Download presentation
Presentation is loading. Please wait.
Published byDominic Garrett Modified over 9 years ago
1
Need for Internal Dosimetry: Fundamentals to Clinical Practice – and Opportunity to Personalize Cancer Therapy Darrell R. Fisher Pacific Northwest National Laboratory Richland, Washington Joseph G. Rajendran University of Washington, Seattle, Washington Slides are not to be reproduced without permission of author
2
Educational objectives u Examine the need for and importance of radiation dosimetry in the nuclear medicine clinic –Benefits –Limitations –Approach and relevant quantities u Identify the important aspects of patient-specific dosimetry u Understand why patient-specific dosimetry is critical for achieving optimal therapy outcomes Slides are not to be reproduced without permission of author
3
“The objectives of dosimetry are to predict response and to guide clinical investigation toward safe and effective implementation of diagnostic and therapeutic radiopharmaceuticals. “These objectives can only be met if advances in computational and physics tools are coupled with advances in our understanding of the biokinetic behavior and radiobiology of radiopharmaceuticals. -- Dr. George Sgouros, Chair, MIRD Committee Slides are not to be reproduced without permission of author
4
Purpose for dosimetry Diagnostic nuclear medicine: 1.Provides the fundamental quantities used in radiation protection and risk assessment 2.Provides product safety information (package insert) Therapeutic nuclear medicine: 1.To propose and plan appropriate treatments 2.To correlate dose with biological effects in - Normal organs and tissues, and the whole body - Tumors 3.To predict radiation effects 4.To establish complete patient records Slides are not to be reproduced without permission of author
5
Rationale for therapy A. Treat to a desired radiation dose to the tumor –Thyroid cancer –Requires data on tumor mass, tumor uptake and retention OR B. Treat to a maximum tolerable radiation dose to the limiting normal organ or tissue –Radioimmunotherapy –Requires patient-specific data on patient organ mass, and the uptake, retention, and clearance of the radiopharmaceutical from the major normal organs Slides are not to be reproduced without permission of author
6
Counter-balanced goals u Establish a rigorous scientific rationale for selecting patient-individualized treatment levels u Minimize to the extent possible the complexity, costs, and time required for imaging and dosimetry costrigor Slides are not to be reproduced without permission of author
7
Two approaches 1. Dynamic Modeling Requires an appropriate pharmacokinetic or biokinetic model Requires known parameter values for the model compartments and transfer rates Example: Biokinetic models established by the International Commission on Radiological Protection (ICRP) Implemented using modeling software (SAAM II, STELLA) A1A1 A2A2 A0A0 A3A3 A4A4 VV V2V2 Slides are not to be reproduced without permission of author
8
Two approaches (continued) 2. Direct Measurements Planar imaging (and quantitative SPECT) - Patient positioning, anterior/posterior geometric means - Regions of interest for the major organs, tumor, whole body - Translation from counts to activity - Calibration against a radionuclide standard - Background subtraction, attenuation correction Marrow and tumor biopsy specimens Organ volumetrics by CT scans Activity-time curve-fitting Area-under-curve analysis Dosimetry calculations using the MIRD schema (implemented using software (such as OLINDA-EXM) Slides are not to be reproduced without permission of author
9
Imaging-based time-activity curve In-111 Day 4, posterior Time, hr activity Slides are not to be reproduced without permission of author
10
MIRD schema u Simplifies dose assessment without compromising on essential details u Is evolving to meet 21 st century needs u Extends from the whole-organ to the cellular and multicellular levels u May be applied to uniform or non-uniform radionuclide distributions u Patient-specific methods are preferred over use of generic model assumptions Slides are not to be reproduced without permission of author
12
Relevant quantity u The relevant dosimetric quantity (and unit) is the absorbed dose (Gy) -Relevant to organ and tumor deterministic effects u Not the “effective dose” (Sv) - In the ICRP terminology, effective dose is used in radiation protection and planning, * Defined only for stochastic effects (cancer induction) * Not for individual patient dosimetry Slides are not to be reproduced without permission of author
13
where k is a unit conversion constant A is the activity in the organ (Bq) E is the total energy emitted (J) f is the fraction of energy that is absorbed m is the mass of target tissue (g) B(t) is the biological retention with time t Absorbed dose in biological systems u General equation Slides are not to be reproduced without permission of author
14
Absorbed dose Rearranged MIRD formula D(r k r h ) = à h i i φ i (r k r h ) / m k where à h is the cumulated activity, i is the mean energy emitted per unit cumulated activity, and φ (r k r h ) is the absorbed fraction of energy imparted by a source organ Slides are not to be reproduced without permission of author
15
Absorbed dose Rearranged MIRD formula D(r k r h ) = Ã h i i φ i (r k r h ) / m k total number of transformations Slides are not to be reproduced without permission of author
16
Absorbed dose Rearranged MIRD formula D(r k r h ) = Ã h i i φ i (r k r h ) / m k energy imparted Slides are not to be reproduced without permission of author
17
Absorbed dose Rearranged MIRD formula D(r k r h ) = Ã h i i φ i (r k r h ) / m k organ mass Slides are not to be reproduced without permission of author
18
Absorbed dose D(r k r h ) = Ã h i i φ i (r k r h ) / m k patient- specific quantitative imaging energy transport and deposition patient organ CT imaging specific absorbed fractions Slides are not to be reproduced without permission of author
19
