Download presentation
Presentation is loading. Please wait.
Published byClifton Dalton Modified over 8 years ago
1
Biophotonics lecture 11. January 2012
2
Today: -Correct sampling in microscopy -Deconvolution techniques
3
Correct Sampling
4
What is SAMPLING? Intensity [a.u.] 2 3456 X [µm] 1
5
Aliasing … suppose it is a sine-wave Intensity [a.u.] 2 3456 There are many sine-waves, SAMPLED with the same measurements. Which is the correct one?
6
Intensity [a.u.] 2 3456 X [µm] When sampling at the frequency of the signal, a zero-frequency is recorded!
7
Intensity [a.u.] 2 3456 X [µm]
8
Intensity [a.u.] 2 3456 X [µm] Problem: too high frequencies will be aliased, they will seemingly become lower frequencies
9
But … high frequencies are not transmitted well. Object: Microscope Image: Intensity Spatial Coordinate Intensity Spatial Coordinate OTF
10
Aliasing in Fourier-space Fourier-transform of Image Intensity Aliased Frequencies ½ Sampling Frequency Cut-off frequency =½ Nyquist Rate Sampling Frequency Nyquist Rate
11
Pixel sensitivity Intensity [a.u.] 2 3456 X [µm] 1 Convolution of pixel form factor with sample Multiplication in Fourier-space Reduced sensitivity at high spatial frequency
12
Optical Transfer Function |k x,y | [1/m] contrast Cut-off limit 0 1 rectangle form-factor OTF sampled
13
Consequences of high sampling Confocal: high Zoom more bleaching? No! if laser is dimmed or scan-speed adjusted bad signal to noise ratio? Yes, but photon positions are only measured more accurately binning still possible high SNR. Readout noise is a problem at high spatial sampling (CCD)
14
Optimal Sampling?
15
Regular sampling Reciprocal -Sampling Grid Real-space sampling: Multiplied in real space with band-limited information
16
Regular sampling Reciprocal -Sampling Grid Real-space sampling:
17
Widefield Sampling In-Plane sampling distance Axial sampling distance
18
Confocal Sampling In-Plane sampling distance (very small pinhole) else use widefield equation Axial sampling distance
19
Confocal OTFs WF 1 AU 0.3 AU in-plane, in-focus OTF 1.4 NA Objective WF Limit
20
Hexagonal sampling Advantage: ~17% + less ‚almost empty‘ information collected + less readout-noise approximation in confocal Reciprocal d-Sampling Grid Real-space sampling: Multiplied in real space with band-limited information
21
63× 1.4 NA Oil Objective (n=1.516), excitation at 488 nm, emission at 520 nm l eff = 251.75 nm, a = 67.44 deg widefield in-plane: d xy < 92.8 nm maximal CCD pixelsize: 63×92.8 = 5.85 µm confocal in-plane:d xy < 54.9 nm widefield axial: d z < 278.2 nm confocal axial: d z < 134.6 nm Fluorescence Sampling Example
22
OTF is not zero but very small (e.g. confocal in-plane frequency) Object possesses no higher frequencies You are only interested in certain frequencies (e.g. in counting cells, serious under-sampling is acceptable) Reasons for undersampling
23
If you need high resolution or need to detect small samples sample your image correctly along all dimensions Sampling Summary
24
Maximum Likelihood Deconvolution
27
Image: http://en.wikipedia.org
30
The prior (requires prior knowledge; can imply contraints, e.g. positivity) Constant normalisation factor
31
Constant, therefore obsolete
37
MATLAB demonstration
38
Information & Photon noise Virtual Microscopy Only Noise? FT NO! 10 Photons / Pixel
39
Band Extrapolation? Object Mean Error Energy Mean Energy Relative Energy Regain
40
With Photon Noise
41
Is this always possible? White Noise Object
42
Is this always possible? Unfortunately NOT !
Similar presentations
© 2025 SlidePlayer.com. Inc.
All rights reserved.