Download presentation
1
Basic Electron Microscopy
Arthur Rowe The Knowledge Base at a Simple Level
2
Introduction These 3 presentations cover the fundamental theory of electron microscopy In presentation #2 we cover: lens aberrations and their importance how we correct for lens astigmatism limits to ultimate resolution of the TEM Interactions of electrons with matter
3
aberrations of electromagnetic lenses
the most important ones to consider are: • spherical aberration • chromatic aberration • astigmatism
4
spherical aberration object plane
• arises because a simple lens is more powerful at the edge than at the centre • is not a problem with glass lenses (can be ground to shape) • disc of minimum confusion results instead of point focus: • is not correctable for electromagnetic lenses
5
coping with spherical aberration
• disc of minimum confusion has diameter given by: d = C {C = constant} • hence reducing gives a large reduction in d • but for optimal resolution we need large ! • best compromise is with = 10-3 radians (= f/500) • gives resolution = 0.1 nm - can not be bettered
6
chromatic aberration • light of differentbrought to different focal positions • for electrons can be controlled by fixed KV and lens currents • but of electrons can change by interaction with specimen ! • rule of thumb: resolution >= (specimen thickness)/10
7
astigmatism • arises when the lens is more powerful in one plane
minimal confusion • arises when the lens is more powerful in one plane than in the plane normal to it • causes points to be imaged as short lines, which ‘flip’ through 90 degrees on passing through ‘focus’ (minimal confusion)
8
astigmatism - arises from:
• inherent geometrical defects in ‘circular’ bore of lens inherent inhomogeneities in magnetic properties of pole piece build-up of contamination on bore of pole-piece and on apertures gives rise to non-conducting deposits which become charged as electron strike them hence astigmatism is time-dependent and cannot be ‘designed out’ inevitably requires continuous correction
9
astigmatism - correction:
• with glass optics (as in spectacles) astigmatism is corrected using an additional lens of strength & asymmetry opposed to the asymmetry of the basic (eye) lens with electron optics, same principle employed: electrostatic stigmator lens apposed to main lens strength & direction of its asymmetry user-variable only the OBJECTIVE lens needs accurate correction correction usually good for 1-2 hours for routine work
10
The TEM Column Gun emits electrons Electric field accelerate
Magnetic (and electric) field control path of electrons Electron 200KeV 2x10-12 m Resolution normally 200KeV 2 x m 2Å
11
depth of focus - depth of field
• depth of useful focus (in the specimen) is primarily limited by chromatic aberration effects the absolute depth of focus is larger than this: for all practical purposes, everything is in focus to same level . . . So one cannot rack through focus (as in a light or even scanning electron) microscope depth of field (in the image plane) is - for all practical purposes infinite
12
when electrons hit matter ..
13
when electrons hit matter ..
(1) they may collide with an inner shell electron, ejecting same > the ejected electron is a low-energy, secondary electron - detected & used to from SEM images > the original high-energy electron is scattered - known as a ‘back-scattered’ electron (SEM use) > an outer-shell electron drops into the position formerly occupied by the ejected electron > this is a quantum process, so a X-ray photon of precise wavelength is emitted - basis for X-ray microanalysis
14
when electrons hit matter ..
15
when electrons hit matter ..
(2) they may collide or nearly collide with an atomic nucleus > undergo varying degree ofdeflection (inelastic scattering) > undergo loss of energy - again varying > lost energy appears as X-rays of varying wavelength > this X-ray continuum is identical to that originating from an X-ray source/generator (medical, XRC etc) > original electrons scattered in a forward direction will enter the imaging system, but with ‘wrong’ l > causes a ‘haze’ and loss of resolution in image
16
when electrons hit matter ..
17
when electrons hit matter ..
(3) they may collide with outer shell electrons > either removing or inserting an electron > results in free radical formation > this species is extremely chemically active > reactions with neighbouring atoms induce massive change in the specimen, especially in the light atoms > this radiation damage severely limits possibilities of EM > examination of cells in the live state NOT POSSIBLE > all examinations need to be as brief (low dose) as possible
18
when electrons hit matter ..
19
when electrons hit matter ..
(4) they may pass through unchanged > these transmitted electrons can be used to form an image > this is called imaging by subtractive contrast > can be recorded by either (a) TV-type camera (CCD) - very expensive (b) photographic film - direct impact of electrons Photographic film > silver halide grains detect virtually every electron > at least 50x more efficient than photon capture !
20
when electrons hit matter ..
‘beam damage’ occurs: light elements (H, O) lost very rapidly change in valency shell means free radicals formed . . .& consequent chemical reactions causing further damage beam damage is minimised by use of low temperatures (-160°) high beam voltages minimal exposure times
Similar presentations
© 2024 SlidePlayer.com. Inc.
All rights reserved.