ARRT & Other DIGITAL Terms Defined April 2009 Wk 9/10 RT 255
2
Display Workstations Conventional film/screen radiography uses large multiviewer lightboxes. With early PACS, radiologists thought that they needed 4-6 monitors. Now, the number of monitors has dropped to an average of 2. Development of viewing software and better hardware.
3
Name the 3 types of monitors
4
Two major types of monitors with a third type gaining acceptance: CRT (?) LCD (?) Plasma screen
5
3 types of monitors CRT (cathode ray tube) LCD (liquid crystal display) Plasma screen
6
Medical Use of Monitors Most medical monitors used to display radiographic images are monochrome high-resolution monitors. Early displays used were primarily CRTs, but as the LCD technology has gotten better, more LCDs are taking the place of the older CRTs.
7
Image Display= MONITORS viewing conditions – (i.e.,luminance,ambient lighting) spatial resolution contrast resolution/dynamic range DICOM gray scale function window level and width function ARRT DEFINITIONS
8
viewing conditions luminance,ambient lighting How does this affect viewing images ? Different monitors ARRT DEFINITIONS
9
viewing conditions luminance,ambient lighting How does this affect viewing images? Surrounding light impacted what was seen on image – now : With different monitors: LCD gives more light. LCD can be used in areas with a high amount of ambient light. “dark rooms” not necessary
10
Technologist workstation monitors are used in brightly lighted areas. So monitor luminance, the brightness of a monitor display, is an important consideration. Also, the monitor must allow a technologist to visualize enough detail to discern motion and that the recorded lines are sharp and visible.
11
Monitor Advantages and Disadvantages Most consumers want a monitor that can provide the highest resolution for the best price. Most radiology applications have traditionally used the CRT because of its superior resolution, but as the LCD technology has progressed, more and more departments are buying the more slim and light weight LCDs to replace the bulky CRTs.
12
MONITORS:Display Workstations Early PACS reading rooms required supplemental air-conditioning to offset the heat from multiple CRTs. Resolution and orientation of the monitor is also a factor in determining which type of monitor is to be used. Most cross-sectional imaging is read on a 1K square monitor. Most computed radiography (CR) and digital radiography (DR) images are read on at least a 2K portrait monitor
13
LCD 1.3 megapixels to 5 megapixels. mammography imaging = at least 5 megapixel resolution is required.
14
CRT The CRT is the most popular monitor on the market. It consists of a cathode and anode within a vacuum tube. Cathode boils off a cloud of electrons, and then a potential difference is placed on the tube. A stream of electrons is sent across to the anode, which in the case of the monitor is a sheet of glass coated with a phosphor layer.
15
CRT Electrons strike the phosphor on the glass. Impact causes the glass to emit a color based on the intensity of impact and area that the electrons strike. The electrons interact with a red, green, or blue dot to form the color and image that is being sent from the video card signal.
16
CRT Electron beam starts in the upper left hand corner and scans across the glass from side to side and top to bottom. Once beam reaches the bottom, it starts back over at the top left. Most monitors have 350 lines to be scanned.
17
LCD LCD produces images by shining or reflecting light through a layer of liquid crystal and a series of color filters. Monitor is two pieces of polarized glass with a liquid crystal material between the two. Light is allowed through the first layer of glass.
18
LCD When a current is applied to the liquid crystal, it aligns and allows light in varying intensities through to the next layer of glass through color filters. Light forms the colors and images seen on the display.
19
LCD: Display Workstations LCD has dropped in price and has risen in quality. LCD will soon take over PACS display market because of its size, resolution, and lack of heat production. LCD requires less maintenance.
20
CRT vs LCD CRT Luminance higher in the center Lower measurable black levels Phosphor granularity adds to spatial noise Viewable area smaller than stated size Better color reproduction More responsive on redraw More rugged Aspect ratio 4:3 LCD Less veiling glare Consumes less energy Increased spatial resolution Larger viewing area by described size Display limited to designed resolution Can position screen Smaller footprint and lighter Widescreen aspect ratio 16:9
21
Monitors – more terms… Aspect ratio Ratio is the width of the monitor to the height of the monitor. Most CRT monitors have an aspect ratio of 4:3, and LCD monitors have a ratio of 16:9. Viewable area The viewable area is measured from one corner of the display to the opposite corner diagonally.
