1 Introducing the Varian 356-LC Refractive Index Detector

2 Introduction The Varian 356-LC detector is being introduced to replace the ProStar 355 detector. The 356-LC detector was designed and built at Polymer Labs, UK. The Varian 356-LC is a high performance refractive index detector designed to meet virtually all customer needs for refractive index detection. This universal detector is ideal for the analysis of non-absorbing compounds such as polymers, carbohydrates, alcohols, and fatty acids.

3 Overview The Varian 356-LC has been designed to integrate into the ProStar range of detectors. Status LED Showing Power, Purge, zero and error Leak drain Hinges for “Barbie” shower 211 mm 477 mm 296 mm Access panel to connections On/Off switch

4 Overview Outlet Inlet Waste for internal leaks

5 Principle of Operation
Refractive index detectors are universal detectors designed for analysis of compounds that do not strongly absorb UV light and, therefore, cannot be assayed with a UV detector. The usual compounds analyzed are sugars, polymers, aliphatic amines, organic acids, and alcohols. The Varian 356-LC is functionally identical to other deflection type refractive index detector, it comprises: LED light source, 6 µL flowcell with reference and sample compartments split photodiode detector The complete assembly is housed in a thermally controlled block to maintain baseline stability.

6 Principle of Operation
Generic diagram of the Varian 356-LC Refractive Index Detector

7 Principle of Operation
When both cell compartments contain the same solvent, the light beam is positioned equally on the two detector diodes I1 and I2 When solvent containing a compound passes through the sample side of the cell the difference in refractive index causes the light beam to be proportionally deflected (according to Snell’s refraction law). The change in light intensities between I1 & I2, is proportional to concentration and refractive index of the sample solution. I1 I2 Refractive index is a bulk property of the mobile phase and is very sensitive to changes in ambient temperature RI detectors are also sensitive to flow-rate changes and cannot be used for gradient elution.

8 Internal Flow Path Normal Operating Conditions: Purge Conditions:
The incoming solvent passes into & out of the sample side of the flowcell, then to waste Purge Conditions: The incoming solvent passes into sample & reference side of the flowcell, before going to waste The Varian 356-LC has only 15 µL of volume between inlet and cell to keep dispersion to a minimum The maximum recommended flow rate permitted through the Varian 356-LC Detector is 10 mL/min

9 Operational Parameters
Purge When Purge is selected, the purge indicator light is highlighted, and solvent flow is diverted through the reference side of the RI cell, instead of the sample side; before going to waste. The purge time can be adjusted to compensate for different solvents & eluent flow rates. Auto-Zero The detector output is re-set to zero when the auto-zero function is selected. An error will occur if the auto-zero process takes longer than 5 minsn Response Time The Varian 356-LC has a default response time of 1.0 second, with a settable range of 0.1 –5.0 seconds.

10 Operational Parameters
Sensitivity The sensitivity of the Varian 356-LC can be adjusted to provide greater flexibility. The three possible sensitivity settings are described below: Low - Suitable for highly concentrated samples (e.g. Prep LC) Standard - Default setting, ideal for the majority of LC applications High - Suitable for samples of low concentration in order to maximize signal‑to‑noise. Temperature The temperature of the Varian 356-LC can be set between °C (1 °C increments) or turned off. The default value is 35 °C. It is recommended that the temperature should be set at least 5 °C above ambient temperature in order to ensure a stable thermal environment.

11 Operational Parameters
Polarity The output of the Varian 356-LC refractive index detector can be either positive or negative; therefore it is essential that the polarity of the detector signal be configured. The polarity parameter is used to invert the sign of the refractive index data. A positive polarity leaves the data unaffected, whilst a negative polarity essentially inverts the chromatogram. Baseline Offset The baseline offset of the Varian 356-LC can be set between the ranges –100 to +100 µRIU.

12 Not Ready (i.e. changing temp)
Front Panel Status The Varian 356-LC, like other Varian modules, has no front-panel display or hard-key control. Full functionality is provided by Galaxie CDS, which gives the user greater flexibility and control than the ProStar 355. Three indicator lights are located at the top left of the front panel of the module to display the status of the instrument Status Color Ready Not Ready (i.e. changing temp) Error Green Orange Red Auto-zero on Purge on

13 Galaxie Control The Varian 356-LC can be connected directly to the PC via an RS232 connection, for full Galaxie Control Detector output and status quickly accessed in overview window Status window displays all current detector conditions A/Z & purge function easily accessed

14 Galaxie Control All detector parameters can be changed via the configuration window Send button downloads settings to detector

15 Key Benefits Sensitivity Baseline Stability Superior Peak Shape
The Varian 356-LC provides three sensitivity settings, HIGH, STD & LOW, that can be selected according to the type of analysis. For example, the LOW setting is ideal for preparative LC in order to prevent the signal from saturating at the high concentrations typically encountered. Baseline Stability Flow cell thermostating in the Varian 356-LC eliminates temperature effects on the RI baseline. Stable baselines mean lower detection limits for both analytical and preparative applications. Temperatures are controlled in one degree increments from 30 °C to 50 °C. Superior Peak Shape The custom designed 6µL ‘flow-efficient’ cell for minimal peak dispersion delivers outstanding peak shape, and superior resolution.

