Defibrillator Implants: Bruce Murdoch PhD CHP CLSO Argonne National Laboratory Argonne, IL USA btmurdoch@anl.gov AIHce 2011, Portland OR Roundtable Session RT 201 Occupational Exposure to Electromagnetic Radiation
2
Typical ICD Device Dimensions: 2” X 2.5” X 0.2” AIHce 2011, Portland OR 2
3
Implant Placement EM Vulnerabilities Leads Control Unit Failure Modes False Trigger Stopping AIHce 2011, Portland OR 3 Source: www.defibrillator-help.comwww.defibrillator-help.com
4
AIHce 2011, Portland OR 4 CASE STUDY Where? APS (Advanced Photon Source) Who? Accelerator Operator w/ICD Why? Work Clearance
5
AIHce 2011, Portland OR 5
6
6
7
7 APS Control Room
8
Manufacturer Test Standard ANSI/AAMA PC69 - 2007 AIHce 2011, Portland OR 8
9
Typical Manufacturer (Medtronic) Exposure Limit Recommendations Static B-FieldB ≤ 0.5 mT (5G) 50/60 HzB ≤ 0.1 mT (1G) E ≤ 6000 V/m ≥ 10 kHzB ≤ 1.25 mT (12.5 G) R-F E-Field (≥ 150 kHz)E ≤ 100 V/m AIHce 2011, Portland OR 9
10
ACGIH 2010 TLV Book Exposure Limit Recommendations Static B-Field B ≤ 0.5 mT (5G) ELF (5 Hz–2kHz)B ≤ 0.1 mT (1G) VLF (2kHz–400 kHz)No recommendation R-F (100 kHz–3GHZ) E ≤ 3-10 V/m AIHce 2011, Portland OR 10
11
Static B-Fields MetroLab Model ETM-1 3-Axis Magnetometer AIHce 2011, Portland OR 11
12
ELF Holaday HI-3627 ELF Magnetic Field Meter w/ HI-3602 Probe 30 Hz – 2 KHz AIHce 2011, Portland OR 12
13
VLF Holaday HI-3637 VLF Magnetic Field Meter w/ HI-3091 Probe 2kHz – 400 KHz AIHce 2011, Portland OR 13
14
R-F Narda NBM-520 RF Survey Meter w/ EF3091 E-Field Probe 100 kHz – 3 GHz AIHce 2011, Portland OR 14
15
Spectrum Coverage AIHce 2011, Portland OR 15 RF
16
AIHce 2011, Portland OR 16 The Instrument Ensemble
17
Survey Plan Daily Control Room work area Weekly maintenance walkthroughs ESH Coordinator guidance 84 specific survey locations plus area sweeps Full spectrum survey at each location AIHce 2011, Portland OR 17
18
Control Room Displays and Instruments AIHce 2011, Portland OR 18
19
Control Room Panels AIHce 2011, Portland OR 19
20
RF Distribution System AIHce 2011, Portland OR 20
21
Electrical Panels ELF, VLF: B ≤ 0.1 G RF: E ≤ 4 V/m AIHce 2011, Portland OR 21
22
Electrical Distribution Room ELF: B ≤ 0.4 G @ 6 inches from transformer RF: E ≤ 4 V/m AIHce 2011, Portland OR 22
23
Mezzanine RF System Freq = 325 MHz E(RF) up to 3.5 V/m @ 6 in. AIHce 2011, Portland OR 23
24
Power Control Systems AIHce 2011, Portland OR 24
25
Pulser Power Transformers AIHce 2011, Portland OR 25
26
Pulser Power Transformers AIHce 2011, Portland OR 26
27
RF Distribution Waveguides Freq = 325 mHz AIHce 2011, Portland OR 27
28
Klystron Gallery Freq = 2.865 Mhz AIHce 2011, Portland OR 28
29
Linac Klystrons Freq = 2.8565 GHz E(RF) = 2 V/m @ 1 ft AIHce 2011, Portland OR 29
30
Case Study Results Employee barred from –Fundamental RF Room: 7 V/m RF –Electrical Distribution Room: 4 V/m RF Cautions –Some instrument racks: 4 V/m RF @ 12” –Some waveguides: 3.5 V/m RF @ 12” AIHce 2011, Portland OR 30
31
Credits Technical Assistance –Mary Malik –Adam Sauter Photography –George Joch –Edmund Chang AIHce 2011, Portland OR 31