Download presentation
1
Environmental Laboratory Accreditation Course for Radiochemistry: DAY THREE
Presented by Minnesota Department of Health Pennsylvania Department of Environmental Protection U.S. Environmental Protection Agency Wisconsin State Laboratory of Hygiene
2
Instrumentation & Methods: Laser Phosphorimetry, Uranium
Richard Sheibley Pennsylvania Dept of Env Protection
3
Laser Phosphorimeter UV excitation by pulsed nitrogen laser 337nm
Green luminescence at 494, 516 and 540 Excitation 3-4 X 10-9sec
4
Laser Phosphorimeter Measure luminescence when laser is off
Use method of standard addition
5
Instrumentation & Methods: Alpha Spectroscopy, Uranium
Lynn West Wisconsin State Lab of Hygiene
6
Review of Radioactive Modes of Decay
Properties of Alpha Decay Progeny loses of 4 AMU. Progeny loses 2 nuclear charges Often followed by emission of gamma
7
Review of Radioactive Modes of Decay, Cont.
Properties of Alpha Decay Alpha particle and progeny (recoil nucleus) have well-defined energies spectroscopy based on alpha-particle energies is possible Energy (MeV) Counts 4.5 5.5 a spectroscopy of the elements based on alpha-particle energies is possible Alpha spectrum at the theoretical limit of energy resolution
8
Instrumentation – Alpha Spectroscopy
Types of detectors Resolution Spectroscopy Calibration/Efficiency Sample Preparation Daily Instrument Checks
9
Types of detectors (Alpha Spectroscopy)
Older technology Diffused junction detector (DJD) Surface barrier silicon detectors (SSB) Ion Implanted Layers Fully depleted detectors State-of-the-art technology Passivated implanted planar silicon detector (PIPS)
10
PIPS Good alpha resolution due very thin uniform entrance window
Surface is more rugged and can be cleaned Low leakage current Low noise Bakable at high temperatures
11
Alpha Spectrometer Detector
An example of a passivated implanted planar silicon detector 600 mm2 active area Resolution of 24 keV (FWHM)
12
Alpha Spectrometer
13
Resolution Broadening of peaks is due to various sources of leakage current – “Noise” Low energy tails result from trapping of charge carriers which results from the incomplete collection of the total energy deposited Good resolution increases sensitivity (background below peak is reduced) Resolution of 10 keV is achievable with PIPS (controlled conditions)
14
Typical Alpha Spectrum
U232 Tracer
15
Calibration/Efficiency
Energy calibration Efficiency can be determined mathematically using Monte-Carlo simulation Efficiency can be determined using a NIST traceable standard in same geometry as samples Efficiency determination not always needed with tracers
16
Sample Preparation Final sample must be very thin to insure high resolution and minimize tailing. Also should stable & rugged The following mounting techniques are commonly used: Electrodeposition Micro precipitation Evaporation from organic solutions Organics must be completely removed There are some other techiniques but probably not applicable to radiochemistry Organics must be completely removed to avoid damaging the detectors
17
Sample Preparation Chemical and radiochemical interferences must be removed during preparation Nuclides must be removed which have energies close to the energies of the nuclide of interest, ie 15 to 30 keV Ion exchange Precipitation/coprecipitation techniques Chemical extractions Chemicals which might damage detector must be elimanted
18
Sample Preparation A radioactive tracer is used to determine the recovery of the nuclide of interest Since a tracer is added to every sample, a matrix spiked sample is not required
19
Sample Counting Mounts with a small negative voltage can be used to help attract the recoil nucleus away from the detector Reduces detector contamination
20
Sample Counting Analyst can choose distance from detector
Trade off is between efficiency & resolution Count performed slightly above atm. pressure to reduce contamination
21
Daily instrument checks
One hour background Pulser check Stability check
22
Instrumentation & Methods: Liquid Scintillation Counters & Tritium
Richard Sheibley Pennsylvania Dept of Env Protection
23
Liquid Scintillation Counter
Principle Beta particle emission Energy transferred to Solute Energy released as UV Pulse Intensity proportional to beta particle initial energy
24
Liquid Scintillation Counter
Low energy beta emitters Tritium – 3H Iodine – 125I, 129I, 131I Radon – 222Rn Nickel – 63Ni Carbon – 14C
25
Liquid Scintillation Counter
Energy Spectrum Isotope specific Beta particle Neutrino Total energy constant
26
Liquid Scintillation Counter
Components Vial with Sample + Scintillator Photomultipliers Multichannel Analyzer Timer Data collection & Output
27
Liquid Scintillation Counter
Variables Temperature Counting room Vial type glass vs. plastic Cocktail Energy window
28
Liquid Scintillation Counter
Other considerations Dark adapt Static Quenching
29
Liquid Scintillation Counter
Interferences Chemical Absorbed beta energy Optical Photon absorption
30
Liquid Scintillation Counter
Instrument Normalization Photomultiplier response Unquenched 14C Standard
31
Liquid Scintillation Counter
Performance assessment Carbon-14 Efficiency Tritium Efficiency Chi-square Instrument Background
32
Liquid Scintillation Counter
Method QC Background Reagent background Efficiency Method Quench correction
33
