Cryogenic particle detection at the Canfranc Underground Laboratory First International Workshop for the Design of the ANDES Underground Laboratory Centro Atómico Constituyentes Buenos Aires, Argentina 11-14 April, 2011
2
OutlineOutline The ROSEBUD Collaboration The scintillating bolometer Particle discrimination capability Experimental set-up in the old LSC facilities Main results: Light & heat response to different particles WIMP search prototypes Gamma and neutron spectroscopy ROSEBUD in the new LSC facilities EURECA
3
OutlineOutline The ROSEBUD Collaboration The scintillating bolometer Particle discrimination capability Experimental set-up in the old LSC facilities Main results: Light & heat response to different particles WIMP search prototypes Gamma and neutron spectroscopy ROSEBUD in the new LSC facilities EURECA
4
Institut d’Astrophysique Spatiale IAS (Orsay, France) Universidad de Zaragoza UNIZAR (Spain) Canfranc Underground Laboratory (LSC) ROSEBUD Collaboration (Rare Objects SEarch with Bolometers UndergrounD) N. Coron, C. Cuesta, E. García, C. Ginestra, J. Gironnet, P. de Marcillac, M. Martínez, A. Ortiz de Solórzano, Y. Ortigoza, C. Pobes, J. Puimedón, T. Redon, T. Rolón, M.L. Sarsa, L. Torres and J.A. Villar. Spectrométrie Thermique pour l’Astrophysique et la Physique (STAP) Institut d’Astrophysique Spatiale – IAS (Orsay, France) Nuclear and Astroparticle Physics Group (GIFNA) University of Zaragoza (Spain)
5
* light + heat technique has shown to be also a powerful tool for nuclear physics Testing of particle detector prototypes in a low background environment. R&D line: characterization of scintillating materials at low temperature. All materials tested have shown scintillation at low temperature: CaWO 4, BGO, LiF, Al 2 O 3 and SrF 2. Multi-target approach: use of scintillating bolometers of different materials in the same experimental set-up. Nuclear recoils discrimination against / background through light + heat technique for WIMP search. * Goals of ROSEBUD ROSEBUD scientific website: http://www.unizar.es/lfnae/rosebud/ Goals of ROSEBUD ROSEBUD scientific website: http://www.unizar.es/lfnae/rosebud/ Sapphire bolometers: 25, 50, 200 and 1000 g Ø 25 mm Ø 40 mm Ge optical bolometer
6
OutlineOutline The ROSEBUD Collaboration The scintillating bolometer Particle discrimination capability Experimental set-up in the old LSC facilities Main results: Light & heat response to different particles WIMP search prototypes Gamma and neutron spectroscopy ROSEBUD in the new LSC facilities EURECA
7
7 Thermal model of a simple bolometer Properties of bolometers 1.Wide choice of different absorber materials. 2.High energy resolution FWHM. 3.Low energy threshold for particle detection. 4.Particle identification capability in hybrid measurements of heat-light or heat-ionization energies. Dielectric and diamagnetic crystal The Scintillating Bolometer n
8
8 20 mK Internal reflecting cavity (Cu coated with Ag) Ge-NTD thermistor Optical bolometer (Ge disk) Scintillating crystal (absorber) Cu frame Thermal link BGO 92 g Thermal link Ge-NTD thermistor The Scintillating Bolometer
9
OutlineOutline The ROSEBUD Collaboration The scintillating bolometer Particle discrimination capability Experimental set-up in the old LSC facilities Main results: Light & heat response to different particles WIMP search prototypes Gamma and neutron spectroscopy ROSEBUD in the new LSC facilities EURECA
10
10 Particle Discrimination Capability Nuclear recoils spectrum c / channel 59.5 keV ( 241 Am) / spectrum Light pulse amplitude (mV) Heat pulse amplitude (mV) Sapphire Bolometer (50 g) Calibration 252 Cf & 241 Am internal source Nuclear recoils region Detector response alphas //