22
Plasma Display Plasma monitors are new to the consumer market. They have been used in government and military applications since the late 1960s. They are made up a many small fluorescent lights that are illuminated to form the color of the image. Monitor varies the intensities of the various light combinations to produce a full range of color.
23
Image Display= MONITORS viewing conditions – (i.e.,luminance,ambient lighting) spatial resolution contrast resolution/dynamic range DICOM gray scale function window level and width function
24
MONITORS: Spatial Resolution Spatial resolution refers to the amount of detail present in any image. Phosphor layer thickness and pixel size determines resolution in CR. The thinner the phosphor layer is, the higher resolution. Film/screen radiography resolution at its best is limited to 10 line pairs per millimeter (lp/mm). CR resolution is 2.55 lp/mm to 5 lp/mm, resulting in less detail. ARRT DEFINITIONS
25
MONITOR RESOLUTION
26
digital image characteristics spatial resolution sampling frequency DEL (detector element size) receptor size and matrix size image signal (exposure related) quantum mottle SNR (signal to noise ratio) or CNR (contrast to noise ratio) ARRT definitions
27
Spatial Resolution A radiograph typically does not show soft tissue structures Digital image shows not only the soft tissue but also the edge of the skin. Giving the appearance of more detail.
28
MONITORS contrast resolution /dynamic range Appearance of more detail is due to the wider dynamic recording range / contrast resolution and does not mean that there is additional detail Because so many more densities are recorded in CR (wide dynamic range), images appear more detailed.
29
contrast resolution The contrast resolution of a monitor is the difference between the maximum and minimum luminance of the display.
30
Exposure Latitude or Dynamic Range CR and DR Contain a detector that can respond in a linear manner as compared to S shape of H&D Curve Exposure latitude is wide, allowing the single detector to be sensitive to a wide range of exposures.
31
Why do digital systems have significantly greater latitude? Linear response give the imaging plates greater latitude Area recieving little radiation can be enhanced by the computer Higher densities can be separated and brought down to the visibile density ranges (Brightness in DR replaces density)
32
Monitors - RESOLUTION Pixel is a basic picture element on a display. A pixel is “any of the small discrete elements that together constitute an image.” Resolution -# of pixels contained on a display Relationship: More pixels in an image, the higher the resolution & more information that can be displayed. Resolution also is defined as the process or capability of distinguishing between individual parts of an image that are adjacent.
33
Nyquist frequency The highest spatial frequency that can be recorded by a digital detector. is determined by the pixel pitch. The Nyquist frequency is half the number of pixels/mm. A digital system with a pixel density of 10 pixels/mm would have a Nyquist frequency of 5 line pair/mm.
34
sampling frequency The frequency that a data sample is acquired from the exposed detector. It is expressed in pixel pitch and pixels per mm. Sampling frequency may be determined by receptor size depending on the vendor. KODAK 8x10 better detail than 14x17 Use of the smallest imaging plate possible for each exam results in the highest sampling rate. When the smallest possible imaging plate is selected, a corresponding matrix is used by the computer algorithm to process the image. ARRT definitions
35
Pixel “picture element,” the smallest area represented in a digital image. A digital radiography image consists of a matrix of pixels which is typically several thousand pixels in each direction. Pixel density – A term that describes the number of pixels/mm in an image. Pixel density is determined by the pixel pitch.
36
a pixel or picture element. The typical number of pixels in a matrix range from about 512 × 512 to 1024 × 1024 and can be as large as 2500 × 2500. The more pixels there are, the greater the image resolution. The image is digitized by position (spatial location) and by intensity (gray level). DEL (detector element size) receptor size and matrix size
37
DELs – detector elements DELs collect electrons that are extracted from the detector assembly and converted into a digital value by an ADC. That process creates the image that displays on our monitor. DEL size controls the recorded detail, or spatial resolution, for the flat- panel device. The technologist can’t change the size of the DEL, which is fixed for that piece of equipment..