Tritium 3H (EPA & SM7500-3H B) Prescribed Procedures for Measurement of Radioactivity in Drinking Water EPA August 1980 Standard Methods 17th, 18th, 19th & 20th
34
Interferences Non-volatile radioactive material Quenching materials
Double distill – eliminate radium Static Fluorescent lighting
35
Tritium 3H Method Summary
Alkaline Permanganate Digestion Remove organic material Distillation Collect middle fraction Liquid Scintillation Counting
36
Calibration – Method Raw water tritium standard Background
Distilled Recovery standard Background Deep well water Distilled water tritium standard Distilled water to which 3H added Not distilled
37
Instrument Calibration
Calibrate each day of use Instrument Normalization Performance assessment Carbon-14 Efficiency Tritium Efficiency Instrument Background NIST traceable standards
38
Calculations 3H(pCi/L) = (C-B)*1000 / 2.22*E*V*F Where:
C = sample count rate, cpm B = background count rate, cpm E = counting efficiency F = recovery factor 2.22 = conversion factor, dpm/cpm
39
Calculations Efficiency: E = (D-B)/G Where:
D = distilled water standard count rate, cpm B = background count rate, cpm G = activity distilled water standard, dpm
40
Calculations Recovery correction factor F = (L-B) / (E*M) Where:
L = raw water standard count rate, cpm B = background count rate, cpm E = counting efficiency M = activity raw water standard (before distillation), dpm
41
Quality Control Batch Precision: Sample duplicate OR
Matrix spike duplicate Calculate relative percent difference Calculate control limits Should be < 20% Frequency 1 per 20
42
Quality Control, continued
Accuracy Laboratory fortified blank Matrix spike sample 2 – 10 Xs detection limit Reagent background |reagent background|< detection limit Instrument drift
43
Quality Control, continued
Daily control charts Acceptance limits Corrective action Preventative maintenance
44
Standard Operating Procedure
Written Reflect actual practice Standard format – EMMC or NELAC
45
Demonstration of Proficiency
Initial Method detection limit – MDL 40 CFR 136, Appendix B Alternate procedure 4 reagent blanks < Detection limit (DL) 4 laboratory fortified blanks (LFB) DL < LFB < MCL Evaluate Recovery and Standard Deviation against method criteria
46
Demonstration of Proficiency
Ongoing Repeat initial demonstration of proficiency Alternate procedure 4 Reagent blanks and laboratory fortified blanks Different batches Non-consecutive days Blank < Detection limit (DL) LFB met method precision and accuracy criteria
47
Instrumentation & Methods: Strontium 89, 90
Lynn West Wisconsin State Lab of Hygiene
48
Method Review Strontium 89, 90 EPA 905.0, SM 7500-Sr B
49
Radiochemical Characteristics
Isotope T1/2 Decay Mode MCL pCi/L 89Sr 50.55 days Beta 80 90Sr 29.1 years 8 90Y 64.2 hours N/A
50
Strontium (EPA 905.0, SM 7500-Sr B)
Prescribed Procedures for Measurement of Radioactivity in Drinking Water EPA August 1980 Standard Methods 17th, 18th, 19th & 20th
52
Strontium Chemistry Chemically similar to Ca
+2 oxidation state in solution Insoluble salts include: CO3 & NO3 “Real Chemistry”
53
Interferences Radioactive barium and radium Non-radioactive strontium
Precipitated as carbonate Removed using chromate precipitation Non-radioactive strontium Cause errors in recovery Calcium Removed by repeated nitrate precipitations
54
905.0 Method Summary Isolate Strontium Measure total strontium
Allow strontium to decay Isolate strontium 90 daughter – yttrium 90 Measure yttrium 90
55
905.0 Method Summary 1 L acidified sample Isolate Strontium
Add stable Sr carrier Precipitate alkaline and rare earths as carbonate Re-dissolve
56
905.0 Method Summary Isolate Strontium(continued)
Precipitate as nitrate Re-dissolve Precipitate as carbonate Determine chemical yield
57
905.0 Method Summary Measure total strontium activity Determine 90Sr
Yttrium in growth – 2 weeks Isolate yttrium Determine 90Y
58
905.0 Method Summary Determine 89Sr Calculated
Total strontium minus 90Sr
59
905.0 Method Summary Calculations include Recovery correction
In-growth correction – yttrium Total strontium Strontium 90 Decay correction – yttrium Isolation of Y to end of count time
60
Calculation total strontium
Total strontium activity (D) D = C / 2.22*E*V*R where: C = net count rate, cpm E = counter efficiency for 90Sr V = sample volume, liters R = fractional chemical yield 2.22 = conversion factor dpm/pCi
61
Calculations cont. See handout
62
Calculations cont. Verify computer programs
Decay constants and time intervals must be in the same units of time Minimum background count time should be equal to the minimum sample count time
63
Instrumentation Low background gas flow proportional counter
P-10 counting gas (10% CH4 & 90% Ar) Due to in growth and short half-life of 90Y, time is critical
64
Instrument Calibration
Isotope specific calibration 89Sr 90Sr 90Y Use NIST traceable standards Perform yearly or after repairs
65
Quality Control Batch Precision: Sample duplicate OR
Matrix spike duplicate Calculate relative percent difference Calculate control limits Should be < 20% Frequency 1 per 20
66
Quality Control, continued
Batch Accuracy Laboratory fortified blank Matrix spike sample 2 – 10 Xs detection limit Reagent background |reagent background|< detection limit Instrument drift
67
Quality Control, continued
Daily control charts Acceptance limits Corrective action Preventative maintenance
Similar presentations
© 2024 SlidePlayer.com. Inc.
All rights reserved.