Presentation is loading. Please wait.

Presentation is loading. Please wait.

LN2 filling system 1.Filling the buffer tank 2.Filling the collector tube 3.Filling the detectors 4.Venting the collector tube 5.Monitoring then go to.

Similar presentations


Presentation on theme: "LN2 filling system 1.Filling the buffer tank 2.Filling the collector tube 3.Filling the detectors 4.Venting the collector tube 5.Monitoring then go to."— Presentation transcript:

1 LN2 filling system 1.Filling the buffer tank 2.Filling the collector tube 3.Filling the detectors 4.Venting the collector tube 5.Monitoring then go to 1 or 2

2 1. Filling the buffer tank Automatic? Alarm on the buffer tank to alert operators?

3 2. Filling the collector tube Logic sequence: – Open VP – Open VIN – While FP wait for TP to get to LN2 value – Close VP

4 3. Filling the detectors Logic sequence: – (VIN already open, VP closed) – Open VDi – Wait until TVi is at LN2 temp – Close VDi

5 4. Venting the collector tube Logic sequence: – Close VIN – Open VP

6 5. Monitoring Logic sequence: – Check P,V against alarm threshold ->Step 1: Fill the buffer – Check Tdi against limit value ->Step 2: Fill all the detectors – Check time since last fill ->Step 2: Fill all the detectors

7 Implementation – collector VIN 24V NI 9476 VP 24V NI 9476 NI 9217NI 9425 10 – 40 V DC NI 9217 Collector Safety Valve(s) NI 94761 + 1/column NI 94251 /column NI 92171 +1 24 V AC1 +1/column 10-40 VDC1/column

8 Implementation - detector VDi 24V NI 9476 NI 9217 Collector Safety Valve(s) NI 94761/detector NI 92171 24 V AC1 /detector Dewar TDi

9 Implementation – monitoring V Collector P NI 9217 NI 9425? NI 9217 NI 9425? Dewar Crystal NI 9217

10 Custom modules To NI 9217 PTi From NI 9476 12 V Power On/Off Open / Close Opened / Closed Powered Y/N To NI 9425 Read PT100 Distribute valve signal Command one valve 10-40 V DC Read flux sensors M1 M2 M3 M4 12V

11 M3 working principle 12 V Power On/Off Open / Close Opened / Closed Powered Y/N M3 12V NI 9476 Valve 12 V

12 Connection diagram Compact Rio Crate (6 x NI 9217, 1 x NI 9476, 1 x NI 9425) PT100 Prox M1 12V Prox M2 M3 Valve 12V? M4