38
Detective Quantum Efficiency Known as the fill factor, the larger the area of the TFT photodiodes, the more radiation can be detected and the greater amount of signal generated. Consequently, the greater the area of the TFT array, the higher the DQE. Over 1 million pixels are read & converted
39
FILL FACTOR A field-effect transistor (FET) or silicon TFT Isolates each pixel element Reacts like a switch to send the electrical charges to the image processor
40
Detective Quantum Efficiency How efficiently a system converts the x-ray input signal into a useful output image is known as detective quantum efficiency, or DQE. DQE is a measurement of the percentage of x- rays that are absorbed when they hit the detector.
41
Detective Quantum Efficiency In other words, CR records all of the phosphor output. Systems with higher quantum efficiency can produce higher-quality images at a lower dose. Indirect and direct DR capture technology has increased DQE over CR. However, DR direct capture technology, because it does not have the light conversion step and consequently no light spread, increases DQE the most.
42
SNR (signal to noise ratio) or CNR (contrast to noise ratio) SNR (signal to noise ratio): there is always a very small electric current flowing in any circuit - is called background electronic noise. It is similar to the fog on a radiograph in that it conveys no information and serves only to obscure the electronic signal. CNR (contrast to noise ratio): measure for assessing the ability of imaging an procedure to generate clinically useful image contrast. gives an objective measure of useful contrast. ARRT definitions
43
Image Display spatial resolution contrast resolution/dynamic range
44
What is a 3-D array of Pixels ? A voxel (a volumetric pixel) is a volume element, representing a 3-D value space. A pixel which represents 2D image data.
45
Pixel Pitch The space from the center of a pixel to the center of the adjacent pixel. It is measured in microns (μm). Pixel pitch is determined by sampling frequency for cassette-based PSP systems and by DEL spacing for TFT flat panel.
46
The Nyquist Theorem Theorem states that when sampling a signal, the sampling frequency must be greater than twice the bandwidth of the input signal so that the reconstruction of the original image will be nearly perfect. At least twice the number of pixels needed to form the image must be sampled. If too few pixels are sampled, the result is a lack of resolution.
47
Monitors: Display Workstations Pixels are arranged in a matrix. Common screen resolutions found on today’s monitors are the following: 1280 × 1024 (1K) 1600 × 1200 (2K) 2048 × 1536 (3K) 2048 × 2560 (5K)
48
Monitors – DOT PITCH Dot pitch is the measurement of how close the dots are located to one another within a pixel. The smaller the dot pitch of a display, the finer the resolution. Dot pitch may be expressed as aperture grille pitch or slot pitch.
49
Monitors – REFRESH RATE Refresh rate or vertical scanning rate Refresh rate is a measure of how fast the monitor rewrites the screen or the number of times that the image is redrawn on the display each second. Refresh rate helps to control the flicker seen by the user. The higher the refresh rate, the less flicker will be seen. Most refresh rates on today’s computer are set between 60 and 75 Hz; the image is redrawn 60 to 75 timers per second.
50
Image Display= MONITORS viewing conditions – (i.e.,luminance,ambient lighting) spatial resolution contrast resolution/dynamic range DICOM gray scale function window level and width function
51
MONITOR RESOLUTION DICOM gray scale function – calibration of monitors to the same standard & communication of images Window level affects brightness (density) - B Window width function affects contrast/gray scale - A “windowing and level”
52
MONITOR Image Manipulation and Enhancement Functions Window/level This is a default function of the left mouse button. By depressing and holding down the mouse button and moving the mouse up and down and left and right, the window and level can be adjusted. Window (width) represents the range of gray values. Level represents the center value of the range. A change in the window and level appears to change the brightness and contrast of the image.
53
Image Manipulation and Enhancement Functions Other Tools: Annotations Annotations are NOT to be used to label left or right to indicate the patient’s side. Annotations are used to indicate prone or supine, 30 minutes, upright or flat. Any other image information is appropriate. Radiologist will place arrows or circles around pathologic or questionable areas.
54
MONITOR Image Manipulation and Enhancement Functions Pan, zoom, and magnify Tools are used primarily by the radiologist to increase the size of an area on the image. Magnify usually magnifies a square area of the image Pan and zoom functions are usually used together. Image is first zoomed up to the desired magnification level then Pan icon is activated. Zoomed image can be moved around to view the different areas of the image.
