Venkateswarlu Gaddam, Gautam Sharma, Neelesh Kumar, Amod Kumar, S.K. Mahna NIT Kurukshetra Central Scientific Instruments Organisation (CSIO) Council for Scientific and Industrial Research(CSIR) Chandigarh,India
2
Introduction Objective Materials and Methods Observations Results Conclusion References
3
GRF- Ground Reaction Force It is equal in magnitude and opposite in direction to the force that the body exerts on the supporting surface through the foot. Vertical GRF (Fz) & its importance- It is considered to be prominent amongst the three components of the GRF and has importance in the gait analysis, clinical assessment etc. Force Plate (FP)- To measure GRF in the human and animal locomotion. Available FPs- AMTI FP, Bertec FP, Kistler FP Kistler FP- Standard Reference Tool Developed FP Calibration Techniques Static calibration Dynamic Calibration
4
To measure static & dynamic calibration of Developed FP Comparison of their (kistler FP & Developed FP) results
5
GRF MEASUREMENT WITH Kistler FP Vertical component of ground reaction force Subject walking on force plate Kistler Data Acquisition UnitPC with Analysis Software
6
Type- 9286B (Switzerland) Sensing Devices- Piezoelectric sensors Cost-17 lacks Plates- aluminum
7
DAQ system for Bioware software Type-5691A
8
To measure & analyze the acquired signal Force calculation Moment calculation Center of pressure (COP) calculation
9
Developed FP ( Design & Sensor arrangementDeveloped FP ( Design & Sensor arrangement) Experimental Setup Ni-DAQSoftware Tool Calibration Techniques Results
10
Plates (2)-Brass Sensors- Linear Pot Length- 60 cm Breadth- 36 cm Height- 7.4 cm Springs arrangement
13
NI USB-6259 32 inputs,16 bit,32Ms/s Multifunction I/O with correlated digital I/O for USB
14
LabVIEW –Laboratory Virtual Instrument Engineering Workbench LabVIEW Signal Express-8.6 Major Role Simulation Data Acquisition Instrument Control Testing & Measurements Industrial automation Data Analysis
15
Calibration- calibration is a comparison between measurements-one of known magnitude or set with one device and another measurement made in a similar way as possible with a second device. Purpose- to check if the developed FP gives the same measurement readings as the Kistler FP, thereby checking the repeatability, sensitivity, linearity of VGRF for the Developed FP. Static Calibration: Input- Fixed Loads Output- Volts
16
Dynamic Calibration Subject- Human being Input- Applied Force from human feet during walking Output- M-Shape curve in Volts
17
Static Calibration Dynamic Calibration Load Vs Voltage
18
(N) 539646735
19
Range of Weight ( Kg) Sensitivity of Kistler FP (mv/N) Sensitivity of Developed FP (mv/N) 55-6612.2241.11 66-756.66728.88
20
To evaluate Gait Parameters which are used for Prosthetic Knee quantification In clinical assessments to know data analysis of a patient Inexpensive
21
[1]. Senanayake S.M.N.A.(2004) Walking, Running and Kicking using Body Mounted Sensors. Proceedings of the 2004 IEEE. Conference on Robotics, Automation and Mechatronics Singapore. Pg 1141-1146. [2]. Hynd D.,Hughes S C.,Evins D J.( 2000) The development of a long, dual- platform triaxial walkway for the measurement of forces and temporal-spatial data in the clinical assessment of gait. Proc Instn Engrs vol 214 Part H: pg 193- 201. [3]. Liu T., Inoue Y., Shibata K.(2009) 3D Force Sensor Designed Using Pressure Sensitive Electric Conductive Rubber. Journal of system design and dynamics, VOL 3 No 3 pg 282-295. [4]. Roland E.S., Hulla M.L., Stover S.M., (2005) Design and demonstration of dynamometric horseshoe for measurinh ground reaction loads of horses during racing conditions. Journal of Biomechanics 38 : pg 2102-2112.
22
THANK YOU!