15/07/2010Rui De Oliveira1 Update on large size GEM manufacturing Rui de Oliveira
2
Process Characterization Some pictures Large volume production outside CERN 15/07/2010Rui De Oliveira2
3
15/07/2010Rui De Oliveira3 Chemical Polyimide etching Second Polyimide etching Copper electro etching Stripping reality
4
15/07/2010Rui De Oliveira4 connected Non etched Top Electrode Lower voltage than the Chemical reaction +3V Resist layer +3V0V Bath at +3V -everything at ground will be protected -everything at +3V will be chemically etched - An example of active corrosion protection
5
Starting structure
6
Conventional isotropic metal etching
7
The hole is created but still an angle in the copper
8
With over etch
9
Going back to Polyimide etching for 30 sec The hole become bi-conical
10
10
11
11 TOP BOTTOM
12
12 TOP BOTTOM
13
13 TOP BOTTOM
14
14 TOP BOTTOM Sample 1 Sample 2
15
15 TOP BOTTOM Sample 3 Sample 4
16
16 Drift GEM Anode 3.05 mm 2.15 mm GEM active area: 10 x 10 cm 2 Gas mixture: Ar/CO 2 70/30 Gas flow: ~ 5 l/h Water content: ~ 100 ppm H 2 O Radiation source: Cu X–ray tube
17
17 Run_0110 Both GEMs in “open–bottom” config E d = 2 kV/cm E i = 3 kV/cm DV GEM = 480 V Energy resolution = 18.4 % FWHM/peak @ 8.04 keV Run_0104 Both GEMs in “open–top” config E d = 2 kV/cm E i = 3 kV/cm DV GEM = 480 V Energy resolution = 19.5 % FWHM/peak @ 8.04 keV
18
18 E d = 2 kV/cm E i = 3 kV/cm Rate bottom ~ 21 kHz/mm 2 Rate top ~ 18 kHz/mm 2 Spark V bottom ~ 540 V Spark V top ~ 530 V Max G bottom ~ 1390 Max G top ~ 1080
19
19 Drift GEM3T Anode 3.25 mm GEMs active area: 10 x 10 cm 2 Gas mixture: Ar/CO 2 70/30 Gas flow: ~ 5 l/h Radiation: Ag X–ray tube, Cu filter 2D (X–Y) readout, all strips shorted GEMs in open–bottom configuration HV provided by voltage divider 2 HV filters on the line (600 k total) GEM3B GEM2T GEM2B GEM1T GEM1B 2.215 mm 2.225 mm 2.15 mm
20
20 E d = 2.09 kV/cm DV GEM1 = 391 V E t1 = 3.09 kV/cm DV GEM2 = 354 V E t2 = 3.08 kV/cm DV GEM3 = 309 V E i = 3.19 kV/cm Rate ~ 2.9 kHz Energy resolution=20.6% FWHM/peak @ 8.04 keV
21
21 E d ~ 2 kV/cm E t1 ~ 3 kV/cm E t2 ~ 3 kV/cm E i ~ 3 kV/cm Rate ~ 420 kHz
22
22 Cu X–rays 2 mm ø collimator Ni #1 absorber DV div = 4200 V gain ~ 2560 rate = 300 Hz T = (2000 ± 900) s DG = (1.7 ± 0.2) %
24
15/07/2010Rui De Oliveira24
25
15/07/2010Rui De Oliveira25
26
15/07/2010Rui De Oliveira26