Download presentation
Presentation is loading. Please wait.
Published byBrandon Golden Modified over 9 years ago
1
Infrared Spectroscopy Department of Natural Sciences UM-Dearborn Written by Bette Kreuz Produced by Ruth Dusenbery 2000
2
Info Slide 1 n Infrared spectroscopy is an instrumental method of analysis that can be used to identify and quantify samples ranging from pharmaceuticals to diesel emissions.
3
Slide 1 Introduction
4
Info Slide 2 n All modern instruments are Fourier Transform instruments. n In all transmission experiments radiation from a source is directed through the sample to a detector.
5
Slide 2 SampleCompartment IR Source Detector
6
Info Slide 3 n The measurement of the type and amount of light transmitted by the sample gives information about the structure of the molecules comprising the sample.
7
Slide 3 InfraredSource Detector SampleCompartment
8
Info Slide 4 n In the IR region of the electromagnetic spectrum, the absorption of radiation by a sample is due to changes in the vibrational energy states of a molecule.
9
Slide 4
10
Info Slide 5 n The pattern of absorption as a function of wavelength is called an IR spectrum.
11
Slide 5
12
Info Slide 6 n To obtain an IR spectrum, the sample must be placed in a “container” or cell that is transparent in the IR region of the spectrum. n Sodium chloride or salt plates are a common means of placing the sample in the light beam of the instrument.
13
Slide 6 IR transparent Salt Plates
14
Info Slide 7 n These plates are made of salt and must be stored in a water free environment such as the dessicator shown here.
15
Slide 7 DessicatorWater-freeEnvironmentforWater-sensitive Salt Plates.
16
Info Slide 8 n The plates must also be handled with gloves to avoid contact of the plate with moisture from one’s hands.
17
Slide 8
18
Info Slide 9 n To run an IR spectrum of a liquid sample, a drop or two of the liquid sample is applied to a salt plate.
19
Slide 9
20
Info Slide 10 n A second salt plate is placed on top of the first one such that the liquid forms a thin film “sandwiched” between the two plates.
21
Slide 10
22
Info Slide 11 n The two plates are then secured in a sample holder that is compatible with the particular instrument being used.
23
Slide 11
24
n The cell holder is then placed in the beam of the instrument. Info Slide 12
25
Slide 12
26
n The light beam traverses the sample compartment, as illustrated by the red line. Info Slide 13
27
Slide 13 Light Path (shown by red line)
28
Info Side 14 n The sample is then scanned by the instrument utilizing predesignated parameters. n A relevant background scan should already have been taken.
29
Slide 14 Click Here to Start Scan
30
Info Slide 15 n A satisfactory spectrum has well defined peaks-but not so intense as to cause flattening on the bottom of the peaks. n Major peaks can be labeled using the peak function of the software using the peak function of the software
31
Slide 15 Well-defined peaks are labeled with the Wavenumbers of the Absorption Maxima.
32
Info Slide 16 n The spectrum can then be printed using the print function of the software.
33
Slide 16
34
Info Slide 17 n A sample of a printout of an IR spectrum.
35
Slide 17 Sample of a printout of an IR spectrum.
36
Info Slide 18 n The salt plates are cleaned by rinsing into a waste container with a suitable organic solvent-commonly cyclohexane. n NEVER WATER!
37
Slide 18 CYCLOHEXANESolvent
38
Info Slide 19 n Cloudy plates must be polished to return them to a transparent condition. n To polish cloudy windows, rotate salt plate on polishing cloth.
39
Slide 19
40
Info Slide 20 n The clean plates and cell holder are stored in the moisture free atmosphere of a dessicator.
41
Slide 20
Similar presentations
© 2024 SlidePlayer.com. Inc.
All rights reserved.