PPE Use Standard (SSP-0011) Y Full Face Shield / Goggles Hood Clothing—Chemical Protection Gloves—Chemical Personal Fall Protection Respirator / Filter.

Similar presentations


Presentation on theme: "PPE Use Standard (SSP-0011) Y Full Face Shield / Goggles Hood Clothing—Chemical Protection Gloves—Chemical Personal Fall Protection Respirator / Filter."— Presentation transcript:

1 PPE Use Standard (SSP-0011) Y Full Face Shield / Goggles Hood Clothing—Chemical Protection Gloves—Chemical Personal Fall Protection Respirator / Filter Mask Supplied Air Job Hazards Y es Job Hazards Y es Access and EgressMoving Parts Activities of OthersNoise Adverse WeatherRail Movement Chemical ExposureSharp Edges ElectricitySlips, Trips, and Falls FatigueStatic Electricity Fire and ExplosionTrip Injuries Ground ContaminationTrapped Pressure Manual HandlingWorking at Height Other: References / Documents  MSDS on gasoline.  Unit Safe Upper & Lower Limits Spreadsheet AreaUnitOperating Procedure #Certified byRevision #DatePage 3LabOPS-LB-PS-01809.20.20111 of 2 Sulfur by Horiba SLFA-2800 EnvironmentalSpecial Instructions Sulfur by Horiba SLFA-2800 1Turn on POWER switch located on the rear of the unit. 2After approximately 40 seconds, the READY menu will appear and the lamp at right back rear of unit will light up. 3 Ensure the following parameters are displayed: a.Press the COND key (F1) if need to change measurement time, repetition count, calibration curve used or unit (% or ppm). b.Use and keys to select desired condition, then press ENTER. c.Once all parameters are correct, press EXIT key to return to the READY menu. 4 Press ID# key to enter up to 6 digit ID number for sample(s). a.Using keypad, depress # (ex. 1, 2, 3) that corresponds to alphabetical character desired – each one press of the key will scroll to next alphabetical character. (Ex., to select the letter “C”, depress 2 ABC button and then depress 3 more times). b.To fix selected value, release the finger from the key for more than 1 second. c.Once sample ID is completed, press ENT key to make curser to next POS #. d.Once all sample IDs have been entered, press EXIT key to return to READY menu. 5 Prepare sample(s) as follows: a.Place sample cell on clean counter mat. b.Using plastic pipet, transfer enough sample into cup to fill at least halfway. c.Place yellow x-ray film over top of cell. d.Affix retaining ring (flange down) onto cell ensuring x-ray film is taut and no wrinkles are present in film. e.Flip cell over and puncture top with thumbtack. f.Repeat steps 5.1-5.5 until all samples are prepared – label sample cups if necessary. EquipmentMaterials and Reagents Scope / Objective Note: This procedure is only valid at the time of printing. The control version can be found on the Meraux Refinery intranet. Completed Date:Prepared By:Approved By:

2 AreaUnitOperating Procedure #Certified byRevision #DatePage 3LabOPS-LB-PS-01809.20.20112 of 2 Sulfur by Horiba SLFA-2800 Continued 6 Load samples in instrument as follows: a.Push lever located under sample-chamber lid to the open position and lift open cover. b.Load samples in carousel in positions corresponding to sample IDs typed in previously. c.Close sample-chamber lid ensuring lid is completely closed. 7With the READY menu displayed, press the MEASURE key located to right of control panel. 8 Upon completion of the measurement(s), a beep sound is heard – sample cups must be removed immediately once analysis of samples is finished. NOTE: Do not leave sample cups in x-ray as moisture will develop inside chamber. 9Discard spent samples into trash can. Note: This procedure is only valid at the time of printing. The control version can be found on the Meraux Refinery intranet. Completed Date:Prepared By:Approved By:


Download ppt "PPE Use Standard (SSP-0011) Y Full Face Shield / Goggles Hood Clothing—Chemical Protection Gloves—Chemical Personal Fall Protection Respirator / Filter."