55
Image Manipulation and Enhancement Functions Measurements Size of a pixel is a known so the software has the ability to measure structures on the image based on this. Angle measurement. –Can give an angle measurement between two structures –Commonly used when reading spine studies
56
Image Manipulation and Enhancement Functions Measurements –Region of interest –Measurement tool determines the pixel intensity of a certain area. –Each type of tissue or fluid has a different intensity of reading. –Radiologist can make a determination whether something is solid or fluid. –Each pixel can have a gray level between 0 (2 0 ) and 4096 (2 12 ). The gray level will be a factor in determining the quality of the image
57
DR
58
Monitor :Navigation Functions Hanging protocols Can be viewed: 1:1 4:1 etc Protocol can also be specified to show the previous exam on one monitor and the current exam on the other Once set, the most efficient study navigation is determined.
59
Image Management Functions Patient demographics must be correct. –If demographics are not correct at the archive level, the images could be lost. Changes should only be made when the information is absolutely known to be wrong. Many hospitals allow only certain persons the access to change demographics just to keep the errors to a minimum.
60
Image Management Functions Query/retrieve icon Used to retrieve on demand any studies from the archive Allows user to query a study on multiple fields –Patient’s name or identification –Date of service –Modality –Diagnosis code or comment field
61
WINDOW LEVEL / WIDTH Which one controls Denisty (brightness) ? Contrast What else control these in DIGITAL IMAGING?
62
Enhanced Visualization Image Processing Takes image diagnostic quality to a new level Increases latitude while preserving contrast Process decreases windowing and leveling Virtually eliminates detail loss in dense tissues
63
Contrast / Denisty (Brightness) CR and DR Kilovoltage peak still influences subject contrast, but radiographic contrast is primarily controlled by an image processing look-up table. LUT Milliampere-second setting has more control over image noise, whereas density is controlled by image-processing algorithms.
64
Physician Review Stations Monitors Step-down model of the radiologist’s reading station (lower res) Some functions reduced One of the most important features is ability to view current and previous reports with images. Many vendors are integrating the RIS functions with PACS software.
65
Technologist QC Stations review images after acquisition but before sending them to the radiologist May be used to improve or adjust image-quality characteristics May be used to verify patient demographic information Placed between the CR and DR acquisition modalities as a pass- through to ensure that the images have met the departmental quality standard
66
Technologist QC Station Generally has a 1K monitor Does not have the resolution capabilities of the radiologist’s reading station Care required of technologist when manipulating images not to change the appearance too much from original acquired image
67
The File Room Workstation Workstation may be used to look up exams for a physician or to print copies of images for the patient to take to an outside physician. Many hospitals are moving away from printing films because of the cost. Hospitals are moving toward burning compact disks (CDs) with the patient’s images.
69
Image Receptors digital image characteristics –spatial resolution –sampling frequency –DEL (detector element size) –receptor size and matrix size –image signal (exposure related) –quantum mottle –SNR (signal to noise ratio) or –CNR (contrast to noise ratio) ARRT definitions
70
image signal (exposure related) Exposure Indicators The amount of light given off by the imaging plate is a result of the radiation exposure that the plate has received. The light is converted into a signal that is used to calculate the exposure indicator number, which is a different number from one vendor to another.
71
Digital artifacts Grid Lines: Appear as grid cutoff. Moire (Aliasing) –wavy artifact occurs because the grid lines and the scanning laser are parallel. –When the spatial frequency is greater than the Nyquist frequency Maintenance (e.g., detector fog): When errors occur in equipment performance, corrective action must occur. These corrections will generally be done by service personnel employed by the vendor. non-uniformity, erasure - blub problems
72
How else can Morie OCCUR?
73
quantum mottle ARRT definitions failure of an imaging system to record densities usually caused by a lack of x- ray photons. “PHOTON STARVED” KVP & MAS HOW IS THIS AFFECTED IN DR / CR?
74
Image Receptors digital image characteristics –spatial resolution –sampling frequency –DEL (detector element size) –receptor size and matrix size –image signal (exposure related) –quantum mottle –SNR (signal to noise ratio) or –CNR (contrast to noise ratio) ARRT definitions
75
Image Acquisition and Readout PSP (photo-stimulable phosphor) flat panel detectors –(direct and indirect)
76
CR Imaging Plate Construction Image recorded on a thin sheet of plastic known as the imaging plate - PSP Consists of several layers: Phosphor?
