Download presentation
Presentation is loading. Please wait.
Published byAdela Lee Modified over 9 years ago
1
Microplate reader spectrophotometer
2
The Beer-Lambert Law A=abc Now let us look at the Beer-Lambert law and explore it's significance. A is absorbance (no units) “a” is the molar absorbtivity with units of L mol -1 cm -1 “b” is the path length of the sample - that is, the path length of the cuvette (cm) “c” is the concentration of the compound in solution, expressed in mol L-1
3
Light Source: Must be able to generate many wavelengths. The light goes through a monochromator excitation filter to make the light beam a single wavelength.
4
Absorbance or transmittance can be measured with a microplate reader. Sample: The sample is in a microplate which is transparent to the light. This means only the sample interacts with the single wavelength of light. Detector: The detector is a photoelectric cell. When light strikes the detector, a flow of electrons (a current) is created in the detector. The current generated is proportional to the light intensity. The detector functions as a result of the light transmitted through the emission filter.
5
Example: The absorbance maxima of DNA is at 260 nm. This is what we would set our reader at.
6
The absorbance maxima of this substance is 300nm. This is the wavelength that the microplate reader should be set at. You should know this ahead of time!
7
The Absorbance Maxima for Food Coloring
8
The Whole Process
9
Using a Spectrophotometer
10
Light Source: Must be able to generate many wavelengths. The light goes through a monochromator excitation filter to make the light beam a single wavelength.
11
Absorbance or transmittance can be measured with a spectrophotometer. Sample: The sample is in a cuvette which is transparent to the light. This means only the sample interacts with the single wavelength of light. Detector: The detector is a photoelectric cell. When light strikes the detector, a flow of electrons (a current) is created in the detector. The current generated is proportional to the light intensity. The detector functions as a result of the light transmitted through the emission filter. Readings are absorbance or transmittance.
12
Example: The absorbance maxima of DNA is at 260 nm. This is what we would set our reader at.
13
The Absorbance Maxima for Food Coloring
Similar presentations
© 2025 SlidePlayer.com. Inc.
All rights reserved.