11
OutlineOutline The ROSEBUD Collaboration The scintillating bolometer Particle discrimination capability Experimental set-up in the old LSC facilities Main results: Light & heat response to different particles WIMP search prototypes Gamma and neutron spectroscopy ROSEBUD in the new LSC facilities EURECA
12
Canfranc Underground Laboratory (LSC) 2450 m.w.e. Tobazo Peak Spanish Pyreness ROSEBUD 12 Muon flux decreased by a factor ~10 5 10 m 2 1995 - today 118 m 2 Experimental set-up in the old LSC facilites The underground laboratory 2.5·10 -3 m -2 s -1
13
13 LiF Mass 33 g Monitoring of neutrons through 6 Li(n,t) Sapphire (Al 2 O 3 ) Mass 50 g 27 Al: ↓A sensitive to SI and SD interactions High β/γ background rejection ↓Z & low energy threshold BGO (Bi 4 Ge 3 O 12 ) Mass 46 g 209 Bi: ↑A, sensitive to SI and SD interactions 207 Bi contamination (clean BGOs available) / spectrometer ↑Z T = 20 mK Experimental set-up in the old LSC facilites The scintillating bolometers
14
14 Experimental set-up in the old LSC facilites The dilution refrigerator and shielding
15
Faraday cage (2 2 3 m 3 ) Pumping and control systems 15 Experimental set-up in the old LSC facilites The Faraday Cage and the cryogenic pumping system
16
OutlineOutline The ROSEBUD Collaboration The scintillating bolometer Particle discrimination capability Experimental set-up in the old LSC facilities Main results: Light & heat response to different particles WIMP search prototypes Gamma and neutron spectroscopy ROSEBUD in the new LSC facilities EURECA
17
17 Sapphire Light REF ( / : : NR) 59.5 keV 122 keV 136.5 keV 241 Am + 57 Co Heat signal amplitude (mV) Light signal amplitude (mV) Light output (γ) = 13.5 keV / MeV REF (γ / NR) = 17.5 ± 1.5 (@ 200 keV ) Light output (α) = 1.3 keV / MeV REF ( / α) = 10.3 ± 1.0 (@ 5.3 MeV ) / events events 210 Po
18
18 BGO Light REF ( / : : NR) 252 Cf REF( / : ) 209 Bi + 232 Th daughters
19
19 Sapphire Thermal REF ( 206 Pb nuclear recoils: / ) 206 Pb recoils at 103.08 ± 0.10 keV from 210 Po source 206 Pb recoil Spectrum of the events in the NR band Thermal REF of NR Relevant for the calibration of the dark matter signal Al 2 O 3
20
BGO irradiation with 241 Am source BGO Thermal REF ( 237 Np nuclear recoils: / ) 237 Np recoiling nuclei at 92.40 ± 0.12 keV from 241 Am source
21
21 Energy partition in Sapphire and BGO scintillating bolometers Heat pulse amplitude (mV) Light pulse amplitude (mV) 57 Co + 241 Am 59.5 keV ( 241 Am) 136.5 keV ( 57 Co) 122.1 keV ( 57 Co) Al 2 O 3 + h + 0 = 1 0 = 0.478 0.093 = 0.058 0.006 h = 0.464 0.093 BGO 0 = 0.110 0.104 = 0.112 0.013 h = 0.778 0.103 Al 2 O 3
22
OutlineOutline The ROSEBUD Collaboration The scintillating bolometer Particle discrimination capability Experimental set-up in the old LSC facilities Main results: Light & heat response to different particles WIMP search prototypes Gamma and neutron spectroscopy ROSEBUD in the new LSC facilities EURECA
23
23 C.L (1 tailed) E (keV) 90 %8.7 95 %10.0 99 %13.0 99.9 %17.1 99.99 %21.3 Sapphire Kyropoulos grown C.L (1 tailed)E (keV) 90%23.5 99.9%33.3 BGO Czochralski grown BGO Discrimination of NR down to ≈25 keV Discrimination of NR down to ≈10 keV WIMP searches Particle discrimination power
24
OutlineOutline The ROSEBUD Collaboration The scintillating bolometer Particle discrimination capability Experimental set-up in the old LSC facilities Main results: Light & heat response to different particles WIMP search prototypes Gamma and neutron spectroscopy ROSEBUD in the new LSC facilities EURECA
25
25 222 Rn inside the Pb shielding The BGO allows to note the background level increase and also to identify its origin The BGO scintillating bolometer as -ray spectrometer
