Measuring end-expiratory lung volume Giacomo Bellani, M.D., Ph.D. University of Milano-Bicocca Monza, Italy giacomo.bellani1@unimib.it
2
Conflicts of interest Personal: Lecturing fees from GE Institutional, research grants from: Draeger Maquet Chiesi Farmaceutici
3
Which «lung volume» ? Functional Residual Capacity (FRC) Volume of gas in the lung at the end of a physiologic expiration Relaxation volume of respiratory system End Expiratory Lung Volume (EELV) Volume of gas in the lung at end expiration during mechanical ventilation with PEEP EELV @ 0 cmH 2 O= FRC
4
How do we measure End- expiratory lung volume? Why should we measure end-expiratory lung volume?
5
Techniques CT scan Precise quantitative assessment Radiological exposure+patient transfer Closed circuit Need for a dedicated tracer+closed system No need for «fast» response concentration measurements - offline measurement Open circuit muti-breath washin/washout Potentially no need for a gas tracer Fast response measurement, synchrony with tidal ventilation
6
Techniques CT scan Precise quantitative assessment Radiological exposure+patient transfer Closed circuit Need for a dedicated tracer+closed system No need for «fast» response concentration measurements - offline measurement Open circuit muti-breath washin/washout Potentially no need for a gas tracer Fast response measurement, synchrony with tidal ventilation
7
Closed Dilution Technique V f = V i + EELV Mass conservation CiVi FRC ? CfVf Vi Cf Ci ViEELV -= V i * C i = V f * C f
9
Techniques CT scan Precise quantitative assessment Radiological exposure+patient transfer Closed circuit Need for a dedicated tracer+closed system No need for «fast» response concentration measurements - offline measurement Open circuit muti-breath washin/washout Potentially no need for a gas tracer Fast response measurement, synchrony with tidal ventilation
10
70 % 30 % O2O2 N2N2 50 % FiN 2 =70% VtVt FeN 2 600 ml 1400 ml 1000 ml
11
Multibreath nitrogen washout Wrigge H et al Intensive Care Med 1998; 24: 487 Off-line correction of viscosity effect on sidestream delay time and pneumothacografic measurments
12
Derives N 2 concentration from: 100%- [CO 2 ]- [CO 2 ] Uses end-tidal concentration (no need for synchronization)
14
What about Pressure Support ? 0 1000 2000 3000 4000 01000200030004000 EELV by He dilution (ml) EELV by GE Engstrom (ml) Volume Control R 2 = 0.8353 Pressure Support Using a 20% variation in FiO 2 : Bellani G, unpublished
15
FRC INview (Engstrom carestation)
18
How do we measure End- expiratory lung volume? Why should we measure end-expiratory lung volume?
19
EELV is profoundly reduced in ARDS Estimate of alveolar recruitment Determinant of VILI Why should we measure end- expiratory lung volume?
20
EELV is profoundly reduced in ARDS Estimate of alveolar recruitment Determinant of VILI Why should we measure end- expiratory lung volume?
21
Alveolar recruitment by P-V curve Maggiore S. et al, Am J Resp Crit Care Med, 2001
22
Effect of recruitment on FRC (0 cmH 2 O)
23
Patroniti N et al., CCM 2010
24
A simplified approach Assessing alveolar recruitment by EELV PEEP = 0 Compliance=30 ml/cmH2O EELV= 1000 ml PEEP=10 Expected EELV increase: 30 * 10= 300 ml No recruitment EELV= 1300 ml Recruitment EELV = 1600 ml
25
0 0.2 0.4 0.6 0.8 1 1.2 1.4 1.6 1.8 2 4 14 PEEP (cmH2O) EELV (l) 0 5 10 15 20 25 30 35 Crs (ml/cmH2O) 681012 1350 ml Crs 29 expected EELV increase 174 ml 1524 l 1950 l true EELV increase 600 ml ∆EELV / PEEP change = 600 / 6 = 100 ml/cmH2O
28
EELV is profoundly reduced in ARDS Estimate of alveolar recruitment Determinant of VILI Why should we measure end- expiratory lung volume?
29
EELV V E (L/min) RATIO NORMAL ARDS 2500< 7< 2.8 1000> 15 SPECIFIC HYPERVENTILATION
30
Vt/FRC
31
End-expirationEnd-inspiration PET EELV normally-aerated Vt normally-aerated Ki normally-aerated High activity Low activity - 1000 100 -500 Regional distension and inflammation Bellani G et al., Am J Resp Crit Care Med, 2011
32
The role of tidal volume/EELV Bellani G et al., Am J Resp Crit Care Med, 2011
34
Conclusions: o Measurement of end-expiratory lung volume is now available: RESEARCH TOOL → CLINICAL TOOL o Useful in estimating lung recruitment o Can it help in setting PEEP? Yes o Is it being used to set PEEP? Don’t know o Possible role in adjusting tidal volume?
Similar presentations
© 2025 SlidePlayer.com. Inc.