1 Possible Radiographic Errors Radiology
2
2 During Exposure Film is bent during set up and exposure Causing distortion Film is bent during set up and exposure Causing distortion
3
3 Elongation Vertical angulation error; not enough positive angle on max or negative angle on md
4
4
5
5 Foreshortening Also vertical angulation; too much angulation; correct by decreasing vertical angulation
6
6
7
7 Overlapping Incorrect horizontal angulation; cone is angled toward mesial or distal surfaces of teeth, instead of at interproximal areas *Evaluate film placement in regard to curve of arch, NOT pt;s face/cheeks Incorrect horizontal angulation; cone is angled toward mesial or distal surfaces of teeth, instead of at interproximal areas *Evaluate film placement in regard to curve of arch, NOT pt;s face/cheeks
8
8
9
9 Conecutting X-ray beam missed part of the film; film is only partially exposed
10
10
11
11 Clear Film Film was not exposed to any radiation
12
12 Double Exposure Film is exposed twice; results in blurred images or dark x-ray
13
13
14
14 Blurred Image Result from movement of pt’s head or tubehead or from film moving in pt’s mouth
15
15 Underexposed Film Light and has thin image; check the settings for mAs, kVs and exposure time; also occurs if cone is not positioned close enough to pt’s face
16
16
17
17 Overexposed Film Dark image; check settings again
18
18
19
19 Radiopaque Film Metal objects; check for and remove
20
20 Backward Film Placing film reversed causes image to be white and a herringbone pattern to be seen
21
21
22
22 Film Processing
23
23 Light film image Underprocessed; developing time too short, developing temp too low, developing solution weak from overuse and needs to be changed
24
24 Dark film image Over developing: developing solution temp too high, devel solution too stong, film left in devel too long
25
25 Fogged Film Gray appearance, film image detail is lost, contrast is lessened; improper storage conditions, outdated films, light leaks in processing room, light leaks on daylight loaders
26
26
27
27 Partial Image Solution levels are low
28
28
29
29 Film Artifacts Result from improper handling; water spots on unprocessed films leaves clear area; fixer on unprocessed film leaves white area; developer leaves black area; also, yellow/brown stain from not rinsing properly, black artifacts from static electicity, fluroide leaves fingerprint smudge; films which touch leave shape of edge or corner of film
30
30 Torn or Scratched Rough handling; leaves white area or mark on processed film
31
31
32
32 Air Bubbles on Film Trapped on film if not agitated when placed in processing solutions; leaves round white spots
33
33 Reticulation When film has been exposed to high temp then low temp; looks dried with tiny cracks
34
34 Streaks Dirty rollers or x-ray racks