life-Time Imaging Microscopy (FLIM) 7. Fluorescence microscopy
20
IPC Friedrich-Schiller-Universität Jena 20 Methods of time-resolved fluorescence diagnostics Time-resolved measurements TCSPC = time-correlated single photon counting Detection of single photons of a periodic light signal Light intensity is so weak, that the probability to detect a photon within one period is very small. Periods with more than one photon are extremely rare For every detected photon its delay time with respect to the excitation pulse is determined A delay distribution builds up over many pulse Time resolution up to 25 ps 7.4 Fluorescence Life-Time Imaging Microscopy (FLIM) 7. Fluorescence microscopy
21
IPC Friedrich-Schiller-Universität Jena 21 Probe Stop watch = TAC: Time-to-Amplitude Converter converts time between a start and a stop pulse by charging a capacitor with constant current Start can be reference (from laser) and photon is stop -> Problem is loss of much time (due to reset time) -> Reverse counting (start = photon, stop=next laser pulse) Histogram of arrival times after excitation -> fluorescence life-time. 7.4 Fluorescence Life-Time Imaging Microscopy (FLIM) Methods of time-resolved fluorescence diagnostics Time-resolved measurements TCSPC = time-correlated single photon counting 7. Fluorescence microscopy
22
IPC Friedrich-Schiller-Universität Jena 22 Fluorescence intensity image of a vacuole which is labeled by fluorescent phospholipids FLIM image and corresponding distribution of life-times. Long life-times (red) are found in the cell membrane while the cytoplasma exhibits shorter life-times pointing towards a less ordered environment. 7.4 Fluorescence Life-Time Imaging Microscopy (FLIM) Methods of time-resolved fluorescence diagnostics Time-resolved measurements TCSPC = time-correlated single photon counting 7. Fluorescence microscopy
23
IPC Friedrich-Schiller-Universität Jena 23 Methods of time-resolved fluorescence diagnostics Steady state measurements Phase modulation Intensity of a continuous wave (CW) source is modulated at high frequency by a standing wave acousto-optic modulator ( 50 MHz) which will modulate the excitation intensity at double frequency. Detected fluorescence is modulated at the same frequency. The observed phase shift with respect to the excitation and the modulation depth M (ratio of Ac signal to DC signal) depends on the fluorescence life-time of the excited fluorophores. Fluorescence lifetimes phase and phase can be calculated and should be identical for single exponential decays. 7.4 Fluorescence Life-Time Imaging Microscopy (FLIM) 7. Fluorescence microscopy
24
IPC Friedrich-Schiller-Universität Jena 24 Methods of time-resolved fluorescence diagnostics Steady state measurements Phase modulation Measurement values: Demodulation (modulation depth) M Phase shift Modulation of excitation light with, which is characterized by modulation depth M E = a/d and E : leads to an accordingly modulated fluorescence signal F(t) with demodulation M F = A/D und phase F 7.4 Fluorescence Life-Time Imaging Microscopy (FLIM) Time Intensity Excitation light Fluorescence 7. Fluorescence microscopy
25
IPC Friedrich-Schiller-Universität Jena 25 Methods of time-resolved fluorescence diagnostics Steady state measurements Phase modulation Rate equation of change of number of excited molecules F(t) ~ N(t) Relationship between fluorescence life-time and fluorescence emission behavior upon intensity modulated excitation light Relationship between measurement parameters: M = M F / M E as well as = E - E and life-time : Absorption rate: 7.4 Fluorescence Life-Time Imaging Microscopy (FLIM) 7. Fluorescence microscopy
26
IPC Friedrich-Schiller-Universität Jena 26 Methods of time-resolved fluorescence diagnostics Steady state measurements Phase modulation Continuous intensity modulated excitation of fluorescence transforms the determination of fluorescence decay-times to measurements of phase shifts and demodulation of the fluorescence signal Demodulation M and phase shift of the fluorescence depend on the fluorescence life- time as well as on the modulation frequency = 2 of the excitation light. The simulation shows the dependency of M and for a decay time of = 4 ns and a frequency of = 40 MHz Choosing frequency at 1 7.4 Fluorescence Life-Time Imaging Microscopy (FLIM) 7. Fluorescence microscopy
27
IPC Friedrich-Schiller-Universität Jena 27 Methods of time-resolved fluorescence diagnostics Steady state measurements Phase modulation Frozen section of portio biopsies in the spectral region of Em >500 nm ( exc = 457 nm; = 40 MHz) Top right: Fluorescence intensity Bottom right: corresponding HE stain image. 7.4 Fluorescence Life-Time Imaging Microscopy (FLIM) 7. Fluorescence microscopy