Optimization of the production and insertion of 3-dimensional polyimide electrodes into rats Lab of Professor Patrick Rousche Hananeh Esmailbeigi Daniela Aizpitarte
2
Overall Picture Create a 3d electrode Will be filled with neurotrophic factors to increase neuron growth in the region Goal is to decrease noise and measure strong long term signals Has many uses for neural prosthetics Made of polyimide-a very flexible polymer
3
1 st Project: Creating a rolling system Current system is time consuming Requirements: Steady the two needles Keep the electrode perpendicular to the needle Be able to roll electrodes quickly Allow the needles to be removed after use Image from www.sanimedtech.com
4
Final Electrode Roller
5
2 nd Project: Implantation A hypodermic needle needs to accompany the electrode Rolled tip is too blunt to penetrate the pia without buckling Needs to be simple to make Can’t increase damage to the surrounding tissue
6
Injection Needle Idea How it works: The two halves are gently separated One half is pulled up at a time Made of stainless steel Advantages: Simple construction Disadvantages: Hard to control
7
Final Inserter Design Glass capillary tube 4mm long Tip sanded down to a 45 degree angle Success when used in surgery
8
Force of Implantation A glass capillary tube at a 45 degree angle was sanded down and inserted 4mm into 0.5% agar gel Three different relatives speeds were tested: fast, medium, slow Fast= 5-7mm/s Medium=2.5-3.5mm/s Slow=0.3-0.5mm/s
9
Fast Trials
10
Medium Trials
11
Conclusions Inserting faster exerts less force (speed dependent) Diameter had to be taken into account Steel and glass are of the same hardness, and should exert the same force Average peak (mN) Standard Deviation (mN) Fast 7.37 0.97 Medium 10.5 1.39 Steel at Medium 7.63 1.03 Per surface area of 1.2mm diameter circle
12
Electrode sticking test: Water
13
Water dish test
14
Conclusions Water is easily lifted up the electrode and gives varying measurements Force changes on the order of 0.01 mN ▫Too small to have great significance
15
Wet electrode testing: Steel and glass
16
Visual Comparison
17
GlassSteel Average Peak Height -0.158mN Standard Deviation 0.076mN Average Max Peak -0.160mN Standard Deviation 0.056mN
18
Agar gel and electrode
19
Average
20
AgarGlass and Steel Average Peak Height -.217mN Standard Deviation 0.062mN Average Peak Height -0.158mN Standard Deviation 0.076mN Average Max Peak -0.160mN Standard Deviation 0.056mN
21
Conclusion The effect of water on the electrode is minimal The electrode prefers the agar over glass capillary tube ▫Reason why the surgery worked ▫Actually pulls away the agar gel Electrode rolling technique overall didn’t work better than hand rolling
22
Acknowledgments I would like to thank the financial support from the EEC-NSF Grant # 0755115 for funding the REU program. Also, I would like to thank Professor Rousche and the rest of the REU program for giving me this project.