16 Key Benefits: Superior Peak Shape
At 1 mL/min the Varian 356-LC peak width* is 50% narrower than the ProStar 355 Varian 356-LC ProStar 355 *measure at ½ height

17 Key Benefits: Superior Resolution
The Varian 356-LC provides better resolution between peaks than the ProStar 355. Resolution 1.17 This is beneficial when analysing compounds of similar properties under isocratic conditions. Resolution 1.18

18 Key Benefits: Automating the 356-LC
Fully Automated via Galaxie Galaxie Software from Varian provides control and automation capabilities. The detector can be automatically purged prior to analysis for overnight operation. The Varian 356-LC can be externally controlled, permitting a device (for instance, a pump) to automatically activate functions such as electronic optical zero and reference cell flush. Digital Data Acquisition Unlike the ProStar 355 the Varian 356-LC offers the user a digital data signal, which can be taken directly into a PC, using an RS232 cable, without the need for an MIB800 module.

19 Automation via Galaxie Software
Detector settings can be saved in a Galaxie method along with pump program for individual applications Automatic Purge and A/Z are possible within a Galaxie method Varian 356-LC can be automated using a sequence of Galaxie methods

20 Applications: Carbohydrates in Fruit Juice
A cost-effective and simple separation to identify food adulteration in fruit juice can be achieved using the Varian 356-LC and PL-Hiplex columns.

21 Applications: Maltodextrin in Sports Drink
The low dispersion characteristics of the Varian 356-LC provides excellent resolution for maltodextrin analysis in sports beverages.

22 Applications: Epoxy Resins
The Varian 356-LC is ideal for GPC analysis, when combined with Galaxie and Cirrus GPC software.

23 Product Positioning The Varian 356-LC and PL-ELS 2100 ELSD are both universal detectors and should have distinct positions in the market place: ELSDs are typically more expensive than RID, and there is no justification for the higher cost and added complexity of method development if the RI detector can already adequately perform the analysis. The Varian 356-LC should be positioned as a high performance general use refractive index detector with low noise, excellent thermal control & advanced capabilities such as automatic purge & auto zero using Galaxie software.

24 Varian 356-LC Features & Benefits
Feature Benefit Low dead-volume flow-cell Improved resolution and sensitivity Active temperature compensation Lower base-line drift and improved reproducibility Galaxie Control Improved productivity through automated, unattended sequence runs Easy to use and consistent user interface Improved traceability using Galaxie operation logs Digital data output Simple to set up Better value – do not need an expensive analog to digital interface board Long-Life LED Reduced maintenance and quick start-up as the LED is always on and has a indefinite life time

25 Varian 356-LC Specifications
RI range (RIU) 1.0~1.75 Range (µRIU) x10-6 RIU/FS Linearity (µRIU) 600 Sensitivity (mV/µRIU) LOW (2), STD (4), HIGH (8) Short-term noise (nRIU) <5.0 Drift (nRIU/hr) <250 Response time (s) 0.1–5.0 Temperature control (°C) OFF, 30–50 (1 °C increments) Cell volume (µL) 6.0 Light source LED 880 nm Analog output 1V FSD Digital output 24 bit (10 Hz) via serial port Flow rate range (mL/min) 0.1–10 Dimensions (w x d x h) (unpacked) 296 x 475 x 212 mm  PC control (software) Galaxie Software 200mm 200mm 415mm 450mm 450mm

26 Ordering Information Varian 356-LC Detector PL0880-0000 Packing List
Manual for Varian 356-LC detector PL 1.5M PTFE Outlet Tubing assembly PL Aux I/O Cable (5 m) PL RS232 Communication cable PL Green Half Moon Door Door Cap Accessories Analogue Cable PL Mains Inlet Fuses (pk 5) PL Field Upgrades LED assembly PL Flow Cell PL Cell Seals (Top) PL Cell Seals (Bottom) PL Purge Valve PL

27 Product Information Folder (PIF)
* The Varian 356-LC PIF contains additional information *