Elements of DR Imaging Systems
Capture element The element/material used to capture the x-ray photon Coupling element Transfers the x-ray generated signal to the collection element Fiber optics, lens, contact layer, or a-Se Collection element Photodiode, CCD, TFT, CMOS Photodiodes, CCDs, and CMOS are light sensitive devices that collect light photons TFT collect electrons
2
Digital Radiography Type CR Indirect DR Direct DR Capture Element
Barium fluorohalide CsI CsI – cesium iodide/GdOD gadolinium oxysulfide CsI – cesium iodide a-Se amorphous selenium Coupling element Lens or fiber optics a-Si Amorphous silicon Fiber optics or lens a-Se Collection element CMOS TFT CCD
3
Indirect w/ CCD and fiber optics
Multiple detectors Single detector CCD Fiber optic coupling CsI capture element
4
Indirect w/CMOS and lens
5
Indirect TFT a-Si Coupling element
6
Direct TFT
7
Review of Indirect DR Basically there are three types of indirect systems CCD A multitude of small CCD (charged coupled devices) receive light from the cesium iodide scintillator. TFT The light from scintillator is recorded by a TFT (thin film transistor) that is also called a flat panel detector. CMOS (complementary metal oxide semiconductor) A lens focuses light on the CMOS chip
8
Flat panel technology (TFT) indirect and direct DR applications