Inspection Camera Z.C. Liu, Z.Q. Li, D. Li, J. Gao IHEP 2010-12-7
2
Inspection cavity type 1.3GHz single cell elliptical cavity 1.3GHz multi-cell elliptical cavity Low frequency elliptical cavity like 500MHz, 650MHz… Other type like spoke cavity…
3
Inspecting what? Equator welding seams ~3-6 mm wide Iris welding seams ~3-6 mm wide Pits, cratches… Multi-tens or multi-hundreds microns A resolution of multi- microns is needed for the inspection system
4
Support tube Sketch map Camera lens 2x lens +1 close-up lens Mirror Monitor Linear rail table Holding cavity Cavity moving direction
5
Image precision analyzing Focal length ( mm ) Precision ( um/pixel ) ( 21 million pixels camera ) 0.9m1.3m1.5m1.8m2m2.2m 30012.821.325.632.036.340.6 400-14.417.622.425.628.8 500---16.719.221.8 6003.27.49.612.814.917.1 800-45.68.09.611.2 1000---4.96.47.7 Camera + long focal length lens + 2x lens: Note: 1 ) ”-” can not form an image 2 ) Blue:equal to kyoto camera (7.5-15um/pixel);Red:higher than kyoto camera. 3 ) Long focal length lens: SIGMA APO 50-500mm F4.5-6.3 DG OS HSM. 4) The image precision will be increased while using +1 close-up lens, for example, with +1 close-up lens and a 500mm focal length, the precision is 0.64um/pixel at 550mm away from the camera CCD. However more lens the picture sharpness more worse.
11
Light system 1) It is not easy to get a good picture by only changing the brightness of the light in region 1 2) It is easy to get a good picture in region 2 and 3 Too bright Good ① ② ③ Light Camera position Mirror-like surface
12
Light system 1) It is easier to get a good picture by changing the brightness of the light in region 1 than without ground glass 2) It is easy to get a good picture in region 2 and 3 Bright Good ① ② ③ Light Camera position Ground glass Mirror-like surface
13
Light system 1) It is not easy to get a good picture at a position for the whole inspection area. Bright Good Light Camera position Ground glass Rough mirror-like surface Bright
14
Light solution Tunable terms Light brightness Light angle Light source Light spectrum Solution: Light brightness Ground glass, paper, reflection Point –like light source Led light Light source movable
15
Light system 1) moving the light source to an angle with no direct reflection from the inspection area to the camera. 2) Reflection light from different material will be tested as light source 3) take different light angle pictures to form a composite one. Bright Good Light Camera position Ground glass Or paper etc.. Rough mirror-like surface Bright
16
Image of plane picture without close-up lens 1mm 200um
17
Image of iris without close-up lens 1mm A resolution of 9um/pixel. 200um
18
Cost of the system Cost($) Camera and lens5500 Support and transfer system25000 Small parts500 Total31000
19
Problems need to be solved The camera is shaking while taking pictures even with tripod. The sharpness of the image need to be improved. Light system need to be improved for equator inspection The real resolution of the system now is only about half of the calculation results. Using close-up lens.