Absorbed dose D(r k r h ) = Ã h i i φ i (r k r h ) / m k patient- specific quantitative imaging energy transport and deposition patient organ mass by CT imaging specific absorbed fractions Three items to focus on for “personalized dosimetry” Slides are not to be reproduced without permission of author
20
Dosimetry is not a “precise science” u It is not possible to determine exact dose values for any tissue in the intact patient u Uncertainties: - Applying a mathematical model to represent the actual size and mass of the patient and internal organs: ± 20 to 60% -Measurement of activity in a source organ: ± 100% -Total error: ± 0 to 300% (ICRP-53) - Probable goodness of an organ dose estimate: ± 30% u The quality and accuracy of the direct measurements are the most important aspect of internal dose assessment Slides are not to be reproduced without permission of author
21
Model and reality Gastrointestinal tract Pelvic CT slice Slides are not to be reproduced without permission of author
22
Limitations u Requires well-trained people who understand the appropriate application and factors that can lead to errors - Common problem: confusing “biological” and “effective” u Mismatch between model and patient u Input and transposition errors u Chronic problem with software errors and validation Slides are not to be reproduced without permission of author
23
Relative improvement in dosimetry bias and precision Improvement value Difficulty and complexity Organ activity assessment by quantitative planar imaging highmoderate Phantom development and radiation transport calculations lowhigh CT-volumetrics to determine organ mass highlow Slides are not to be reproduced without permission of author
24
Phantom development and radiation transport MIRD “stylized” Realistic Tomographic, voxel, and hybrid models provide measured improvements in ability to calculate the self- organ and cross-organ gamma component Not yet scalable or patient-specific Revised S values not yet available Slides are not to be reproduced without permission of author
25
Newborn model courtesy Wes Bolch, University of Florida Slides are not to be reproduced without permission of author
26
Correcting for organ mass u A correction may be made for patient-specific organ masses that differ from the standard values (OLINDA-EXM) u This correction is appropriate when most of the organ dose is due to non-penetrating radiations u Alternative approach: multiply the source organ residence time by the ratio of the MIRD organ mass to the actual organ mass patient ( h ) (m MIRD /m actual ) Rajendran et al., 2004. J. Nucl. Med. 45(6):1059-1064. Slides are not to be reproduced without permission of author
27
Red marrow - Sensitive to ionizing radiation - Target tissue for certain radionuclide treatments (ablation therapy) - Often the dose-limiting normal tissue in radioimmunotherapy u In some protocols, marrow dosimetry is essential for treatment planning and evaluating biological response after treatment Slides are not to be reproduced without permission of author
28
Marrow dosimetry challenges u The geometry of red marrow structure with respect to bony spicules is difficult to describe and model u The amount of radionuclide in marrow is difficult to measure and quantify u Marrow is also irradiated by activity in circulating blood u High-dose irradiation of marrow may alter the cellularity of marrow and therefore marrow composition and volume Slides are not to be reproduced without permission of author
29
Approaches to marrow dosimetry u Make no direct measurements of marrow and assume that red marrow is part of the “remainder” tissues u Measure blood serum radioactivity, obtain the hematocrit, and assume that marrow uptake is proportional to activity in circulating blood (AAPM Task Group) u Measure the blood serum radioactivity (Sgouros method) u Direct imaging of a marrow space - repetitive counts over the same region (acetabulum, sacrum, pelvis, femur) - biopsy specimen (weighed and counted) Slides are not to be reproduced without permission of author
30
Sacral marrow scintigraphy u Obtain sacral marrow region-of-interest u Determine the time-activity curve u Assume that the sacral marrow = 9.9% of total-body marrow Advantage: may be more accurate than serum activity counting Disadvantage: activity in the sacrum is not always distinguishable from background (Siegel et al., 1989) Slides are not to be reproduced without permission of author
31
u Serial gamma camera (a/p) imaging to determine the marrow time-activity curves u Bone marrow biopsies obtained at 24-h post-infusion to determine %ID/g, to normalize the marrow time-activity curve Anterior view Posterior view Iliac Crest Sacrum Direct measurement with calibration Slides are not to be reproduced without permission of author
32
Normalizing a time-activity curve Plot the iliac or sacral direct measurements to a time-activity curve Correct the needle biopsy measurement for bone, blood, and fatty marrow Normalize the curve to the biopsy measurement normalize Slides are not to be reproduced without permission of author
33
Dose and biological response u Radiation dose correlates with biological response u The correlation always improves with ability to assess with accuracy both dose and end point Wessels et al, JNM 49:1884; 2008 Kidney dose vs. renal damage Slides are not to be reproduced without permission of author
34
Challenges u Uncertainties in quantitative (direct) imaging u Variability in red marrow volumes for different regions of the skeleton in a patient u Accuracy of biopsy specimen measurements –Variable fluid, cellular, fat, and bone contents –Small sample size u Marrow activity in the rib cage may interfere with lung activity assessment Slides are not to be reproduced without permission of author
35
Summary u Personalized dosimetry is important for selecting appropriate therapy amounts –based on absorbed dose to the normal limiting organ u Most important efforts in dose improvement will be: –Quantitative imaging for major normal organs –CT organ volumetrics –Red marrow imaging calibrated by biopsy (activity/gram) Slides are not to be reproduced without permission of author