26
Heat channel energy resolution
27
1keV25meV capture elastic scattering total Thermal n 27 Resonance capture The LiF scintillating bolometer as neutron spectrometer 6 Li(n,t) neutron detection efficiencies of LiF bolometers
28
NR Irradiation of a 33 g LiF and a 50 g Al 2 O 3 scintillating bolometers with 252 Cf 6 Li(n, ) resonance Previous work presented at TAUP09: J Phys: Conf Series 203 (2010)012139 Hypothesis: fast neutron flux inside the lead shielding We estimated the region of three parameters ( 0, ,T) compatible with experimental data LiF & Al 2 O 3 6 Li(n, ) Q th = 4.78 MeV / events NR / events events Scintillating bolometers as neutron spectrometers Fast neutron flux inside the shielding Scintillating bolometers as neutron spectrometers Fast neutron flux inside the shielding
29
Present work: Testing hypothesis about the fast neutron flux inside the shielding Al 2 O 3 heat NR spectra measured MCNP-PoliMi Spectra shape in good agreement Comparison of full experimental data with MC calculation is in progress = −0.9 T = 1.48 MeV Scintillating bolometers as neutron spectrometers Fast neutron flux inside the shielding Scintillating bolometers as neutron spectrometers Fast neutron flux inside the shielding
30
OutlineOutline The ROSEBUD Collaboration The scintillating bolometer Particle discrimination capability Experimental set-up in the old LSC facilities Main results: Light & heat response to different particles WIMP search prototypes Gamma and neutron spectroscopy ROSEBUD in the new LSC facilities EURECA
31
31 ROSEBUD in the new LSC facilities http://www.lsc-canfranc.es/
32
ROSEBUD in the new new LSC facilities Hall B ROSEBUD in the new new LSC facilities Hall B 3 3 4.5 m³
33
ROSEBUD in the new new LSC facilities Hall B ROSEBUD in the new new LSC facilities Hall B
34
OutlineOutline The ROSEBUD Collaboration The scintillating bolometer Particle discrimination capability Experimental set-up in the old LSC facilities Main results: Light & heat response to different particles WIMP search prototypes Gamma and neutron spectroscopy ROSEBUD in the new LSC facilities EURECA
35
EURECA project Universidad de Zaragoza Institut d’Astrophysique Spatiale IAS Target mass 1 ton. Semiconductor bolometers built with the technology of EDELWEISS (LSM). Scintillating bolometers built with the technology of CRESST (LNGS) and ROSEBUD (LSC). To be carried out at the Modane Underground Laboratory (LSM) in France. 35 UNIZAR and IAS integrated in EURECA http://www.eureca.ox.ac.uk/ See also Gilles Gerbier’s talk!
36
ROSEBUD is a collaborative effort dedicated to the development of scintillating bolometers for nuclear and particle physics experiments, focusing on rare event search experiments. Scintillating bolometers characterized by ROSEBUD (Al 2 O 3, BGO and LiF) have shown excellent capabilities for particle discrimination and background rejection. ROSEBUD is currently moving to the Hall B in the new LSC facilities planning to restart measurements in 2012. New materials are being characterized in IAS. UNIZAR and IAS also participate in EURECA. Conclusions
37
Radiopurity measurements of materials (crystals, cryogenic resins, shieldings, detector components, dilution unit pieces) at LSC using ultra-low background germanium detectors. Development of new scintillating bolometers. Test of scintillating bolometers at the Canfranc Underground Laboratory (LSC) in order to characterize and optimize scintillators in a low background environment, evaluating these for use as potential dark matter targets in EURECA. UNIZAR and IAS participation in EURECA