77
Imaging Plate Phosphor? BARIUM FLUORO HALIDE WITH A EUROPIUM BASE
78
Two types of digital radiography Indirect capture DR Machine absorbs x-rays and converts them to light. CCD or thin-film transistor (TFT) converts light to electric signals. Computer processes electric signals. Images are viewed on computer monitor. Digital Radiography
79
Direct capture DR Photoconductor absorbs x-rays. TFT collects signal. Electrical signal is sent to computer for processing. Image is viewed on computer screen. Digital Radiography
80
Image Acquisition and Readout flat panel detectors Phosphors? –direct = Am SELENIUM – indirect = Am SILICON
81
Digital Systems electronic collimation grayscale rendition or look-up table (LUT) edge enhancement/ –noise suppression contrast enhancement system malfunctions (e.g., ghost image, banding, erasure, dead pixels, readout problems, printer distortion)
82
Image Data Recognition and Preprocessing shutter Agfa uses the term collimation, Kodak uses the term segmentation, and Fuji uses the phrase “exposure data recognition.” All systems use a region of interest to define the area where the part to be examined is recognized and the exposure outside the region of interest is subtracted.
83
Excessive exposure to receptor without rescaling. Excessive exposure to receptor with rescaling. So in essence, rescaling provided an acceptable image, despite an excess level of exposure to the receptor. What about the dose to the patient?
84
Rescaling the image pixel values to appear appropriate, display properly, can lead to overexposing a patient. The visual cue to the technologist that overexposure has occurred isn’t present. With an analog system, a technologist would have seen the image on the left as it came out of the processor and used the excessive density of the image as a visual cue to repeat the image. Rescaling forces a technologist to look elsewhere for signs that a proper exposure was used to produce an image.
85
Dose creep refers to the potential to gradually increase patient exposure over time. However, a technologist lacks visual feedback that additional radiation is being used to produce the images
86
DAP The dose area product (DAP) meter is a device that may be interlinked with the x-ray unit to determine the actual patient entrance skin exposure dose with accurately calibrated equipment. Currently, no standards are established for using a dose area product meter These two radiographs show the difference in entrance skin exposure measured by the DAP meter. The area with a smaller exposure field size carries a lower exposure without reducing image quality
87
Image Evaluation: Brightness and Contrast in Images Exposure Field Recognition Error Gross Overexposure Excessive Scatter Striking the Receptor Excessive Fog on the Receptor Grid Cutoff Intra-Field or Off-Focus Radiation Wrong Menu Selection
88
ARRT SPECS - DIGITAL PACS HIS (hospital information system) - work list RIS (radiology information system) DICOM Workflow (inappropriate documentation, lost images, mismatched images, corrupt data)
89
PACS Image is stored on a computer: retrieval and viewing system for digital imaging examinations. The PACS software provides “tools” that enable a PACS operator to manipulate images. The basic image manipulations possible are: magnification and minification of an image; altering image brightness and contrast; and annotating the image using text, symbols, lines and arrows.
90
Picture Archival and Communication Systems Networked group of computers, servers, and archives to store digital images Can accept any image that is in DICOM format Serves as the file room, reading room, duplicator, and courier Provides image access to multiple users at the same time, on- demand images, electronic annotations of images, and specialty image processing
91
RIS Data Storage Less ERRORS A radiology information system, or RIS, is a data system for patient-related functions in the radiology department. Examples of functionality a RIS provides within a radiology department include (but are not limited to): scheduling appointments; collecting and displaying orders for radiologic examinations; storing and displaying patient data; tracking patients providing patient and order data to a PACS; storing and distributing radiology reports; providing billing services; and providing a database to track and project trends.
92
HIS A hospital information system, or HIS, is a paper and/or data system that manages the administrative, financial and clinical information necessary to operate a hospital or health care system
93
Health Level 7 (HL7) HL7 are the software standards established for exchanging electronic information in health care. PACS is mainly concerned with images and data associated with images, whereas HL7 sets standards for transmitting text-based information throughout a medical center.
94
Next ODIA - Register Review Modules 5 & 6 (& 11 ??) Do study test at end of modules - ?Print out certificate of completion